Comparison of Vascular Responses Following New-Generation Biodegradable and Durable Polymer-Based Drug-Eluting Stent Implantation in an Atherosclerotic Rabbit Iliac Artery Model

被引:36
|
作者
Nakazawa, Gaku [1 ]
Torii, Sho [1 ]
Ijichi, Takeshi [1 ]
Nagamatsu, Hirofumi [1 ]
Ohno, Yohei [1 ]
Kurata, Fumi [1 ]
Yoshikawa, Ayako [1 ]
Nakano, Masataka [1 ]
Shinozaki, Norihiko [1 ]
Yoshimachi, Fuminobu [1 ]
Ikari, Yuji [1 ]
机构
[1] Tokai Univ, Sch Med, Dept Cardiol, 143 Shimokasuya, Isehara, Kanagawa 2591193, Japan
来源
JOURNAL OF THE AMERICAN HEART ASSOCIATION | 2016年 / 5卷 / 10期
关键词
atherosclerosis; endothelialization; polymer; stent; BARE-METAL STENTS; MYOCARDIAL-INFARCTION; SPIRIT III; THROMBOSIS; RESTENOSIS; PATHOLOGY; TRIAL; PREDICTORS; MICRORNAS; HUMANS;
D O I
10.1161/JAHA.116.003803
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Incomplete endothelialization is the primary substrate of late stent thrombosis; however, recent reports have revealed that abnormal vascular responses are also responsible for the occurrence of late stent failure. The aim of the current study was to assess vascular response following deployment of biodegradable polymer-based Synergy (Boston Scientific) and Nobori (Terumo) drug-eluting stents and the durable polymer-based Resolute Integrity stent (Medtronic) in an atherosclerotic rabbit iliac artery model. Methods and Results-A total of 24 rabbits were fed an atherogenic diet, and then a balloon injury was used to induce atheroma formation. Synergy, Nobori, and Resolute Integrity stents were randomly implanted in iliac arteries. Animals were euthanized at 28 days for scanning electron microscopic evaluation and at 90 days for histological analysis. The percentage of uncovered strut area at 28 days was lowest with Synergy, followed by Resolute Integrity, and was significantly higher with Nobori stents (Synergy 1.1 +/- 2.2%, Resolute Integrity 2.0 +/- 3.9%, Nobori 4.6 +/- 3.0%; P< 0.001). At 90 days, inflammation score was lowest for Synergy (0.27 +/- 0.45), followed by Nobori (0.62 +/- 0.59), and was highest for Resolute Integrity (0.89 +/- 0.46, P< 0.001). Foamy macrophage infiltration within neointima (ie, neoatherosclerosis) was significantly less with Synergy (0.62 +/- 0.82) compared with Nobori (0.85 +/- 0.74) and Resolute Integrity (1.39 +/- 1.32; P=0.034). Conclusions-The biodegradable polymer-coated thin-strut Synergy drug-eluting stent showed the fastest stent strut neointimal coverage and the lowest incidence of neoatherosclerosis in the current animal model.
引用
收藏
页数:8
相关论文
共 50 条
  • [11] Meta-Analysis of Randomized Clinical Trials Comparing Biodegradable Polymer Drug-Eluting Stent to Second-Generation Durable Polymer Drug- Eluting Stents
    El-Hayek, Georges
    Bangalore, Sripal
    Dominguez, Abel Casso
    Devireddy, Chandan
    Jaber, Wissam
    Kumar, Gautam
    Mavromatis, Kreton
    Tamis-Holland, Jacqueline
    Samady, Habib
    JACC-CARDIOVASCULAR INTERVENTIONS, 2017, 10 (05) : 462 - 473
  • [12] Thrombogenicity and Early Vascular Healing Response in Metallic Biodegradable Polymer-Based and Fully Bioabsorbable Drug-Eluting Stents
    Koppara, Tobias
    Cheng, Qi
    Yahagi, Kazuyuki
    Mori, Hiroyoshi
    Sanchez, Oscar David
    Feygin, Julia
    Wittchow, Eric
    Kolodgie, Frank D.
