Quantification of Renal Perfusion: Comparison of Arterial Spin Labeling and Dynamic Contrast-Enhanced MRI

被引:50
作者
Winter, Jeff D. [1 ]
St Lawrence, Keith S. [2 ,3 ]
Cheng, Hai-Ling Margaret [1 ,4 ]
机构
[1] Hosp Sick Children, Res Inst, Toronto, ON M5G 1X8, Canada
[2] Lawson Hlth Res Inst, Imaging Div, London, ON, Canada
[3] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
[4] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
perfusion; arterial spin labeling; dynamic contrast enhanced MRI; kidney; cortex; rabbit; BLOOD-FLOW; KIDNEYS;
D O I
10.1002/jmri.22660
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To provide the first comparison of absolute renal perfusion obtained by arterial spin labeling (ASL) and separable compartment modeling of dynamic contrast-enhanced (DCE) magnetic resonance imaging (MRI). Moreover, we provide the first application of the dual bolus approach to quantitative DCE-MRI perfusion measurements in the kidney. Materials and Methods: Consecutive ASL and DCE-MRI acquisitions were performed on six rabbits on a 1.5 T MRI system. Gadolinium (Gd)-DTPA was administered in two separate injections to decouple measurement of the arterial input function and tissue uptake curves. For DCE perfusion, pixel-wise and mean cortex region-of-interest tissue curves were fit to a separable compartment model. Results: Absolute renal cortex perfusion estimates obtained by DCE and ASL were in close agreement: 3.28 +/- 0.59 mL/g/min (ASL), 2.98 +/- 0.60 mL/g/min (DCE), and 3.57 +/- 0.96 mL/g/min (pixel-wise DCE). Renal medulla perfusion was 1.53 +/- 0.35 mL/g/min (ASL) but was not adequately described by the separable compartment model. Conclusion: ASL and DCE-MRI provided similar measures of absolute perfusion in the renal cortex, offering both noncontrast and contrast-based alternatives to improve current renal MRI assessment of kidney function.
引用
收藏
页码:608 / 615
页数:8
相关论文
共 50 条
  • [41] Dynamic contrast-enhanced MRI quantification of synovium microcirculation in experimental arthritis
    Doria, AS
    Noseworthy, M
    Oakden, W
    Moineddin, R
    Rayner, T
    Tassos, V
    Engelberts, D
    Pritzker, K
    Rogers, M
    Jong, R
    Babyn, P
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (04) : 1165 - 1171
  • [42] Multi-delay arterial spin labeling perfusion MRI in moyamoya disease–comparison with CT perfusion imaging
    Rui Wang
    Songlin Yu
    Jeffry R. Alger
    Zhentao Zuo
    Juan Chen
    Rong Wang
    Jing An
    Bo Wang
    Jizong Zhao
    Rong Xue
    Danny J. J. Wang
    European Radiology, 2014, 24 : 1135 - 1144
  • [43] Operator dependency of arterial input function in dynamic contrast-enhanced MRI
    Kleppesto, Magne
    Bjornerud, Atle
    Groote, Inge Rasmus
    Kim, Minjae
    Vardal, Jonas
    Larsson, Christopher
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2022, 35 (01) : 105 - 112
  • [44] Correction of arterial input function in dynamic contrast-enhanced MRI of the liver
    Wang, Hesheng
    Cao, Yue
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (02) : 411 - 421
  • [45] Operator dependency of arterial input function in dynamic contrast-enhanced MRI
    Magne Kleppestø
    Atle Bjørnerud
    Inge Rasmus Groote
    Minjae Kim
    Jonas Vardal
    Christopher Larsson
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2022, 35 : 105 - 112
  • [46] Quantitation of renal perfusion using arterial spin labeling with FAIR-UFLARE
    Karger, N
    Biederer, J
    Lüsse, S
    Grimm, J
    Steffens, JC
    Heller, M
    Glüer, CC
    MAGNETIC RESONANCE IMAGING, 2000, 18 (06) : 641 - 647
  • [47] Comparison of tumor blood perfusion assessed by dynamic contrast-enhanced MRI with tumor blood supply assessed by invasive imaging
    Graff, BA
    Benjaminsen, IC
    Brurberg, KG
    Ruud, EBM
    Rofstad, EK
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (03) : 272 - 281
  • [48] Medullary hemangioblastoma in a child with von Hippel-Lindau disease: vascular tumor perfusion depicted by arterial spin labeling and dynamic contrast-enhanced imaging
    Goo, Hyun Woo
    Ra, Young-Shin
    JOURNAL OF NEUROSURGERY-PEDIATRICS, 2015, 16 (01) : 50 - 53
  • [49] Absolute quantification of cerebral blood flow: correlation between dynamic susceptibility contrast MRI and model-free arterial spin labeling
    Knutsson, Linda
    van Westen, Danielle
    Petersen, Esben T.
    Bloch, Karin Markenroth
    Holtas, Stig
    Stahlberg, Freddy
    Wirestam, Ronnie
    MAGNETIC RESONANCE IMAGING, 2010, 28 (01) : 1 - 7
  • [50] Morphological, contrast-enhanced and spin labeling perfusion imaging for monitoring of relapse after RF ablation of renal cell carcinomas
    Boss, Andreas
    Martirosian, Petros
    Schraml, Christina
    Clasen, Stephan
    Fenchel, Michael
    Anastasiadis, Artistotelis
    Claussen, Claus D.
    Pereira, Philippe L.
    Schick, Fritz
    EUROPEAN RADIOLOGY, 2006, 16 (06) : 1226 - 1236