Rapid decline in acute stimulation thresholds with steroid-eluting active-fixation pacing leads

被引:33
作者
Kistler, PM [1 ]
Kalman, JM [1 ]
Fynn, SP [1 ]
Singarayar, S [1 ]
Roberts-Thomson, KC [1 ]
Lindsay, CB [1 ]
Khong, U [1 ]
Sparks, PB [1 ]
Strathmore, N [1 ]
Mond, HG [1 ]
机构
[1] Royal Melbourne Hosp, Dept Cardiol, Melbourne, Vic 3050, Australia
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2005年 / 28卷 / 09期
关键词
pacing; pacemakers; surgery;
D O I
10.1111/j.1540-8159.2005.00209.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Rapid Decline in Acute Stimulation Thresholds with Steroid-Eluting Active-Fixation Pacing Leads. Background and Aim: There is an increasing use of active-fixation leads for cardiac pacing, yet concerns remain regarding initial high stimulation thresholds. The aim was to perform a detailed analysis of pacing parameters at the time of implantation to determine when lead repositioning should be considered. Methods: We performed a prospective observational study of consecutive new pacemaker implants. Detailed analysis of pacing parameters was collected at 2-minute intervals for 10 minutes, and at day 1 and week 8 following implant. Results: Ninety-four patients underwent implantation of 79 dual-chamber and 15 single-chamber pacemakers using active-fixation leads in both chambers. An initial threshold of > 1 V was demonstrated in 45194 (48%) ventricular leads (mean threshold 1.5 +/- 0.3 V). This declined rapidly to 0.9 +/- 0.3 V at 4 minutes (P < 0.01), 0.7 +/- 0.3 V at 10 minutes (P < 0.01), and 0.6 +/- 0.3 V at day 1 (P < 0.01). At day 1, 43145 leads were < 1 V There were 79 atrial leads. An initial threshold of > 1 V (mean 1.7 +/- 0.6 V) was demonstrated in 41/79 (52%) leadsfalling significantly to 1.1 +/- 0.5 V at 4 minutes (P < 0.01), 0.9 +/- 0.4 V at 10 minutes (P < 0.01), and 0.6 +/- 0.2 V at day 1 (P < 0.01). At 10 minutes, 32 of 41 leads demonstrated a threshold of < 1 V with all leads < 1 V at day 1. Thresholds were maintained medium term. Conclusions: Active-fixation leads are commonly associated with initially high thresholds that fall rapidly. An initial threshold of 2 V should be provisionally accepted and retested at 4 minutes. The majority will have a threshold of < 1 V the following day. A failure of a high threshold to decline at 4 minutes requires lead repositioning.
引用
收藏
页码:903 / 909
页数:7
相关论文
共 26 条
  • [11] ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD
    GREENE, TO
    PORTNOW, AS
    HUANG, SKS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01): : 21 - 25
  • [12] Hidden-Lucet F, 2000, PACE, V23, P1798
  • [13] Chronic steroid-eluting lead performance: A comparison of atrial and ventricular pacing
    Hua, W
    Mond, HG
    Strathmore, N
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 17 - 24
  • [14] THE RIGHT ATRIAL FREE WALL - AN ALTERNATIVE PACING SITE
    JAMIDAR, H
    GOLI, V
    REYNOLDS, DW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (05): : 959 - 963
  • [15] Single center experience with femoral extraction of permanent endocardial pacing leads
    Jarwe, M
    Klug, D
    Beregi, JP
    Le Franc, P
    Lacroix, D
    Kouakam, C
    Guédon-Moreau, L
    Zghal, N
    Kacet, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (08): : 1202 - 1209
  • [16] ACTIVE FIXATION ATRIAL LEADS - RANDOMIZED COMPARISON OF 2 LEAD DESIGNS
    KAY, GN
    ANDERSON, K
    EPSTEIN, AE
    PLUMB, VJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (08): : 1355 - 1361
  • [17] Selective site pacing: Defining and reaching the selected site
    Lieberman, R
    Grenz, D
    Mond, HG
    Gammage, MD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 883 - 886
  • [18] Implantable transvenous pacing leads: The shape of things to come
    Mond, HG
    Grenz, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 887 - 893
  • [19] Selective site pacing: The future of cardiac pacing?
    Mond, HG
    Gammage, MD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 835 - 836
  • [20] MOND HG, 1992, PACE, V15, P95