Monte Carlo and water calorimetric determination of kilovoltage beam radiotherapy ionization chamber correction factors

被引:5
|
作者
Bancheri, Julien [1 ]
Ketelhut, Steffen [2 ]
Buermann, Ludwig [2 ]
Seuntjens, Jan [1 ]
机构
[1] McGill Univ, Med Phys Unit, Cedars Canc Ctr, Montreal, PQ H4A 3J1, Canada
[2] Phys Tech Bundesanstalt PTB, Bundesallee 100, Braunschweig, Germany
基金
加拿大自然科学与工程研究理事会;
关键词
medical physics; kilovoltage dosimetry; reference dosimetry; ionization chambers; correction factors; water calorimetry; radiation primary standard; ENERGY-ABSORPTION COEFFICIENTS; X-RAYS; TRANSPORT; DOSIMETRY; RATIOS; DEPTH;
D O I
10.1088/1361-6560/ab82e7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The in-phantom calibration method for radiotherapy kilovoltage x-ray beams requires ionization chamber correction factors. The overall ionization chamber correction factor accounts for changes in the chamber response due to the displacement of water by the chamber cavity and wall, the presence of the stem and the change in incident photon energy and angular distribution in the phantom to that in air. A waterproof sheath, if required, is accounted for in a sheath correction factor. The aim of this study is to determine chamber correction factors through Monte Carlo (MC) simulations and water calorimetry measurements. Correction factors are determined for the PTW TM30013, NE2571, IBA FC65-G, IBA FC65-P and Exradin A12 ionization chambers. They are compared to experimental values obtained at the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB) with their water calorimetry-based absorbed dose to water primary standard and at other national metrological institutes. An uncertainty analysis considers the contributions to the uncertainty on the chamber correction factors from the field size, photon cross sections, photon fluence spectra and chamber wall and central electrode dimensions. The MC calculated chamber correction factors are within 2.2% of unity with a standard uncertainty of 0.3%. For the 50 kV and 100 - 140 kV radiation beam qualities, the calculated correction factors deviate from the measured correction factors (with a standard uncertainty of 1%) by up to 2.6%. The calculated chamber correction factors for the PTW TM30013 and Exradin A12 are consistent with those derived from the BIPM kilovoltage primary standard. The inconsistencies between the calculated and experimental chamber correction factors indicate the need to further investigate the accuracy of kilovoltage absorbed dose to water primary standards and the use of MC simulations to determine kilovoltage beam chamber correction factors.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Monte Carlo study of ionization chamber magnetic field correction factors as a function of angle and beam quality
    Malkov, Victor N.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2018, 45 (02) : 908 - 925
  • [2] Monte Carlo calculation of beam quality correction factors for PTW cylindrical ionization chambers in photon beams
    Gimenez-Alventosa, Vicent
    Gimenez, Vicente
    Ballester, Facundo
    Vijande, Javier
    Andreo, Pedro
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (20)
  • [3] Monte Carlo simulated beam quality and perturbation correction factors for ionization chambers in monoenergetic proton beams
    Kretschmer, Jana
    Dulkys, Anna
    Brodbek, Leonie
    Stelljes, Tenzin Sonam
    Looe, Hui Khee
    Poppe, Bjoern
    MEDICAL PHYSICS, 2020, 47 (11) : 5890 - 5905
  • [4] Monte Carlo-derived ionization chamber correction factors in therapeutic carbon ion beams
    Khan, Ahtesham Ullah
    Simiele, Eric A.
    DeWerd, Larry A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (19)
  • [5] Kilovoltage beam Monte Carlo dose calculations in submillimeter voxels for small animal radiotherapy
    Bazalova, Magdalena
    Zhou, Hu
    Keall, Paul J.
    Graves, Edward E.
    MEDICAL PHYSICS, 2009, 36 (11) : 4991 - 4999
  • [6] Monte Carlo calculated beam quality correction factors for two cylindrical ionization chambers in photon beams
    Alissa, Mohamad
    Zink, Klemens
    Tessier, Frederic
    Schoenfeld, Andreas A.
    Czarnecki, Damian
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 94 : 17 - 23
  • [7] An Ionization Chamber Array with High Spatial Resolution for External Beam Radiotherapy
    Togno, M.
    Wilkens, J.
    Menichelli, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 321 - 321
  • [8] Monte Carlo calculated ionization chamber correction factors in clinical proton beams - deriving uncertainties from published data
    Baumann, Kilian-Simon
    Goma, Carles
    Wulff, Joerg
    Kretschmer, Jana
    Zink, Klemens
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 113
  • [9] Calculation of beam quality correction factors for various thimble ionization chambers using the Monte Carlo code PENELOPE
    Erazo, Fabian
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (02): : 163 - 170
  • [10] Monte Carlo calculation of beam quality correction factors in proton beams using detailed simulation of ionization chambers
    Goma, Carles
    Andreo, Pedro
    Sempau, Josep
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (06) : 2389 - 2406