Wireless and Batteryless Blood Pressure Sensor

被引:0
作者
Sosa, J. [1 ]
Garcia-Alonso, S. [1 ]
Monzon-Verona, J. M.
Montiel-Nelson, J. A. [1 ]
Beriain, A. [2 ]
Fernandez, E. [2 ]
Solar, H. [2 ]
Garcia-Alonso, A. [2 ]
Gutierrez, I. [2 ]
机构
[1] Univ Las Palmas Gran Canaria, Dept Elect Engn & Automat, E-35017 Las Palmas Gran Canaria, Spain
[2] Univ Navarra, CElT, COMMIC, San Sebastian, Spain
来源
2011 IEEE 54TH INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS) | 2011年
关键词
OPTIMIZATION; MICROSYSTEM; MEMS;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A low-power invasive blood pressure transducer has been designed. It is based on a structure of four high resistive strain gauges in a full Wheatstone bridge configuration. For obtaining the required sensitivity, the gauge was allocated performing a careful mechanical simulation of the membrane behaviour. The electronic conditioning circuit and the ADC were optimized to compensate the noise introduced by the high resistive gauges. A batteryless RFID tag designed to incorporate and to supply the sensor was fabricated and measured. It power supplies the sensor at a distance longer than 50 cm from a standard reader, while the reported ranges -to the best of our knowledge- are shorter than 10 cm. This wireless and batteryless sensor is especially suitable for surgery rooms and intensive care units.
引用
收藏
页数:4
相关论文
共 20 条
  • [11] Kester W., 2005, DATA CONVERSION HDB
  • [12] OPTIMIZATION OF WHEATSTONE BRIDGE SENSITIVITY
    MAISEL, JE
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1977, 26 (01) : 17 - 21
  • [13] MEMS in medicine and biology
    Panescu, Dorin
    [J]. IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2006, 25 (05): : 19 - 28
  • [14] Pardo D, 2007, 2007 IEEE INTERNATIONAL CONFERENCE ON RFID, P116
  • [15] IC PIEZORESISTIVE PRESSURE SENSOR FOR BIOMEDICAL INSTRUMENTATION
    SAMAUN
    WISE, KD
    ANGELL, JB
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) : 101 - 109
  • [16] TAKAGISHI E, 1978, IEEE T INSTRUM MEAS, V27, P304
  • [17] Full Passive UHF Tag With a Temperature Sensor Suitable for Human Body Temperature Monitoring
    Vaz, A.
    Ubarretxena, A.
    Zalbide, I.
    Pardo, D.
    Solar, H.
    Garcia-Alonso, A.
    Berenguer, R.
    [J]. IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS, 2010, 57 (02) : 95 - 99
  • [18] Analog-to-digital converter survey and analysis
    Walden, RH
    [J]. IEEE JOURNAL ON SELECTED AREAS IN COMMUNICATIONS, 1999, 17 (04) : 539 - 550
  • [19] Security and privacy in RFID and applications in telemedicine
    Xiao, Y
    Shen, XM
    Sun, B
    Cai, L
    [J]. IEEE COMMUNICATIONS MAGAZINE, 2006, 44 (04) : 64 - 72
  • [20] Power and energy optimization of the digital core of a Gen2 long range full passive RFID sensor tag
    Zalbide, I.
    Vicario, J.
    Velez, I.