Combining SPECT medical imaging and computational fluid dynamics for analyzing blood and dialysate flow in hemodialyzers

被引:14
作者
Eloot, S
D'asseler, Y
De Bondt, P
Verdonck, P
机构
[1] Univ Ghent, Hydraul Lab, Inst Biomed Technol, B-9000 Ghent, Belgium
[2] Univ Ghent, Dept Elect & Informat Syst, Inst Biomed Technol, B-9000 Ghent, Belgium
[3] Ghent Univ Hosp, Dept Nucl Med, Inst Biomed Technol, B-9000 Ghent, Belgium
关键词
flow distribution; dialyzer; medical imaging; numerical model; CFD;
D O I
10.1177/039139880502800713
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For a better insight in dialyzer efficiency with respect to local mass transport in a low flux dialyzer (Fresenius F6HPS), blood and dialysate flow distributions were visualized with computational fluid dynamic (CFD) simulations, which were validated with single photon emission computed tomography (SPECT) imaging. To visualize blood-side flow while avoiding transport through the fiber membrane, a bolus of 99m-Technetium labeled MAA (Macro Aggregated Albumin) was injected in the flow using an electronic valve. Water was used to simulate blood, but flow rate was adjusted according to laws of dynamic similarity to account for the viscosity difference (factor 2.75). For the visualization of dialysate flow, a bolus of 99m-Technetium labeled DMSA (Dimercaptosuccinic Acid) was injected, while pressurized air in the blood compartment avoided transmembrane flow. For each test series, 3D acquisitions were made on a two respectively three-headed SPECT camera. By evaluating the images at different time steps, dynamic 3D intensity plots were obtained, which were further used to derive local flow velocities. Additionally, three-dimensional CFD models were developed for simulating the overall blood and dialysate flow, respectively. In both models, the whole fiber compartment was defined as a porous medium with overall axial and radial permeability derived theoretically and from in vitro tests. With the imaging as well as with the computational technique, a homogeneous blood flow distribution was found, while vortices and fluid stagnation were observed in the dialyzer inlet manifold. The non-homogeneous dialysate distribution, as found with SPECT imaging, implies the occurrence of non-efficient sites with respect to mass transfer. The discrepancy between the dialysate results of both techniques indicated that the assumption of a constant fiber bundle permeability in the CFD model was too optimistic. In conclusion, medical imaging techniques like SPECT are very helpful to validate CFD models, which can be further applied for dialyzer design and optimization.
引用
收藏
页码:739 / 749
页数:11
相关论文
共 20 条
[1]   In vitro evaluation of the hydraulic permeability of polysulfone dialysers [J].
Eloot, S ;
De Wachter, D ;
Vienken, J ;
Pohlmeier, R ;
Verdonck, P .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2002, 25 (03) :210-216
[2]   Computational flow modeling in hollow-fiber dialyzers [J].
Eloot, S ;
De Wachter, D ;
Van Tricht, I ;
Verdonck, P .
ARTIFICIAL ORGANS, 2002, 26 (07) :590-599
[3]   Effects of novel manufacturing technology on blood and dialysate flow distribution in a new low flux "α polysulfone" hemodialyzer [J].
Gastaldon, F ;
Brendolan, A ;
Crepaldi, C ;
Frisone, P ;
Zamboni, S ;
D'Intini, V ;
Poulin, S ;
Hector, R ;
Granziero, A ;
Martins, K ;
Gellert, R ;
Inguaggiato, P ;
Ronco, C .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (02) :105-112
[4]   Hollow fiber shape alters solute clearances in high flux hemodialyzers [J].
Leypoldt, JK ;
Cheung, AK ;
Chirananthavat, T ;
Gilson, JF ;
Kamerath, CD ;
Deeter, RB .
ASAIO JOURNAL, 2003, 49 (01) :81-87
[5]   Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates [J].
Leypoldt, JK ;
Cheung, AK ;
Agodoa, LY ;
Daugirdas, JT ;
Greene, T ;
Keshaviah, PR .
KIDNEY INTERNATIONAL, 1997, 51 (06) :2013-2017
[6]  
Mockros L F, 1985, Trans Am Soc Artif Intern Organs, V31, P628
[7]   Dialysate pressure isobars in a hollow-fiber dialyzer determined from magnetic resonance imaging and numerical simulation of dialysate flow [J].
Osuga, T ;
Obata, T ;
Ikehira, H ;
Tanada, S ;
Sasaki, Y ;
Naito, H .
ARTIFICIAL ORGANS, 1998, 22 (10) :907-909
[8]   Increasing dialysate flow rate increases dialyzer urea mass transfer-area coefficients during clinical use [J].
Ouseph, R ;
Ward, RA .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 37 (02) :316-320
[9]   Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers: A nonintrusive experimental study using MRI [J].
Poh, CK ;
Hardy, PA ;
Liao, ZJ ;
Huang, ZP ;
Clark, WR ;
Gao, DY .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (04) :481-489
[10]   Effect of spacer yarns on the dialysate flow distribution of hemodialyzers: A magnetic resonance imaging study [J].
Poh, CK ;
Hardy, PA ;
Liao, ZJ ;
Huang, ZP ;
Clark, WR ;
Gao, DY .
ASAIO JOURNAL, 2003, 49 (04) :440-448