The validity and accuracy in foot-to-foot bioelectrical impedance analysis measuring models referenced by dual-energy X-ray absorptiometry in body composition in standing position

被引:0
作者
Hsieh, Kuen-Chang [2 ]
Lu, Hsueh-Kuan [3 ]
Chen, Chun-Hao [4 ]
Jang, Tsong-Rong [5 ,6 ]
Chen, Yu-Yawn [1 ]
Kao, Ming-Feng [1 ]
机构
[1] Natl Taiwan Coll Phys Educ, Dept Phys Educ, Taipei, Taiwan
[2] Charder Elect Co Ltd, Res Ctr, Taipei, Taiwan
[3] Natl Taiwan Coll Phys Educ, Sport Sci Res Ctr, Taipei, Taiwan
[4] Tunghai Univ, Phys Educ Off, Taichung, Taiwan
[5] Natl Taiwan Coll Phys Educ, Athlet Dept, Taipei, Taiwan
[6] Natl Taiwan Coll Phys Educ, Grad Sch, Taipei, Taiwan
关键词
Hand-to-foot bioimpedance analysis; foot-to-foot bioimpednace analysis; relative body fat (BF); fat free mass (FFM); dual-energy X-ray absorptiometry (DEXA); OBESE WOMEN; VALIDATION; LEG; FAT; RESISTANCE; DEVICES; SYSTEM;
D O I
暂无
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The criterion validity of foot-to-foot (Z(F-F)) by bioelectrical impedance analysis (BIA) in a standing position measuring models was referenced by dual-energy X-ray absorptiometry (DEXA) and hand-to-foot (Z(H-F)). In order to cross match analysis, both of the Z(H-F) and Z(F-F) measuring by BIA were performed in the same event to the identical subject, simultaneously. 105 males (7 to 70 years old) and 108 females (7 to 67 years old) were measured by BIA and DEXA. The Z(F-F) value was 481.85 +/- 63.37 ohm in male and 554.35 +/- 74.97 ohm in female. The Z(H-F) value was 586.82 +/- 81.56 ohm in male and 703.28 +/- 97.70 ohm in female. Regression equations for Z(F-F) and Z(H-F) were Z(F-F) = 0.926 Z(H-F) - 63.093, (R = 0.85) in male and Z(F-F) = 0.909 Z(H-F) - 86.673 (R = 0.86) in female. The fat free mass (FFM), measured by DEXA was 49.42 +/- 10.03 kg in male and 34.52 +/- 4.45 kg in female. The correlation (R value) for Z(F-F) to DEXA was 0.93 and for Z(H-F) was 0.96 in male, and 0.85 and 0.87 in female, respectively. The high correlation between Z(H-F) and Z(F-F) vs. whole body DEXA in both male and female renders it applicable to develop the clinical instrument by foot-to-foot measuring models by BIA in standing position.
引用
收藏
页码:3222 / 3231
页数:10
相关论文
共 30 条
[1]  
BLAND JM, 1998, LANCET, V8, P307
[2]  
Bolanowski M, 2001, Med Sci Monit, V7, P1029
[3]   Body composition analysis by leg-to-leg bioelectrical impedance and dual-energy X-ray absorptiometry in non-obese and obese individuals [J].
Boneva-Asiova, Z. ;
Boyanov, M. A. .
DIABETES OBESITY & METABOLISM, 2008, 10 (11) :1012-1018
[4]   Accuracy of Bioelectrical Impedance Consumer Devices for Measurement of Body Composition in Comparison to Whole Body Magnetic Resonance Imaging and Dual X-Ray Absorptiometry [J].
Bosy-Westphal, Anja ;
Later, Wiebke ;
Hitze, Britta ;
Sato, Tetsuya ;
Kossel, Elke ;
Glueer, Claus-Christian ;
Heller, Martin ;
Mueller, Manfred James .
OBESITY FACTS, 2008, 1 (06) :319-324
[5]  
Burger HC, 1943, ACTA MED SCAND, V114, P584
[6]  
GRAVES J E, 1989, American Journal of Human Biology, V1, P603, DOI 10.1002/ajhb.1310010511
[7]  
HORTON JW, 1935, J FRANKLIN I, V20, P557
[8]   Body composition profiles of elite American heptathletes [J].
Houtkooper, LB ;
Mullins, VA ;
Going, SB ;
Brown, CH ;
Lohman, TG .
INTERNATIONAL JOURNAL OF SPORT NUTRITION AND EXERCISE METABOLISM, 2001, 11 (02) :162-173
[9]  
JANSSEN I, 2000, AM J CLIN NUTR, V64, pS423
[10]   Bioelectrical impedance analysis principles and methods [J].
Kyle, UG ;
Bosaeus, I ;
De Lorenzo, AD ;
Deurenberg, P ;
Elia, M ;
Gómez, JM ;
Heitmann, BL ;
Kent-Smith, L ;
Melchior, JC ;
Pirlich, M ;
Scharfetter, H ;
Schols, AMWJ ;
Pichard, C .
CLINICAL NUTRITION, 2004, 23 (05) :1226-1243