    Virmani, Renu
    Joner, Michael
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2015, 8 (06)
  • [13] Late neointimal volume reduction is observed following biodegradable polymer-based drug eluting stent in porcine model
    Ijichi, Takeshi
    Nakazawa, Gaku
    Torii, Sho
    Nagamatsu, Hirofumi
    Yoshikawa, Ayako
    Nakamura, Shintaro
    Souba, Junko
    Isobe, Atsushi
    Hagiwara, Hitomi
    Ikari, Yuji
    IJC HEART & VASCULATURE, 2021, 34
  • [14] Outcomes of Acute Myocardial Infarction Patients Implanted With Biodegradable Polymer Biolimus-Eluting Stents Versus New-Generation Durable Polymer Drug-Eluting Stents: A Retrospective Analysis
    Choe, Jeong Cheon
    Cha, Kwang Soo
    Jang, Hye Yoon
    Choi, Jong Hyun
    Kim, Bo Won
    Ahn, Jinhee
    Park, Jin Sup
    Lee, Hye Won
    Oh, Jun-Hyok
    Choi, Jung Hyun
    Lee, Han Cheol
    Hong, Taek Jong
    Ahn, Youngkeun
    Jeong, Myung Ho
    ANGIOLOGY, 2017, 68 (08) : 698 - 706
  • [15] Long-Term Vascular Responses to Resolute® and Xience V® Polymer-Based Drug-Eluting Stents in a Rabbit Model of Atherosclerosis
    Hoymans, Vicky Y.
    Van Dyck, Christophe J.
    Haine, Steven E.
    Frederix, Geert
    Fransen, Erik
    Timmermans, Jean-Pierre
    Vrints, Christiaan J.
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2014, 27 (04) : 381 - 390
  • [16] Strut Coverage and Vessel Wall Response to a New-Generation Paclitaxel-Eluting Stent With an Ultrathin Biodegradable Abluminal Polymer Optical Coherence Tomography Drug-Eluting Stent Investigation (OCTDESI)
    Guagliumi, Giulio
    Sirbu, Vasile
    Musumeci, Giuseppe
    Bezerra, Hiram G.
    Aprile, Alessandro
    Kyono, Hiroyuki
    Fiocca, Luigi
    Matiashvili, Aleksandre
    Lortkipanidze, Nikoloz
    Vassileva, Angelina
    Popma, Jeffrey J.
    Allocco, Dominic J.
    Dawkins, Keith D.
    Valsecchi, Orazio
    Costa, Marco A.
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (04) : 367 - U154
  • [17] Vascular responses to coronary calcification following implantation of newer-generation drug-eluting stents in humans: impact on healing
    Torii, Sho
    Jinnouchi, Hiroyuki
    Sakamoto, Atsushi
    Mori, Hiroyoshi
    Park, Joohyung
    Amoa, Falone C.
    Sawan, Mariem
    Sato, Yu
    Cornelissen, Anne
    Kuntz, Salome H.
    Kutyna, Matthew
    Paek, Ka Hyun
    Fernandez, Raquel
    Braumann, Ryan
    Mont, Eric K.
    Surve, Dipti
    Romero, Maria E.
    Kolodgie, Frank D.
    Virmani, Renu
    Finn, Aloke, V
    EUROPEAN HEART JOURNAL, 2020, 41 (06) : 786 - +
  • [18] Tacrolimus-eluting stent with biodegradable polymer is more effective than sirolimus- and everolimus-eluting stent in rabbit iliac artery restenosis model
    Dae Sung Park
    Jun-Kyu Park
    Myung Ho Jeong
    In-Ho Bae
    So-Youn Lee
    Eun Jae Jang
    Kyung-Seob Lim
    Jong Min Kim
    Jung Ha Kim
    Dae Young Hyun
    Yun Ah Jeong
    Hyun Kuk Kim
    Doo Sun Sim
    Macromolecular Research, 2015, 23 : 1034 - 1041
  • [19] Tacrolimus-eluting stent with biodegradable polymer is more effective than sirolimus- and everolimus-eluting stent in rabbit iliac artery restenosis model
    Park, Dae Sung
    Park, Jun-Kyu
    Jeong, Myung Ho
    Bae, In-Ho
    Lee, So-Youn
    Jang, Eun Jae
    Lim, Kyung-Seob
    Kim, Jong Min
    Kim, Jung Ha
    Hyun, Dae Young
    Jeong, Yun Ah
    Kim, Hyun Kuk
    Sim, Doo Sun
    MACROMOLECULAR RESEARCH, 2015, 23 (11) : 1034 - 1041
  • [20] In vivo comparison of a polymer-free Biolimus A9-eluting stent with a biodegradable polymer-based Biolimus A9 eluting stent and a bare metal stent in balloon denuded and radiated hypercholesterolemic rabbit iliac arteries
    Waksman, Ron
    Pakala, Rajbabu
    Baffour, Richard
    Seabron, Rufus
    Hellinga, David
    Chan, Rosanna
    Su, Shih-Horng
    Kolodgie, Frank
    Virmani, Renu
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 80 (03) : 429 - 436