Effect of the Anatomic Reservoir on Low-Flow Oxygen Delivery Via Nasal Cannula: Constant Flow Versus Pulse Flow With Portable Oxygen Concentrator

被引:19
|
作者
Zhou, Steven [1 ]
Chatburn, Robert L. [1 ,2 ]
机构
[1] Cleveland Clin, Resp Inst, Cleveland, OH 44195 USA
[2] Case Western Reserve Univ, Dept Med, Cleveland Clin, Lerner Coll Med, Cleveland, OH 44106 USA
关键词
lung simulator; nasal cannula; chronic obstructive pulmonary disease; oxygen therapy; OBSTRUCTIVE PULMONARY-DISEASE; LUNG-DISEASE; EXERCISE; SYSTEM; EFFICACY; REST; PERFORMANCE; THERAPY; COPD;
D O I
10.4187/respcare.02878
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The F-IO2 for a nasal cannula with constant flow (CF) depends on the anatomic reservoir (AR), which is affected by changes in frequency and end-expiratory flow. Conversely, pulse flow (PF) devices do not require the AR. The purpose of this study was to compare the F-IO2 delivered by a nasal cannula supplied by CF via oxygen tank with that delivered by PF delivered via portable oxygen concentrator. Hypotheses were (1) a lung model of COPD with non-zero end-expiratory flow decreases F-IO2 for CF more than for PF, and (2) CF and PF perform differently in terms of F-IO2 delivery, despite having equivalent settings. METHODS: Normal and COPD lung models were simulated (Ing Mar Medical ASL 5000) using published human data: normal: breathing frequency = 15 breaths/min, R-in = 4 cm H2O center dot s center dot L-1, R-out = 4 cm H2O center dot s center dot L-1, C = 60 mL center dot cm H2O-1, tidal volume (V-T) = 685 mL, P-max = 11.95 cm H2O, increase = 33%, and release = 28; COPD: breathing frequency = 20 breaths/min, R-in = 12 cm H2O center dot s center dot L-1, R-out = 25 cm H2O center dot s center dot L-1, C = 66 mL center dot cm H2O-1, V-T = 685 mL, P-max = 24.52 cm H2O, increase = 35%, and release = 23%. CF was 1-5 L/min. Portable oxygen concentrators used were Solo(2) (Invacare), XPO2 (Invacare), FreeStyle (AirSep), Focus (AirSep), One G3 (Inogen), and LifeChoice ActivOx (Inova Labs). RESULTS: CF produced significantly higher F-IO2 at all settings for normal lungs but lower for COPD lungs compared with Solo(2). COPD reduced the F-IO2 for CF but had a smaller variable effect for PF. Data show there is no equivalency between PF setting and CF rates for the portable oxygen concentrators tested. CONCLUSIONS: CF oxygen delivery via a nasal cannula is significantly reduced by elimination of the AR in a model of COPD, yielding clinically important decreases in F-IO2. PF (delivered with a portable oxygen concentrator) is relatively unaffected. This study supports the recommendation that clinicians and caretakers should titrate the PF setting to each patient's unique oxygen requirements.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 50 条
  • [41] Non-rebreather mask and low-flow nasal cannula vs high-flow nasal cannula in severe COVID-19 pneumonia in the emergency department
    Kamil, Muhammad Khidir Mohd
    Yoong, Khadijah Poh Yuen
    Azhar, Abdul Muhaimin Noor
    Bustam, Aida
    Abdullah, Ahmad Hariz
    Yusuf, Mohd Hafyzuddin Md
    Zambri, Aliyah
    Zahedi, Ahmad Zulkarnain Ahmad
    Shafie, Hidayah
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2023, 63 : 86 - 93
  • [42] Acute Responses to Oxygen Delivery via High Flow Nasal Cannula in Patients with Severe Chronic Obstructive Pulmonary Disease-HFNC and Severe COPD
    Attaway, Amy H.
    Faress, Jihane
    Jacono, Frank
    Dasarathy, Srinivasan
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (09)
  • [43] For Critically Ill Patients, Is High-Flow Nasal Cannula Oxygen Delivery a Suitable Alternative to Mechanical Ventilation?
    Nishimura, Masaji
    RESPIRATORY CARE, 2015, 60 (02) : 307 - 308
  • [44] Narrative review of practical aspects of aerosol delivery via high-flow nasal cannula
    Li, Jie
    Fink, James B.
    ANNALS OF TRANSLATIONAL MEDICINE, 2021, 9 (07)
  • [45] Comparisons of effects of high-flow and nasal cannula oxygen on exercise performance in patients with COPD
    Chen, Yen-Huey
    Cheng, Shiao-Lan
    Wu, Huang-Pin
    Lin, Hui-Ling
    EUROPEAN RESPIRATORY JOURNAL, 2020, 56
  • [46] High-flow nasal cannula oxygen supply as treatment in hypercapnic respiratory failure
    Lepere, Victoria
    Messika, Jonathan
    La Combe, Beatrice
    Ricard, Jean-Damien
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (09) : 1914.e1 - 1914.e2
  • [47] Evaluation of oxygen administration with a high-flow nasal cannula to clinically normal dogs
    Daly, Jennifer L.
    Guenther, Christine L.
    Haggerty, Jamie M.
    Keir, Iain
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2017, 78 (05) : 624 - 630
  • [48] Right Versus Left Prong Nasal Cannula Flow Delivery and the Effects of Nasal Cycling on Inspired FIO2 in an Adult Anatomic Model
    Marshall, S. Gregory
    Henry, Nicholas R.
    Russian, Christopher J.
    RESPIRATORY CARE, 2016, 61 (04) : 397 - 404
  • [49] Efficacy of High-Flow Nasal Cannula Oxygen Therapy in Patients with Mild Hypercapnia
    Su, Lingling
    Zhao, Qinyu
    Liu, Taotao
    Xu, Yujun
    Li, Weichun
    Zhang, Aiping
    LUNG, 2021, 199 (05) : 447 - 456
  • [50] High-flow nasal cannula oxygen for bronchiolitis in a pediatric ward: a pilot study
    Bressan, Silvia
    Balzani, Marco
    Krauss, Baruch
    Pettenazzo, Andrea
    Zanconato, Stefania
    Baraldi, Eugenio
    EUROPEAN JOURNAL OF PEDIATRICS, 2013, 172 (12) : 1649 - 1656