Depressor effect of closed-loop chip system in spontaneously hypertensive rats

被引:4
作者
Gao, Xing-Ya [1 ]
Huang, Xing-Lin [1 ]
Wang, Han-Jun [1 ]
Zhou, Li-Min [1 ]
Xu, Yao [1 ]
Wang, Wei [2 ]
Zhu, Guo-Qing [1 ]
机构
[1] Nanjing Med Univ, Dept Physiol, Nanjing 210029, Peoples R China
[2] Univ Nebraska, Coll Med, Dept Cellular & Integrat Physiol, Omaha, NE 68198 USA
来源
AUTONOMIC NEUROSCIENCE-BASIC & CLINICAL | 2007年 / 137卷 / 1-2期
关键词
baroreflex; hypertension; blood pressure; chip;
D O I
10.1016/j.autneu.2007.09.001
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
We previously reported that a closed-loop chip system was designed to decrease arterial pressure in normal rabbits and rats. In the present study, the depressor effects of the chip system were investigated in spontaneously hypertensive rats (SHR) and Wistar-Kyoto rats (WKY). The arterial pressure was recorded, sampled, operated and processed in the chip system. The chip system instantaneously controlled arterial pressure by stimulating the left aortic depressor nerve according to the feedback signals of arterial pressure. The closed-loop chip system effectively decreased mean arterial pressure (MAP) and heart rate (HR) in both SHR and WKY rats. It decreased the duration and the maximal MAP level of the pressor response evoked by either intravenous injection of phenylephrine or cutaneous nociceptive stimulation in SHR, but had no significant effect on the magnitude of the increase in MAP. Furthermore, the chip system significantly increased the baroreflex gain in SHR, but not in normal WKY rats. These results suggest that the closed-loop chip system effectively decreases the arterial pressure and increases baroreflex gain in SHR. The chip system does not abolish the arterial pressure responses to accidental pressor events, but decreases the duration and the maximal MAP level of the pressor responses. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:84 / 91
页数:8
相关论文
共 30 条
[1]   RELATIONSHIP OF AORTIC-WALL AND BARORECEPTOR PROPERTIES DURING DEVELOPMENT IN NORMOTENSIVE AND SPONTANEOUSLY HYPERTENSIVE RATS [J].
ANDRESEN, MC ;
KRAUHS, JM ;
BROWN, AM .
CIRCULATION RESEARCH, 1978, 43 (05) :728-738
[2]   BARORECEPTOR FUNCTION AND CHANGES IN STRAIN SENSITIVITY IN NORMOTENSIVE AND SPONTANEOUSLY HYPERTENSIVE RATS [J].
ANDRESEN, MC ;
KURAOKA, S ;
BROWN, AM .
CIRCULATION RESEARCH, 1980, 47 (06) :821-828
[3]   TREATMENT OF HYPERTENSION WITH AN IMPLANTABLE ELECTRONIC DEVICE [J].
BILGUTAY, AM ;
LILLEHEI, CW .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1965, 191 (08) :649-&
[4]   BILATERAL CAROTID-SINUS NERVE-STIMULATION IN TREATMENT OF HYPERTENSION [J].
BREST, AN ;
BACHRACH, B ;
WIENER, L .
AMERICAN JOURNAL OF CARDIOLOGY, 1972, 29 (06) :821-&
[5]   COMPARISON OF AORTIC BARORECEPTOR DISCHARGE IN NORMOTENSIVE AND SPONTANEOUSLY HYPERTENSIVE RATS [J].
BROWN, AM ;
SAUM, WR ;
TULEY, FH .
CIRCULATION RESEARCH, 1976, 39 (04) :488-496
[6]   AORTIC-WALL PROPERTIES AND BARORECEPTOR BEHAVIOR AT NORMAL ARTERIAL-PRESSURE AND IN ACUTE HYPERTENSIVE RESETTING IN DOGS [J].
COLERIDGE, HM ;
COLERIDGE, JCG ;
POORE, ER ;
ROBERTS, AM ;
SCHULTZ, HD .
JOURNAL OF PHYSIOLOGY-LONDON, 1984, 350 (MAY) :309-326
[7]   Resetting blood pressure by a closed-loop implanted chip system in normotensive rats [J].
Gao, XY ;
Wang, HJ ;
Zhang, Y ;
Lu, ZH ;
Wang, W ;
Zhu, GQ .
LIFE SCIENCES, 2006, 78 (10) :1129-1134
[8]  
Gao XY, 2005, FASEB J, V19, pA1298
[9]  
Gao XY, 2004, J NANJING MED U, V18, P279
[10]   Adrenergic and reflex abnormalities in obesity-related hypertension [J].
Grassi, G ;
Seravalle, G ;
Dell'Oro, R ;
Turri, C ;
Bolla, GB ;
Mancia, G .
HYPERTENSION, 2000, 36 (04) :538-542