A convenient method to verify the accuracy of oscillometric blood pressure monitors by the auscultatory method: A smartphone-based app

被引:8
作者
Zhang, Zhi [1 ]
Xi, Weichun [1 ]
Wang, Bingjiang [2 ]
Chu, Guang [1 ]
Wang, Fang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Gen Hosp, Sch Med, Dept Cardiol, Shanghai, Peoples R China
[2] Nanjing Med Univ, Huaian Peoples Hosp 1, Dept Cardiol, Huaian, Peoples R China
基金
中国国家自然科学基金;
关键词
auscultatory; blood pressure measurement; blood pressure monitor; hypertension; oscillometric sphygmomanometer; HYPERTENSION; VALIDATION; ERROR;
D O I
10.1111/jch.13460
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
It is recommended that oscillometric devices be calibrated by auscultation when first used, but this is difficult in practice. Here, we introduce a smartphone-based technique to verify the accuracy of blood pressure monitors (BPMs). We enrolled 99 consecutive subjects and tested 6 brands of BPMs in this study. During measurements of electronic oscillometric BPMs, Korotkoff sounds were simultaneously collected using a stethoscope head beneath a cuff connected to a smartphone, and an app named Accutension Stetho could then yield an auscultatory BP reading as a reference. Next, differences in BP between the different BPMs and Accutension Stetho were determined. The percentage of BP differences falling within 5, 10, and 15 mm Hg; the mean (MD) value; and the standard deviation (SD) of BP differences and deflation errors were analyzed among all the BPMs. We found that the percentages of SBP differences falling within 5 mm Hg of the 6 BPMs were 80%, 79%, 77%, 72%, 68%, and 63%, in turn. The deflation rates among the 6 BPMs were 2.23, 3.48, 6.10, 2.44, 3.66, and 4.85 mm Hg/beat, respectively. Deflation errors, which were defined as deflation prior to the end of the Korotkoff sounds, existed in 4 BPMs. In conclusion, Accutension Stetho could detect BP differences between oscillometric BPM readings and simultaneous auscultatory readings. Diastolic BP was overestimated when the device deflated prior to the end of the Korotkoff sounds. Using the app, it is possible to evaluate the accuracy of BPMs among the same subjects.
引用
收藏
页码:173 / 180
页数:8
相关论文
共 18 条
[1]   The Accutension Stetho, an automated auscultatory device to validate automated sphygmomanometer readings in individual patients [J].
Alpert, Bruce S. .
JOURNAL OF HUMAN HYPERTENSION, 2018, 32 (06) :455-459
[2]   Automatic blood pressure measurement: the oscillometric waveform shape is a potential contributor to differences between oscillometric and auscultatory pressure measurements [J].
Amoore, John N. ;
Lemesre, Yann ;
Murray, Ian C. ;
Mieke, Stephan ;
King, Susan T. ;
Smith, Fiona E. ;
Murray, Alan .
JOURNAL OF HYPERTENSION, 2008, 26 (01) :35-43
[3]   From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: does easiness come with reliability? [J].
Benmira, A. ;
Perez-Martin, A. ;
Schuster, I. ;
Aichoun, I. ;
Coudray, S. ;
Bereksi-Reguig, F. ;
Dauzat, M. .
EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (02) :179-189
[4]   Validation of a smartphone auscultatory blood pressure kit Accutension XYZ-110 in adults according to the ANSI/AAMI/ISO 81060-2:2013 standard [J].
Chu, Guang ;
Zhang, Zhi ;
Xu, Mengdan ;
Huang, Daini ;
Dai, Qiuyan .
BLOOD PRESSURE MONITORING, 2017, 22 (05) :290-294
[5]   Reliability of Resting Blood Pressure Measurement and Classification Using an Oscillometric Device in Children with Chronic Kidney Disease [J].
Flynn, Joseph T. ;
Pierce, Christopher B. ;
Miller, Edgar R., III ;
Charleston, Jeanne ;
Samuels, Joshua A. ;
Kupferman, Juan ;
Furth, Susan L. ;
Warady, Bradley A. .
JOURNAL OF PEDIATRICS, 2012, 160 (03) :434-U124
[6]   Oscillometric blood pressure estimation: Past, present, and future [J].
Forouzanfar, Mohamad ;
Dajani, Hilmi R. ;
Groza, Voicu Z. ;
Bolic, Miodrag ;
Rajan, Sreeraman ;
Batkin, Izmail .
IEEE Reviews in Biomedical Engineering, 2015, 8 :44-63
[7]   Limitations of current validation protocols for home blood pressure monitors for individual patients [J].
Gerin, W ;
Schwartz, AR ;
Schwartz, JE ;
Pickering, TG ;
Davidson, KW ;
Bress, J ;
O'Brien, E ;
Atkins, N .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :313-318
[8]   A procedure for evaluation of non-invasive blood pressure simulators [J].
Gersak, Gregor ;
Zemva, Ales ;
Drnovsek, Janko .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2009, 47 (12) :1221-1228
[9]   CARDIAC CYCLE PHASE UNCERTAINTY - ANOTHER SOURCE OF ERROR IN INDIRECT BLOOD-PRESSURE MEASUREMENT [J].
HATSELL, CP .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (04) :157-158
[10]   Influence of Different Presentations of Oscillometric Data on Automatic Determination of Systolic and Diastolic Pressures [J].
Jazbinsek, Vojko ;
Luznik, Janko ;
Mieke, Stephan ;
Trontelj, Zvonko .
ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (03) :774-787