Comparative biomechanical analysis of rigid vs. flexible fixation devices for the lumbar spine: A geometrically patient-specific poroelastic finite element study

被引:13
作者
Khalaf, Kinda [1 ,2 ]
Nikkhoo, Mohammad [3 ]
机构
[1] Khalifa Univ Sci & Technol, Dept Biomed Engn, POB 127788, Abu Dhabi, U Arab Emirates
[2] Hlth Engn Innovat Ctr, POB 127788, Abu Dhabi, U Arab Emirates
[3] Islamic Azad Univ, Dept Biomed Engn, Sci & Res Branch, Tehran, Iran
关键词
Poroelastic modeling; Finite element analysis; Patient-specific; Lumbar fixation device; Dynamic rod; Interspinous process devices; INTERBODY FUSION; MOTION SEGMENT; INTERVERTEBRAL DISCS; INTRADISCAL PRESSURE; LOAD-BEARING; ADJACENT; MODEL; STABILIZATION; DEGENERATION; VALIDATION;
D O I
10.1016/j.cmpb.2021.106481
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Background and Objective: Lumbar spinal stenosis (LSS), or the narrowing of the spinal canal, continues to be the leading preoperative diagnosis for adults older than 65 years who undergo spine surgery. Although the treatment of LSS depends on its severity, the optimal surgical technique towards decreasing the risk of adjacent segment disease (ASD) remains elusive. This study aimed to comparatively analyze spinal biomechanics with rigid and flexible fixation devices (i.e., rigid and dynamic posterolateral fusion (PLF) and interspinous process (ISP) devices) during daily activities. Methods: Using a validated parametric poroelastic finite element modeling approach, 8 subject-specific pre-operative models were developed, and their validity was evaluated. Parametric FE models of the lumbar spines were then regenerated based on post-operation images for (A) rigid PLF (B) dynamic PLF (C) rigid ISP device (Coflex) and (D) flexible ISP device (DIAM) at L4-L5 level. Biomechanical responses for instrumented and adjacent intervertebral discs (IVDs) were analyzed and compared subject to static and cyclic loading. Results: The preoperative models were well comparable with previous works in literature. The postoperative results for the PLF and Coflex rigid systems, demonstrated greater ROM; higher values of stress and strain in the AF region; and increased disc height and fluid loss at the adjacent levels, as compared with the pre-op models and the post-op results of the flexible systems (i.e., dynamic PLF and DIAM). The calculated forces on the facet joint were of smaller magnitude for the ISP devices as compared to the PLF, particularly during extension. Conclusions: This study demonstrates that the dynamic PLF construct and DIAM implants could be effective to maintain the natural poroelastic characteristics of adjacent IVDs, which could be beneficial for enhancing long-term clinical outcomes. FEM provides clinicians with an invaluable patient-specific quantitative tool for informed surgical planning and discerning follow-up management. (c) 2021 Published by Elsevier B.V.
引用
收藏
页数:12
相关论文
共 61 条
  • [1] Poroelastic creep response analysis of a lumbar motion segment in compression
    Argoubi, M
    ShiraziAdl, A
    [J]. JOURNAL OF BIOMECHANICS, 1996, 29 (10) : 1331 - 1339
  • [2] We Need to Talk about Lumbar Total Disc Replacement
    Beatty, Stephen
    [J]. INTERNATIONAL JOURNAL OF SPINE SURGERY, 2018, 12 (02) : 201 - 240
  • [3] Biomechanical In Vitro Test of a Novel Dynamic Spinal Stabilization System Incorporating Polycarbonate Urethane Material Under Physiological Conditions
    Beckmann, Agnes
    Nicolini, Luis Fernando
    Grevenstein, David
    Backes, Hermann
    Oikonomidis, Stavros
    Sobottke, Rolf
    Kobbe, Philipp
    Hildebrand, Frank
    Stoffel, Marcus
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2020, 142 (01):
  • [4] Dynamic compared to rigid fixation in lumbar spine: a systematic review
    Botelho, Ricardo Vieira
    Bastianello Junior, Rafael
    de Albuquerque, Luciana DiniGianini
    Bernardo, Wanderley Marques
    [J]. REVISTA DA ASSOCIACAO MEDICA BRASILEIRA, 2014, 60 (02): : 151 - 155
  • [5] Numerical implementation of an osmo-poro-visco-hyperelastic finite element solver: application to the intervertebral disc
    Castro, A. P. G.
    Alves, J. L.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (05) : 538 - 550
  • [6] Chou Dean, 2011, Evid Based Spine Care J, V2, P33, DOI 10.1055/s-0030-1267111
  • [7] Chou WK, 2015, J SPINAL DISORD TECH, V28, pE121, DOI 10.1097/BSD.0000000000000176
  • [8] Comparison of eight published static finite element models of the intact lumbar spine: Predictive power of models improves when combined together
    Dreischarf, M.
    Zander, T.
    Shirazi-Adl, A.
    Puttlitz, C. M.
    Adam, C. J.
    Chen, C. S.
    Goel, V. K.
    Kiapour, A.
    Kim, Y. H.
    Labus, K. M.
    Little, J. P.
    Park, W. M.
    Wang, Y. H.
    Wilke, H. J.
    Rohlmann, A.
    Schmidt, H.
    [J]. JOURNAL OF BIOMECHANICS, 2014, 47 (08) : 1757 - 1766
  • [10] Finite element investigation of the loading rate effect on the spinal load-sharing changes under impact conditions
    El-Rich, Marwan
    Arnoux, Pierre-Jean
    Wagnac, Eric
    Brunet, Christian
    Aubin, Carl-Eric
    [J]. JOURNAL OF BIOMECHANICS, 2009, 42 (09) : 1252 - 1262