AN ADAPTIVE SPECTRAL ESTIMATION TECHNIQUE TO DETECT CAVITATION IN HIFU WITH HIGH SPATIAL RESOLUTION

被引:7
作者
Hsieh, Chang-yu [1 ]
Smith, Penny Probert [1 ]
Mayia, Fares [2 ]
Ye, Guoliang [1 ]
机构
[1] Univ Oxford, Inst Biomed Engn, Oxford OX1 2JD, England
[2] Oxford Radcliffe Trust, Dept Med Phys, Oxford, England
基金
英国工程与自然科学研究理事会;
关键词
HIFU; Ultrasound signal processing; Image processing; INTENSITY FOCUSED ULTRASOUND; NONLINEAR PROPAGATION; LESION FORMATION; SELECTION; ABLATION; SURGERY; BUBBLES; CANCER;
D O I
10.1016/j.ultrasmedbio.2011.04.011
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In ultrasound-guided high-intensity focused ultrasound (HIFU) therapy, the changes observed on tissue are subtle during treatment; some ultrasound-guided HIFU protocols rely on the observation of significant brightness changes as the indicator of tissue lesions. The occurrence of a distinct hyperechogenic region ("bright-up") around the focus is often associated with acoustic cavitation resulting in microbubble formation, but it may indicate different physical events such as larger bubbles from boiling (known to alter acoustic impedance) or sometimes lesion formation. A reliable method to distinguish and spatially localize these causes within the tissue would assist the control of HIFU delivery, which is the subject of this paper. Spectral analysis of the radio frequency (RF) signal underlying the B-mode image provides more information on the physical cause, but the usual techniques that are methods on the Fourier transform require a long series for good spectral resolution and so they give poor spatial resolution. This paper introduces an active spectral cavitation detection method to attain high spatial resolution (0.15 x 0.15 mm per pixel) through a parametric statistical method (ARMA modeling) used on finite-length data sets, which enables local changes to be identified more easily. This technique uses the characteristics of the signal itself to optimize the model parameters and structure. Its performance is assessed using synthesized cavitation RF data, and it is then demonstrated in ex vivo bovine liver during and after HIFU exposure. The results suggest that good spatial and spectral resolution can be obtained by the design of suitable algorithms. In ultrasound-guided HIFU, the technique provides a useful supplement to B-mode analysis, with no additional time penalty in data acquisition. (E-mail: skyevom3@yahoo.com) (C) 2011 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1134 / 1150
页数:17
相关论文
共 42 条
[1]   NEW LOOK AT STATISTICAL-MODEL IDENTIFICATION [J].
AKAIKE, H .
IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1974, AC19 (06) :716-723
[2]   STATISTICAL PREDICTOR IDENTIFICATION [J].
AKAIKE, H .
ANNALS OF THE INSTITUTE OF STATISTICAL MATHEMATICS, 1970, 22 (02) :203-&
[3]  
[Anonymous], 1960, ECONOMETRICA
[4]   Use of overpressure to assess the role of bubbles in focused ultrasound lesion shape in vitro [J].
Bailey, MR ;
Couret, LN ;
Sapozhnikov, OA ;
Khokhlova, VA ;
Ter Haar, G ;
Vaezy, S ;
Shi, XG ;
Martin, R ;
Crum, LA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (05) :695-708
[5]   Quantitative MR temperature monitoring of high-intensity focused ultrasound therapy [J].
Bohris, C ;
Schreiber, WG ;
Jenne, J ;
Simiantonakis, I ;
Rastert, R ;
Zabel, HJ ;
Huber, P ;
Bader, R ;
Brix, G .
MAGNETIC RESONANCE IMAGING, 1999, 17 (04) :603-610
[6]  
Bracewell R.N., 1986, FOURIER TRANSFORM IT, V2nd
[7]  
Broersen P.M., 2006, Automatic Autocorrelation and Spectral Analysis
[8]   ON FINITE-SAMPLE THEORY FOR AUTOREGRESSIVE MODEL ORDER SELECTION [J].
BROERSEN, PMT ;
WENSINK, HE .
IEEE TRANSACTIONS ON SIGNAL PROCESSING, 1993, 41 (01) :194-204
[9]   The quality of models for ARMA processes [J].
Broersen, PMT .
IEEE TRANSACTIONS ON SIGNAL PROCESSING, 1998, 46 (06) :1749-1752
[10]   Automatic spectral analysis with time series models [J].
Broersen, PMT .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2002, 51 (02) :211-216