Validity of auscultatory and Penaz blood pressure measurements during profound heat stress alone and with an orthostatic challenge

被引:16
作者
Ganio, Matthew S.
Brothers, R. Matthew
Lucas, Rebekah A. I.
Hastings, Jeffrey L.
Crandall, Craig G. [1 ]
机构
[1] Texas Hlth Presbyterian Hosp Dallas, Inst Exercise & Environm Med, Dallas, TX 75231 USA
关键词
lower body negative pressure; passive heat stress; arterial blood pressure; STATISTICAL-METHODS; FINGER PRESSURE; TOLERANCE; SKIN; EXERCISE; VALIDATION; FINOMETER; FINAPRES; VELOCITY; DEVICE;
D O I
10.1152/ajpregu.00247.2011
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Ganio MS, Brothers RM, Lucas RAI, Hastings JL, Crandall CG. Validity of auscultatory and Penaz blood pressure measurements during profound heat stress alone and with an orthostatic challenge. Am J Physiol Regul Integr Comp Physiol 301: R1510-R1516, 2011. First published August 10, 2011; doi:10.1152/ajpregu.00247.2011.Despite frequent reporting of blood pressure (BP) during profound passive heat stress, both with and without a hypotensive challenge, the method by which BP is measured often varies between laboratories. It is unknown whether auscultatory and finger BP measures accurately reflect intra-arterial BP during dynamic changes in cardiac output and peripheral resistance associated with the aforementioned conditions. The purpose of this investigation was to test the hypothesis that auscultatory BP measured at the brachial artery, and finger BP measured by the Penaz method, are valid measures of intra-arterial BP during a passive heat stress and a heat-stressed orthostatic challenge, via lower body negative pressure (LBNP). Absolute (specific aim 1) and the change in (specific aim 2) systolic (SBP), diastolic (DBP), and mean BPs (MBP) were compared at normothermia, after a core temperature increase of 1.47 +/- 0.09 degrees C, and during subsequent LBNP. Heat stress did not change auscultatory SBP (6 +/- 11 mmHg; P = 0.16), but Penaz SBP (-22 +/- 16 mmHg; P < 0.001) and intra-arterial SBP (-11 +/- 13 mmHg P = 0.017) decreased. In contrast, DBP and MBP did not differ between methods throughout heat stress. Compared with BP before LBNP, the magnitude of the reduction in BP with all three methods was similar throughout LBNP (P > 0.05). In conclusion, auscultatory SBP and Penaz SBP failed to track the decrease in intra-arterial SBP that occurred during the profound heat stress, while decreases in arterial BP during an orthostatic challenge are comparable between methodologies.
引用
收藏
页码:R1510 / R1516
页数:7
相关论文
共 36 条
[1]  
Association for the Advancement of Medical Instrumentation, 1993, ANSI AAMI SP, P40
[2]   Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine [J].
Atkinson, G ;
Nevill, AM .
SPORTS MEDICINE, 1998, 26 (04) :217-238
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   EFFECT OF REGIONAL AND SYSTEMIC CHANGES IN VASOMOTOR TONE ON FINGER PRESSURE AMPLIFICATION [J].
BOS, WJW ;
VANDENMEIRACKER, AH ;
WESSELING, KH ;
SCHALEKAMP, MADH .
HYPERTENSION, 1995, 26 (02) :315-320
[5]   Methodological assessment of skin and limb blood flows in the human forearm during thermal and baroreceptor provocations [J].
Brothers, R. Matthew ;
Wingo, Jonathan E. ;
Hubing, Kimberly A. ;
Crandall, Craig G. .
JOURNAL OF APPLIED PHYSIOLOGY, 2010, 109 (03) :895-900
[6]   Skin surface cooling improves orthostatic tolerance in normothermic individuals [J].
Durand, S ;
Cui, J ;
Williams, KD ;
Crandall, CG .
AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 2004, 286 (01) :R199-R205
[7]   Consistency of blood pressure differences between the left and right arms [J].
Eguchi, Kazuo ;
Yacoub, Mona ;
Jhalani, Juhee ;
Gerin, William ;
Schwartz, Joseph E. ;
Pickering, Thomas G. .
ARCHIVES OF INTERNAL MEDICINE, 2007, 167 (04) :388-393
[8]   Application of acute maximal exercise to protect orthostatic tolerance after simulated microgravity [J].
Engelke, KA ;
Doerr, DF ;
Convertino, VA .
AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 1996, 271 (04) :R837-R847
[9]   Blood Pressure Measurement During Aerobic Exercise: Subsidies for Cardiac Rehabilitation [J].
Furtado, Emanuel Couto ;
Ramos, Plinio dos Santos ;
Soares de Araujo, Claudio Gil .
ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2009, 93 (01) :45-52
[10]   AN AUTOMATED-SYSTEM FOR BLOOD-PRESSURE DETERMINATION DURING EXERCISE [J].
GLASSER, SP ;
RAMSEY, MR .
CIRCULATION, 1981, 63 (02) :348-353