The spectral distribution of the response of the TlInSe2 non-linear γ-ray detector

被引:7
作者
Abdinova, SG [1 ]
Alekseev, IV [1 ]
机构
[1] Joint Inst Nucl Res, Frank Lab Neutron Phys, Dubna 141980, Moscow Region, Russia
关键词
D O I
10.1016/S0168-9002(98)00215-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The spectral distribution of the response of the TlInSe2 gamma-ray detector was calculated using the Monte Carlo method. Non-linearity of the ampere-watt characteristic (AWC) of the detector and the inhomogeneity of radiation absorption in the crystal were corrected by means of the coefficient s less than or equal to 1, which depends on gamma-quantum energy, crystal thickness and the AWC shape. This method of correction is legitimate when the diffusion effect on distribution of non-equilibrium charge carriers in the semiconductor volume is not considerable. Comparison of the calculated and experimental data was carried out in the 0.02-1.25 MeV range. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:365 / 368
页数:4
相关论文
共 7 条
[1]  
Alekseev I.V., 1992, IZV AN USSR INORG MA, V28, P2404
[2]  
AMATUNI CA, 1984, EFI7355084
[3]   COMPARISON OF DIFFERENT GAAS DETECTORS FOR X-RAY DIGITAL RADIOGRAPHY [J].
BENCIVELLI, W ;
BERTIN, R ;
BERTOLUCCI, E ;
BOTTIGLI, U ;
DAURIA, S ;
DELPAPA, C ;
FANTACCI, ME ;
RANDACCIO, P ;
ROSSO, V ;
STEFANINI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 338 (2-3) :549-555
[4]   A CDTE GAMMA-RAY SPECTROMETER WITH IMAGING CAPABILITIES FOR ASTROPHYSICS - MONTE-CARLO SIMULATION RESULTS [J].
CAROLI, E ;
BALDAZZI, G ;
DICOCCO, G ;
DONATI, A ;
DUSI, W ;
MALAGUTI, G ;
ROSSI, M ;
STEPHEN, JB .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :639-643
[5]   USE OF THE EGS4 MONTE-CARLO CODE TO EVALUATE THE RESPONSE OF HGI2 AND CDTE DETECTORS FOR PHOTONS IN THE DIAGNOSTIC ENERGY-RANGE [J].
CONTI, M ;
DELGUERRA, A ;
MAZZEI, D ;
RUSSO, P ;
BENCIVELLI, W ;
BERTOLUCCI, E ;
MESSINEO, A ;
ROSSO, V ;
STEFANINI, A ;
BOTTIGLI, U ;
RANDACCIO, P ;
NELSON, WR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :591-595
[6]  
MASHKEVICH VP, 1982, SHEIELDING IONISING
[7]  
Ryvkin S. M., 1963, PHOTOELECTRIC PHENOM