Rapid decline in acute stimulation thresholds with steroid-eluting active-fixation pacing leads

被引:33
作者
Kistler, PM [1 ]
Kalman, JM [1 ]
Fynn, SP [1 ]
Singarayar, S [1 ]
Roberts-Thomson, KC [1 ]
Lindsay, CB [1 ]
Khong, U [1 ]
Sparks, PB [1 ]
Strathmore, N [1 ]
Mond, HG [1 ]
机构
[1] Royal Melbourne Hosp, Dept Cardiol, Melbourne, Vic 3050, Australia
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2005年 / 28卷 / 09期
关键词
pacing; pacemakers; surgery;
D O I
10.1111/j.1540-8159.2005.00209.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Rapid Decline in Acute Stimulation Thresholds with Steroid-Eluting Active-Fixation Pacing Leads. Background and Aim: There is an increasing use of active-fixation leads for cardiac pacing, yet concerns remain regarding initial high stimulation thresholds. The aim was to perform a detailed analysis of pacing parameters at the time of implantation to determine when lead repositioning should be considered. Methods: We performed a prospective observational study of consecutive new pacemaker implants. Detailed analysis of pacing parameters was collected at 2-minute intervals for 10 minutes, and at day 1 and week 8 following implant. Results: Ninety-four patients underwent implantation of 79 dual-chamber and 15 single-chamber pacemakers using active-fixation leads in both chambers. An initial threshold of > 1 V was demonstrated in 45194 (48%) ventricular leads (mean threshold 1.5 +/- 0.3 V). This declined rapidly to 0.9 +/- 0.3 V at 4 minutes (P < 0.01), 0.7 +/- 0.3 V at 10 minutes (P < 0.01), and 0.6 +/- 0.3 V at day 1 (P < 0.01). At day 1, 43145 leads were < 1 V There were 79 atrial leads. An initial threshold of > 1 V (mean 1.7 +/- 0.6 V) was demonstrated in 41/79 (52%) leadsfalling significantly to 1.1 +/- 0.5 V at 4 minutes (P < 0.01), 0.9 +/- 0.4 V at 10 minutes (P < 0.01), and 0.6 +/- 0.2 V at day 1 (P < 0.01). At 10 minutes, 32 of 41 leads demonstrated a threshold of < 1 V with all leads < 1 V at day 1. Thresholds were maintained medium term. Conclusions: Active-fixation leads are commonly associated with initially high thresholds that fall rapidly. An initial threshold of 2 V should be provisionally accepted and retested at 4 minutes. The majority will have a threshold of < 1 V the following day. A failure of a high threshold to decline at 4 minutes requires lead repositioning.
引用
收藏
页码:903 / 909
页数:7
相关论文
共 26 条
[11]   ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD [J].
GREENE, TO ;
PORTNOW, AS ;
HUANG, SKS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :21-25
[12]  
Hidden-Lucet F, 2000, PACE, V23, P1798
[13]   Chronic steroid-eluting lead performance: A comparison of atrial and ventricular pacing [J].
Hua, W ;
Mond, HG ;
Strathmore, N .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01) :17-24
[14]   THE RIGHT ATRIAL FREE WALL - AN ALTERNATIVE PACING SITE [J].
JAMIDAR, H ;
GOLI, V ;
REYNOLDS, DW .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (05) :959-963
[15]   Single center experience with femoral extraction of permanent endocardial pacing leads [J].
Jarwe, M ;
Klug, D ;
Beregi, JP ;
Le Franc, P ;
Lacroix, D ;
Kouakam, C ;
Guédon-Moreau, L ;
Zghal, N ;
Kacet, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (08) :1202-1209
[16]   ACTIVE FIXATION ATRIAL LEADS - RANDOMIZED COMPARISON OF 2 LEAD DESIGNS [J].
KAY, GN ;
ANDERSON, K ;
EPSTEIN, AE ;
PLUMB, VJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (08) :1355-1361
[17]   Selective site pacing: Defining and reaching the selected site [J].
Lieberman, R ;
Grenz, D ;
Mond, HG ;
Gammage, MD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06) :883-886
[18]   Implantable transvenous pacing leads: The shape of things to come [J].
Mond, HG ;
Grenz, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06) :887-893
[19]   Selective site pacing: The future of cardiac pacing? [J].
Mond, HG ;
Gammage, MD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06) :835-836
[20]  
MOND HG, 1992, PACE, V15, P95