Adaptive R-Peak Detector in Extreme Noise Using EMD Selective Analyzer

被引:0
作者
Abderahman, Huthaifa N. [1 ]
Dajani, Hilmi R. [1 ]
Groza, Voicu Z. [1 ]
机构
[1] Univ Ottawa, Sch Elect Engn & Comp Sci, Ottawa, ON, Canada
来源
2022 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA 2022) | 2022年
关键词
ECG; R-peak detection; Very low S/N; Single-arm electrodes; Motion artifact; EMD; IMF; Hausdorff Distance; BASE-LINE WANDER; ECG SIGNAL; MOTION ARTIFACT; ALGORITHM; DIFFERENCE; REMOVAL;
D O I
10.1109/MEMEA54994.2022.9856531
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accurately detecting R-peaks in electrocardiogram (ECG) signals is important in various health monitoring applications, such as cuffless blood pressure measurements. In some cases, such as with single-arm ECG measurement, the signal may be buried in large amounts of noise. Moreover, many of the current existing ECG monitors pause the reading until the noise conditions are better, which may lead to a large loss in data. In this work, an adaptive approach is introduced, based on Empirical Mode Decomposition (EMD), to accurately detect the R-peaks in an extremely noisy ECG signal obtained using a single-arm measurement. The proposed algorithm starts by examining the Intrinsic Mode Functions (IMFs) extracted from the recorded signal, using the Hausdorff Distance (HD) as a selection tool, before applying the peak detection algorithm. Experimental measurements used to evaluate the algorithm were obtained from 10 healthy subjects for a total of 30 noisy ECG records, containing close to 2000 QRS complexes in total, the majority of which were at an estimated S/N (Signal-to-noise ratio) below -10 dB. The obtained results show a promising technique for detecting R-peaks in extreme noise with a low percentage of detection error ratio and an average percentage error in R-R interval estimation of 6.8%.
引用
收藏
页数:6
相关论文
共 46 条
[1]   An integrated blood pressure measurement system for suppression of motion artifacts [J].
Abderahman, Huthaifa N. ;
Dajani, Hilmi R. ;
Bolic, Miodrag ;
Groza, Voicu Z. .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2017, 145 :1-10
[2]   Fractal and EMD based removal of baseline wander and powerline interference from ECG signals [J].
Agrawal, Sakshi ;
Gupta, Anubha .
COMPUTERS IN BIOLOGY AND MEDICINE, 2013, 43 (11) :1889-1899
[3]   Electrocardiogram-Assisted Blood Pressure Estimation [J].
Ahmad, Saif ;
Chen, Silu ;
Soueidan, Karen ;
Batkin, Izmail ;
Bolic, Miodrag ;
Dajani, Hilmi ;
Groza, Voicu .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (03) :608-618
[4]  
Anuradha K., 2016, Advances in Natural and Applied Sciences, V10, P83
[5]   A new QRS detection algorithm based on the Hilbert transform [J].
Benitez, DS ;
Gaydecki, PA ;
Zaidi, A ;
Fitzpatrick, AP .
COMPUTERS IN CARDIOLOGY 2000, VOL 27, 2000, 27 :379-382
[6]   Motion Artifact Removal in Ambulatory ECG Signal for Heart Rate Variability Analysis [J].
Berwal, Deepak ;
Vandana, C. R. ;
Dewan, Sourya ;
Jiji, C., V ;
Baghini, Maryam Shojaei .
IEEE SENSORS JOURNAL, 2019, 19 (24) :12432-12442
[7]   ECG signal denoising and baseline wander correction based on the empirical mode decomposition [J].
Blanco-Velasco, Manuel ;
Weng, Binwei ;
Barner, Kenneth E. .
COMPUTERS IN BIOLOGY AND MEDICINE, 2008, 38 (01) :1-13
[8]   Multiresolution wavelet-based QRS complex detection algorithm suited to several abnormal morphologies [J].
Bouaziz, Fatiha ;
Boutana, Daoud ;
Benidir, Messaoud .
IET SIGNAL PROCESSING, 2014, 8 (07) :774-782
[9]   Fast QRS Detection and ECG Compression Based on Signal Structural Analysis [J].
Burguera, Antoni .
IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2019, 23 (01) :123-131
[10]  
Chen Z., 2018, IEEE T IND INFORM, P1, DOI DOI 10.1109/TII.2018.2869843