European multicentre evaluation of the super aution SA-4220 urinalysis analyser

被引:5
作者
Galimany, R
Arambarri, M
Biosca, C
Ceriotti, F
Chapelle, JP
Lopez, R
Souverijn, JHM
Spagnuolo, G
Scopacasa, F
Almirall, L
机构
[1] Hosp Univ Germans Trias & Pujol, Serv Bioquim Clin, E-08916 Barcelona, Spain
[2] Ist Sci H San Raffaele, Milan, Italy
[3] Ctr Hosp Univ Liege, Liege, Belgium
[4] ICS, DAP St Andreu, Lab Bon Pastor, Barcelona, Spain
[5] Leiden Univ, Med Ctr, Leiden, Netherlands
[6] Univ Naples Federico II, Naples, Italy
[7] A Menarini Diagnost, Applicat Lab, St Adria Del Besos, Spain
关键词
urinalysis; automation; multicentre evaluation;
D O I
10.1515/CCLM.1998.164
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A multicentre evaluation of the urine test strip analyser Super Aution-4220 was carried out in six laboratories. The analytical performance of the instrument with regard to imprecision, linearity, detection limit, drift, carry-over and method comparison was studied. Using the Aution stick 8 test strip the pH, glucose, protein, ketones, bilirubin, blood, urobilinogen and leukocyte esterase were analysed. Specific gravity measurements were performed by refractive index method. Within-run and between-run imprecision determined at three levels of analyte were good. No carry-over was observed. Obtained results were linear through all the described analytical range. No significant drift was detected. Method comparison with some quantitative methods was performed and showed a good correlation with most of the analytes. The study of interferences showed minor interferences by common therapeutic drugs with the measurement of some analytes. During the assessment period of about 6 months no breakdown occurred in any laboratory. The Super Aution urine analyser appeared to be a highly automated analyser of urinary test strips. The operation was simple and the maintenance required only a few minutes a day.
引用
收藏
页码:947 / 957
页数:11
相关论文
共 9 条
[1]  
ALUMA A, 1994, SELECCION EVALUACION, P45
[2]   GUIDELINES FOR EVALUATION OF REAGENT STRIPS - EXEMPLIFIED BY ANALYSIS OF URINE ALBUMIN AND GLUCOSE-CONCENTRATION USING VISUALLY READ REAGENT STRIPS [J].
BERG, B ;
HELLSING, K ;
JAGENBURG, R ;
KALLNER, A .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1989, 49 (07) :689-699
[3]   CRITERIA FOR THE ASSESSMENT OF ANALYZER PRACTICABILITY [J].
BIOSCA, C ;
GALIMANY, R ;
ALSINA, MJ ;
ALUMA, A ;
ARMENTER, C ;
BRAVO, J ;
DOLADE, M ;
FERRE, J ;
MARTINEZ, M .
JOURNAL OF AUTOMATIC CHEMISTRY, 1993, 15 (05) :167-170
[4]  
BONINI P, 1998, J AUTOM CHEM, V10, P121
[5]  
COCA C, 1986, ANAL CLIN, V44, P239
[6]  
GOMEZ C, 1990, ESTADISTICA APLICADA
[7]  
HAECKEL R, 1985, J CLIN CHEM CLIN BIO, V23, P473
[8]  
*NAT COMM CLIN LAB, 1991, 16P NCCLS GP
[9]  
TRYDING N, 1996, DRUG EFFECTS CLIN CH