Method for Evaluation of Trustworthiness of Oscillometric Blood Pressure Measurements

被引:0
作者
Koohi, Iraj [1 ]
Ahmad, Saif [1 ]
Batkin, Izmail [1 ]
Groza, Voicu Z. [1 ]
Shirmohammadi, Shervin [1 ]
Dajani, Hilmi R. [1 ]
Petriu, Emil M. [1 ]
机构
[1] Univ Ottawa, Sch Elect Engn & Comp Sci, Ottawa, ON, Canada
来源
2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS | 2015年
关键词
Mean arterial pressure; systolic; diastolic; blood pressure; vessel compliance; oscillometric waveform simulation; PULSE-WAVE VELOCITY; ARTERIAL STIFFNESS;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Simple, unobtrusive, and reliable estimation of cardiovascular parameters is a challenge. We present a novel, simple, and noninvasive method called Ratio2 that provides expected ranges for systolic and diastolic blood pressure values (SBP, DBP) estimated by any algorithm, and an evaluation of vessel compliance. Ratio2 was developed in the frame of the oscillometric blood pressure estimation and it exploits the equality between the arterial blood pressure and the cuff pressure at the mean arterial pressure (MAP). This method is based on the observation that the brachial arterial blood pressure pulses, with MAP used as baseline, are characterized by a peak to trough ratio close to 2. This ratio is employed to characterize expected ranges for estimates of systolic and diastolic blood pressure. Any SBP or DBP measurement which is not contained in these intervals is deemed untrustworthy, and it is marked as such. Ratio2 also provides parameters that are used in a mathematical model of arterial blood pressure (BP) to evaluate vessel stiffness. We tested the performance of the Ratio2 method on 150 oscillometric recordings and their corresponding Omron BP estimates obtained from 10 healthy subjects. Results are encouraging, whereby, (a) out-of-range values obtained with the maximum amplitude algorithm (MAA) and the maximum/minimum slope algorithm (MMSA) methods were successfully detected, and (b) linear correlation between age and vessel compliance is -85% (p<0.005). Therefore, we conclude that the proposed work shows promise towards providing noninvasive BP monitors with an inbuilt mechanism for assessing the fidelity of their BP estimates along with an indicator of vessel compliance.
引用
收藏
页码:267 / 272
页数:6
相关论文
共 50 条
  • [1] Limitations of the of the oscillometric method for blood pressure measurements in dialyzed patients
    Czarkowski, Marek
    Staszkow, Monika
    Kostyra, Kacper
    Shebani, Zuhier
    Rozanowski, Krzysztof
    Matuszkiewicz-Rowinska, Joanna
    Niemczyk, Stanislaw
    MEDICAL SCIENCE MONITOR, 2011, 17 (04): : MT35 - MT40
  • [2] Assessment of Algorithms for Oscillometric Blood Pressure Measurement
    Chen, S.
    Groza, V. Z.
    Bolic, M.
    Dajani, H. R.
    I2MTC: 2009 IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3, 2009, : 1711 - 1715
  • [3] An Experimental Study on the Blood Pressure Cuff as a Transducer for Oscillometric Blood Pressure Measurements
    Bogatu, Laura, I
    Turco, Simona
    Mischi, Massimo
    Muehlsteff, Jens
    Woerlee, Pierre
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2021, 70
  • [4] Quantitative analysis of the effect of clothing on the oscillometric waveform envelope and oscillometric blood pressure measurements
    Kim, Jeonghoon
    Lee, Jongshill
    Lee, Jeyeon
    Park, Hoon-Ki
    Kim, In Young
    BIOMEDICAL ENGINEERING LETTERS, 2025, : 525 - 536
  • [5] NOVEL METHOD FOR ASSESSING ARTERIAL STIFFNESS BASED ON OSCILLOMETRIC BLOOD PRESSURE MEASUREMENT
    Zhang, Chi
    Gao, Yongjia
    Long, Taochen
    Du, Yi
    Hou, Xiaoxu
    Li, Deyu
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2018, 18 (07)
  • [6] A simulation study of the consistency of oscillometric blood pressure measurements with and without artefacts
    Amoore, JN
    BLOOD PRESSURE MONITORING, 2000, 5 (02) : 69 - 79
  • [7] Clinical evaluation of an automated oscillometric blood pressure wrist device
    Cuckson, AC
    Moran, P
    Seed, P
    Reinders, A
    Shennan, AH
    BLOOD PRESSURE MONITORING, 2004, 9 (01) : 31 - 37
  • [8] Acquiring the oscillometric envelopes for non-invasive blood pressure measurements
    Gersak, Gregor
    ELEKTROTEHNISKI VESTNIK-ELECTROCHEMICAL REVIEW, 2009, 76 (03): : 97 - 102
  • [9] Evaluation of arterial stiffness and central blood pressure by oscillometric method in normotensive offspring of hypertensive parents
    Yildirim, Cansu Sivrikaya
    Kosger, Pelin
    Akin, Tugcem
    Ucar, Birsen
    CARDIOLOGY IN THE YOUNG, 2022, 32 (06) : 888 - 895
  • [10] Enhancement of oscillometric index in non-invasive blood pressure measurements
    Jazbinsek, V.
    Luznik, J.
    Trontelj, Z.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 1357 - 1360