Global optimization of default phases for parallel transmit coils for ultra-high-field cardiac MRI

被引:6
作者
Terekhov, Maxim [1 ]
Elabyad, Ibrahim A. [1 ]
Schreiber, Laura M. [1 ]
机构
[1] Univ Hosp Wuerzburg, Comprehens Heart Failure Ctr, Chair Cellular & Mol Imaging, Wurzburg, Germany
关键词
TRANSCEIVER ARRAY; MAGNETIC-RESONANCE; DESIGN; PULSE;
D O I
10.1371/journal.pone.0255341
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The development of novel multiple-element transmit-receive arrays is an essential factor for improving B-1(+) field homogeneity in cardiac MRI at ultra-high magnetic field strength (B-0 > = 7.0T). One of the key steps in the design and fine-tuning of such arrays during the development process is finding the default driving phases for individual coil elements providing the best possible homogeneity of the combined B-1(+)-field that is achievable without (or before) subject-specific B-1(+)-adjustment in the scanner. This task is often solved by time-consuming (brute-force) or by limited efficiency optimization methods. In this work, we propose a robust technique to find phase vectors providing optimization of the B-1-homogeneity in the default setup of multiple-element transceiver arrays. The key point of the described method is the pre-selection of starting vectors for the iterative solver-based search to maximize the probability of finding a global extremum for a cost function optimizing the homogeneity of a shaped B-1(+)-field. This strategy allows for (i) drastic reduction of the computation time in comparison to a brute-force method and (ii) finding phase vectors providing a combined B-1(+)-field with homogeneity characteristics superior to the one provided by the random-multi-start optimization approach. The method was efficiently used for optimizing the default phase settings in the in-house-built 8Tx/16Rx arrays designed for cMRI in pigs at 7T.
引用
收藏
页数:20
相关论文
共 33 条
[1]   Three-dimensional static and dynamic parallel transmission of the human heart at 7 T [J].
Aigner, Christoph Stefan ;
Dietrich, Sebastian ;
Schmitter, Sebastian .
NMR IN BIOMEDICINE, 2021, 34 (03)
[2]   Rapid B1+ Mapping Using a Preconditioning RF Pulse with TurboFLASH Readout [J].
Chung, Sohae ;
Kim, Daniel ;
Breton, Elodie ;
Axel, Leon .
MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (02) :439-446
[3]   Saturated double-angle method for rapid B1 plus mapping [J].
Cunningham, Charles H. ;
Pauly, John M. ;
Nayak, Krishna S. .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (06) :1326-1333
[4]   Design and Application of a Four-Channel Transmit/Receive Surface Coil for Functional Cardiac Imaging at 7T [J].
Dieringer, Matthias A. ;
Renz, Wolfgang ;
Lindel, Tomasz ;
Seifert, Frank ;
Frauenrath, Tobias ;
von Knobelsdorff-Brenkenhoff, Florian ;
Waiczies, Helmar ;
Hoffmann, Werner ;
Rieger, Jan ;
Pfeiffer, Harald ;
Ittermann, Bernd ;
Schulz-Menger, Jeanette ;
Niendorf, Thoralf .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (03) :736-741
[5]   3D Free-breathing multichannel absolute B1+Mapping in the human body at 7T [J].
Dietrich, Sebastian ;
Aigner, Christoph S. ;
Kolbitsch, Christoph ;
Mayer, Johannes ;
Ludwig, Juliane ;
Schmidt, Simon ;
Schaeffter, Tobias ;
Schmitter, Sebastian .
MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (05) :2552-2567
[6]   Design and Implementation of Two 16-Element Antisymmetric Transceiver Coil Arrays for Parallel Transmission Human Cardiac MRI at 7 T [J].
Elabyad, Ibrahim A. ;
Terekhov, Maxim ;
Bille, Maya ;
Schreiber, Laura M. .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2021, 69 (07) :3540-3557
[7]   A Novel Mono-surface Antisymmetric 8Tx/16Rx Coil Array for Parallel Transmit Cardiac MRI in Pigs at 7T [J].
Elabyad, Ibrahim A. ;
Terekhov, Maxim ;
Lohr, David ;
Stefanescu, Maria R. ;
Baltes, Steffen ;
Schreiber, Laura M. .
SCIENTIFIC REPORTS, 2020, 10 (01)
[8]   Design of a novel antisymmetric coil array for parallel transmit cardiac MRI in pigs at 7 T [J].
Elabyad, Ibrahim A. ;
Terekhov, M. ;
Stefanescu, M. R. ;
Lohr, D. ;
Fischer, M. ;
Schreiber, L. M. .
JOURNAL OF MAGNETIC RESONANCE, 2019, 305 :195-208
[9]  
Erturk M Arcan, 2019, Top Magn Reson Imaging, V28, P101, DOI 10.1097/RMR.0000000000000202
[10]   A 16-channel combined loop-dipole transceiver array for 7 Tesla body MRI [J].
Erturk, M. Arcan ;
Raaijmakers, Alexander J. E. ;
Adriany, Gregor ;
Ugurbil, Kamil ;
Metzger, Gregory J. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (02) :884-894