Determination of dose delivery accuracy and image quality in full - Field digital mammography

被引:2
作者
Sosu, E. K. [1 ,2 ,3 ]
Boadu, M. [1 ,2 ]
Mensah, S. Y. [3 ]
机构
[1] GAEC, Radiol & Med Sci Res Inst, Med Radiat Phys Ctr, POB LG 80, Legon, Accra, Ghana
[2] Univ Ghana, Sch Nucl & Appl Sci, Med Phys Dept, Atomic Campus,POB AE 1, Accra, Ghana
[3] Univ Cape Coast, Phys Dept, Cape Coast, Ghana
关键词
Mammography; Mean glandular dose; Image quality; Quality control; Polymethylmethacrylate; Automatic exposure control; FILM MAMMOGRAPHY; BREAST; UK;
D O I
10.1016/j.jrras.2018.02.002
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Accuracy of dose delivered, image quality and some technical parameters of the Fujifilm - Amuletf full field digital mammography X-ray equipment has been undertaken. The study was conducted to review the overall condition of the first full - field digital mammography equipment in Ghana with the aim of optimizing mammography practice. Quality control assessment and Mean glandular dose estimation was performed using the International Atomic Energy Agency Human Health Series 2 and 17 protocol and the European guidelines for quality assurance in breast cancer screening and diagnosis. Quantitative image analysis was performed with Images software using the "Rose Model". The results from the quality control test performed indicates that the system is functioning well. Signal-to-noise and contrast-to-noise values determined from images shows that the images are of standard quality. With the exception of 9.00 mm breast thickness which recorded a displayed Mean Glandular Dose (MGD) of 7.17 mGy, all other MGD calculated or displayed were well within the acceptable level. The percentage difference between the calculated and console displayed MGD was within the acceptable difference level of 50%. The phantom dose values obtained can be used as baseline data for future studies which can assist in setting optimization activities. The mammography X-ray equipment at the Korle-Bu teaching Hospital, Accra - Ghana is functioning under optimized conditions. It is therefore recommended for further usage. (C) 2018 The Egyptian Society of Radiation Sciences and Applications. Production and hosting by Elsevier B.V.
引用
收藏
页码:232 / 236
页数:5
相关论文
共 11 条
[1]   Signal-to-noise optimization of medical imaging systems [J].
Cunningham, IA ;
Shaw, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03) :621-632
[2]   Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols [J].
Dance, D. R. ;
Young, K. C. ;
van Engen, R. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (02) :453-471
[3]   Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol [J].
Dance, DR ;
Skinner, CL ;
Young, KC ;
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3225-3240
[4]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[5]   Estimation of mean-glandular dose from monitoring breast entrance skin air kerma using a high sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeter system in mammography [J].
Dong, SL ;
Chu, TC ;
Lee, JS ;
Lan, GY ;
Wu, TH ;
Yeh, YH ;
Hwang, JJ .
APPLIED RADIATION AND ISOTOPES, 2002, 57 (06) :791-799
[6]   Does practice make perfect when interpreting mammography? Part II [J].
Elmore, JG ;
Miglioretti, DL ;
Carney, PA .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2003, 95 (04) :250-252
[7]   Comparison of Acquisition Parameters and Breast Dose in Digital Mammography and Screen-Film Mammography in the American College of Radiology Imaging Network Digital Mammographic Imaging Screening Trial [J].
Hendrick, R. Edward ;
Pisano, Etta D. ;
Averbukh, Alice ;
Moran, Catherine ;
Berns, Eric A. ;
Yaffe, Martin J. ;
Herman, Benjamin ;
Acharyya, Suddhasatta ;
Gatsonis, Constantine .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (02) :362-369
[8]   Diagnostic performance of digital versus film mammography for breast-cancer screening [J].
Pisano, ED ;
Gatsonis, C ;
Hendrick, E ;
Yaffe, M ;
Baum, JK ;
Acharyya, S ;
Conant, EF ;
Fajardo, LL ;
Bassett, L ;
D'Orsi, C ;
Jong, R ;
Rebner, M .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 353 (17) :1773-1783
[9]  
Sickles EA, 2000, RADIOLOGY, V215, P1
[10]   Towards a proposition of a diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in Belgium [J].
Smans, K ;
Bosmans, H ;
Xiao, M ;
Carton, AK ;
Marchal, G .
RADIATION PROTECTION DOSIMETRY, 2005, 117 (1-3) :321-326