X-Ray Thermoluminescence Dosimetry Characterization of Commercially Available CVD Diamond

被引:3
作者
Gil-Tolano, Maria Ines [1 ]
Melendrez, Rodrigo [1 ]
Alvarez-Garcia, Susana [1 ]
Soto-Puebla, Diego [1 ]
Chernov, Valery [1 ]
Barboza-Flores, Marcelino [1 ]
机构
[1] Univ Sonora, Dept Invest Fis, Hermosillo 83190, Sonora, Mexico
来源
PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE | 2018年 / 215卷 / 22期
关键词
diamond CVD crystal; dosimetry; fading; thermoluminescence; X-ray irradiation; TL CHARACTERIZATION; PERFORMANCE; SENSORS; FILMS;
D O I
10.1002/pssa.201800246
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
High quality commercially available CVD diamonds are being tested in novel clinical high energy photons, hadron therapy, and high energy physics applications. In the present work, the authors report on the thermoluminescence (TL) and dosimetry properties of commercial synthetic CVD diamond sample from Element6, exposed to X-ray irradiation at dose rates of 3.9-120 Gy min(-1) from 0.13 to 90 Gy. The TL glow curves are composed of four peaks with maxima at about 104, 150, 210, and 280 degrees C and activation energies of 0.78, 0.79, 0.81, and 1.0 eV, respectively. The dose response of the area under TL glow curves is linear at doses lower than 0.6 Gy, supralinear between 0.6 and 6 Gy and sublinear at higher doses. The samples exhibited a strong TL fading of the low temperature 104 degrees C peaks that fades away around 120 min after irradiation, which is accompanied by increasing of the 210 and 280 degrees C peaks. The measured TL cycle reproducibility was very good, in spite of the fading exhibited by the lower temperature peak, with values of 0.1-3.0% at different dose rates and doses. It is concluded that CVD commercially single crystal electronic grade diamond may be securely used in clinical radiotherapy.
引用
收藏
页数:6
相关论文
共 29 条
  • [2] Diamond sensors for future high energy experiments
    Bachmair, Felix
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 831 : 370 - 377
  • [3] Diamond detectors for high energy physics experiments
    Baeni, L.
    Alexopoulos, A.
    Artuso, M.
    Bachmair, F.
    Bartosik, M.
    Beacham, J.
    Beck, H.
    Bellini, V.
    Belyaev, V.
    Bentele, B.
    Berdermann, E.
    Bergonzo, P.
    Bes, A.
    Brom, J-M.
    Bruzzi, M.
    Cerv, M.
    Chiodini, G.
    Chren, D.
    Cindro, V.
    Claus, G.
    Collot, J.
    Cumalat, J.
    Dabrowski, A.
    D'Alessandro, R.
    Dauvergne, D.
    de Boer, W.
    Dorfer, C.
    Duenser, M.
    Eremin, V.
    Eusebi, R.
    Forcolin, G.
    Forneris, J.
    Frais-Koelbl, H.
    Gallin-Martel, L.
    Gallin-Martel, M. L.
    Gan, K. K.
    Gastal, M.
    Giroletti, C.
    Goffe, M.
    Goldstein, J.
    Golubev, A.
    Gorisek, A.
    Grigoriev, E.
    Grosse-Knetter, J.
    Grummer, A.
    Gui, B.
    Guthoff, M.
    Haughton, I.
    Hiti, B.
    Hits, D.
    [J]. JOURNAL OF INSTRUMENTATION, 2018, 13
  • [4] FIGURE-OF-MERIT (FOM), AN IMPROVED CRITERION OVER NORMALIZED CHI-SQUARED TEST FOR ASSESSING GOODNESS-OF-FIT OF GAMMA-RAY SPECTRAL PEAKS
    BALIAN, HG
    EDDY, NW
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1977, 145 (02): : 389 - 395
  • [5] Batnitsky S. M., 1993, PHYS MED BIOL, V38, P173
  • [6] Thermoluminescence properties of CVD diamond for clinical dosimetry use
    Benabdesselam, M.
    Serrano, B.
    Iacconi, P.
    Wrobel, F.
    Lapraz, D.
    Herault, J.
    Butler, J. E.
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) : 87 - 90
  • [7] TL characterisation of a CVD diamond wafer for ionising radiation dosimetry
    Benabdesselam, M
    Iacconi, P
    Butler, JE
    Nigoul, JM
    [J]. DIAMOND AND RELATED MATERIALS, 2003, 12 (10-11) : 1750 - 1754
  • [8] Performance of CVD diamond as a thermoluminescent dosemeter
    Benabdesselam, M
    Iacconi, P
    Briand, D
    Butler, JE
    [J]. DIAMOND AND RELATED MATERIALS, 2000, 9 (3-6) : 1013 - 1016
  • [9] Improving diamond detectors: A device case
    Bergonzo, P.
    Tromson, D.
    Descamps, C.
    Hannita, H.
    Mer, C.
    Tranchant, N.
    Nesladek, M.
    [J]. DIAMOND AND RELATED MATERIALS, 2007, 16 (4-7) : 1038 - 1043
  • [10] Dose rate effects on the performance of MWCVD diamond films as TL gamma radiation dosimeter
    Cruz-Zaragoza, E.
    Gastelum, S.
    Melendrez, R.
    Chernov, V.
    Barboza-Flores, M.
    [J]. PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2010, 207 (08): : 1944 - 1948