Method for Evaluation of Trustworthiness of Oscillometric Blood Pressure Measurements

被引:0
作者
Koohi, Iraj [1 ]
Ahmad, Saif [1 ]
Batkin, Izmail [1 ]
Groza, Voicu Z. [1 ]
Shirmohammadi, Shervin [1 ]
Dajani, Hilmi R. [1 ]
Petriu, Emil M. [1 ]
机构
[1] Univ Ottawa, Sch Elect Engn & Comp Sci, Ottawa, ON, Canada
来源
2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS | 2015年
关键词
Mean arterial pressure; systolic; diastolic; blood pressure; vessel compliance; oscillometric waveform simulation; PULSE-WAVE VELOCITY; ARTERIAL STIFFNESS;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Simple, unobtrusive, and reliable estimation of cardiovascular parameters is a challenge. We present a novel, simple, and noninvasive method called Ratio2 that provides expected ranges for systolic and diastolic blood pressure values (SBP, DBP) estimated by any algorithm, and an evaluation of vessel compliance. Ratio2 was developed in the frame of the oscillometric blood pressure estimation and it exploits the equality between the arterial blood pressure and the cuff pressure at the mean arterial pressure (MAP). This method is based on the observation that the brachial arterial blood pressure pulses, with MAP used as baseline, are characterized by a peak to trough ratio close to 2. This ratio is employed to characterize expected ranges for estimates of systolic and diastolic blood pressure. Any SBP or DBP measurement which is not contained in these intervals is deemed untrustworthy, and it is marked as such. Ratio2 also provides parameters that are used in a mathematical model of arterial blood pressure (BP) to evaluate vessel stiffness. We tested the performance of the Ratio2 method on 150 oscillometric recordings and their corresponding Omron BP estimates obtained from 10 healthy subjects. Results are encouraging, whereby, (a) out-of-range values obtained with the maximum amplitude algorithm (MAA) and the maximum/minimum slope algorithm (MMSA) methods were successfully detected, and (b) linear correlation between age and vessel compliance is -85% (p<0.005). Therefore, we conclude that the proposed work shows promise towards providing noninvasive BP monitors with an inbuilt mechanism for assessing the fidelity of their BP estimates along with an indicator of vessel compliance.
引用
收藏
页码:267 / 272
页数:6
相关论文
共 50 条
  • [41] The effect of clothes on sphygmomanometric and oscillometric blood pressure measurement
    Liebl, ME
    Holzgreve, H
    Schulz, M
    Crispin, A
    Bogner, JR
    BLOOD PRESSURE, 2004, 13 (05): : 279 - 282
  • [42] Patient-Specific Oscillometric Blood Pressure Measurement
    Liu, Jiankun
    Cheng, Hao-Min
    Chen, Chen-Huan
    Sung, Shih-Hsien
    Moslehpour, Mohsen
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (06) : 1220 - 1228
  • [43] Challenges and pitfalls of cuff oscillometric blood pressure measurement
    Paolo Palatini
    Hypertension Research, 2025, 48 (6) : 1993 - 1996
  • [44] Determination of Blood Pressure and Hemodynamics from Oscillometric Waveforms
    Jilek, J.
    Stork, M.
    NEW ASPECTS OF SYSTEMS, PTS I AND II, 2008, : 109 - +
  • [45] Model-Based Oscillometric Blood Pressure Estimation
    Forouzanfar, Mohamad
    Dajani, Hilmi R.
    Groza, Voicu Z.
    Bolic, Miodrag
    2014 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2014, : 443 - 448
  • [46] Accuracy of an oscillometric blood pressure monitor in anesthetized pigs
    Reed, Rachel
    Barletta, Michele
    Grimes, Janet
    Mumaw, Jennifer
    Park, Hea Jin
    Giguere, Steeve
    Azain, Michael
    Fang, Xi
    Quandt, Jane
    LABORATORY ANIMALS, 2018, 52 (05) : 490 - 496
  • [47] Evaluation of an indirect oscillometric blood pressure monitor in anaesthetised dogs at three different anatomical locations
    Fujiyama, M.
    Sano, H.
    Chambers, J. P.
    Gieseg, M.
    NEW ZEALAND VETERINARY JOURNAL, 2017, 65 (04) : 185 - 191
  • [48] Automated Noninvasive Central Blood Pressure Measurements by Oscillometric Radial Pulse Wave Analysis: Results of the MEASURE-cBP Validation Studies
    Trinkmann, Frederik
    Benck, Urs
    Halder, Julian
    Semmelweis, Alexandra
    Saur, Joachim
    Borggrefe, Martin
    Akin, Ibrahim
    Kaden, Jens J.
    AMERICAN JOURNAL OF HYPERTENSION, 2021, 34 (04) : 383 - 393
  • [49] CONCORDANCE BETWEEN NOCTURNAL CARDIAC OUTPUT ESTIMATED WITH A NOVEL NON-INVASIVE METHOD USING BRACHIAL OSCILLOMETRIC BLOOD PRESSURE MEASUREMENTS AND ECHOCARDIOGRAPHY
    Fernandez-Ruiz, Ricardo
    Alvarez-Montoya, Diego
    Jaramillo-Gomez, Carlos
    Madrid-Munoz, Camilo
    Gallo-Villegas, Jaime
    Aristizabal-Ocampo, Dagnovar
    JOURNAL OF HYPERTENSION, 2024, 42 (SUPPL 3)
  • [50] Blood Pressure Estimation using Oscillometric Pulse Morphology
    Mafi, Majid
    Rajan, Sreeraman
    Bolic, Miodrag
    Groza, Voicu Z.
    Dajani, Hilmi R.
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 2492 - 2496