Biomechanical comparison of monocortical and bicortical plate fixation for rib fractures in a cadaveric model using a locking plate system

被引:12
作者
Choke, Abby [1 ]
Wong, Yoke Rung [2 ]
Joethy, Janna-Vale [1 ]
机构
[1] Singapore Gen Hosp, Dept Plast Reconstruct & Aesthet Surg, 20 Coll Rd, Singapore 169856, Singapore
[2] Singapore Gen Hosp, Biomech Lab, Singapore 169856, Singapore
关键词
Rib fracture; bicortical; monocortical; unicortical; locking plate; FLAIL CHEST; STABILIZATION; OSTEOSYNTHESIS; MANAGEMENT; WALL;
D O I
10.21037/jtd.2019.12.31
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Surgical stabilization of rib fractures is an established form of treatment for complex rib fractures. Plate fixation with bicortical screws placement can cause injury to intra-thoracic organs and pleural irritation from protruding screw tips. The aim of this study is to compare the biomechanical properties of monocortical and bicortical plate fixation for rib fractures using a locking plate system. Methods: Ten pairs of fresh-frozen cadaveric ribs were harvested. Native ribs were mounted onto a biomechanical tester and statically loaded to failure to induce a rib fracture. The native stiffness of the rib was measured. Next, the ribs were stabilized using the Synthes MatrixRIB (Johnson & Johnson, USA) locking plate. Left-sided ribs were fixed in a bicortical manner and right-sided ribs were fixed in a monocortical manner. The repaired ribs were subjected to cyclic loading of 50,000 cycles between 2 to 6 N to simulate physiological respiration, followed by static loading at a rate of 10 N/min until failure. The pre and postrepaired stiffness were measured. A high-speed camera was used to record the mechanism of failure. Results: One left-sided rib was omitted from the study because the fracture occurred at the drill hole site. Left-sided ribs demonstrated a mean native stiffness of 10.0 N/mm (SD 3.71) and right-sided 11.92 N/mm (SD 3.57). After plate fixation, pre and post cyclic stiffness was 3.32 N/mm (SD 1.21) and 4.41 N/mm (SD 3.59) for the bicortical group; 3.14 N/mm (SD 1.24) and 3.91 N/mm (SD 1.98) for the monocortical group. There is no statistical difference found between the two groups (P=0.872). Conclusions: Our results show that there is no difference in stability between monocortical and bicortical fixation for rib fractures using a locking plate system. Monocortical fixation is recommended to avoid potential complications.
引用
收藏
页码:4966 / 4971
页数:6
相关论文
共 50 条
[41]   Comparison of the Biomechanical Properties of a Ventral Cervical Intervertebral Anchored Fusion Device With Locking Plate Fixation Applied to Cadaveric Canine Cervical Spines [J].
Schoellhorn, Bernhard ;
Buerki, Alexander ;
Stahl, Christina ;
Howard, Judith ;
Forterre, Franck .
VETERINARY SURGERY, 2013, 42 (07) :825-831
[42]   Locking Plate System Versus Standard Plate Fixation in the Management of Mandibular Fractures: Meta-Analysis of Randomized Controlled Trials [J].
Wusiman, Patiguli ;
Tuerxun, Julaiti ;
Yaolidaxi, Buhiliqiemu ;
Moming, Adili .
JOURNAL OF CRANIOFACIAL SURGERY, 2017, 28 (06) :1456-1461
[43]   Biomechanical Comparison of a Locking Compression Plate Combined With an Intramedullary Pin or a Polyetheretherketone Rod in a Cadaveric Canine Tibia Gap Model [J].
Beierer, Lucas H. ;
Glyde, Mark ;
Day, Robert E. ;
Hosgood, Giselle L. .
VETERINARY SURGERY, 2014, 43 (08) :1032-1038
[44]   Biomechanical Comparison of Superior Versus Anterior Plate Position for Fixation of Distal Clavicular Fractures: A New Model [J].
Wilkerson, James ;
Paryavi, Ebrahim ;
Kim, Hyunchul ;
Murthi, Anand ;
Pensy, Raymond A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (01) :E13-E17
[45]   Biomechanical study of a novel self-locking plate system for anterior cervical fixation [J].
Lao, Lifeng ;
Li, Qianyi ;
Zhong, Guibin ;
Song, Chao ;
Li, Yuanchao ;
Xu, Mingze ;
Liu, Zude .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2014, 9 :120
[46]   Fixation of distal fibular fractures: A biomechanical study of plate fixation techniques [J].
Marvan, Jiri ;
Horak, Zdenek ;
Vilimek, Miloslav ;
Horny, Lukas ;
Kachlik, David ;
Baca, Vaclav .
ACTA OF BIOENGINEERING AND BIOMECHANICS, 2017, 19 (01) :33-39
[47]   Biomechanical comparison of a lateral polyaxial locking plate with a posterolateral polyaxial locking plate applied to the distal fibula [J].
Hallbauer, Jakob ;
Klos, Kajetan ;
Rausch, Sascha ;
Graefenstein, Andreas ;
Wipf, Felix ;
Beimel, Claudia ;
Hofmann, Gunther ;
Mueckley, Thomas .
FOOT AND ANKLE SURGERY, 2014, 20 (03) :180-185
[48]   Fixation of Osteoporotic Fractures in the Upper Limb with a Locking Compression Plate [J].
Neuhaus, V ;
King, J. D. ;
Jupiter, J. B. .
ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2012, 79 (05) :404-410
[49]   Angulated locking plate in periprosthetic proximal femur fractures: biomechanical testing of a new prototype plate [J].
Mark Lenz ;
Boyko Gueorguiev ;
Steve Joseph ;
Bas van der Pol ;
Robert Geoff Richards ;
Markus Windolf ;
Karsten Schwieger ;
Piet de Boer .
Archives of Orthopaedic and Trauma Surgery, 2012, 132 :1437-1444
[50]   Supplemental Fixation of Supracondylar Distal Femur Fractures: A Biomechanical Comparison of Dual-Plate and Plate-Nail Constructs [J].
Wright, David J. ;
DeSanto, Donald J. ;
McGarry, Michelle H. ;
Lee, Thay Q. ;
Scolaro, John A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2020, 34 (08) :434-440