Effect of the Anatomic Reservoir on Low-Flow Oxygen Delivery Via Nasal Cannula: Constant Flow Versus Pulse Flow With Portable Oxygen Concentrator

被引:19
作者
Zhou, Steven [1 ]
Chatburn, Robert L. [1 ,2 ]
机构
[1] Cleveland Clin, Resp Inst, Cleveland, OH 44195 USA
[2] Case Western Reserve Univ, Dept Med, Cleveland Clin, Lerner Coll Med, Cleveland, OH 44106 USA
关键词
lung simulator; nasal cannula; chronic obstructive pulmonary disease; oxygen therapy; OBSTRUCTIVE PULMONARY-DISEASE; LUNG-DISEASE; EXERCISE; SYSTEM; EFFICACY; REST; PERFORMANCE; THERAPY; COPD;
D O I
10.4187/respcare.02878
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The F-IO2 for a nasal cannula with constant flow (CF) depends on the anatomic reservoir (AR), which is affected by changes in frequency and end-expiratory flow. Conversely, pulse flow (PF) devices do not require the AR. The purpose of this study was to compare the F-IO2 delivered by a nasal cannula supplied by CF via oxygen tank with that delivered by PF delivered via portable oxygen concentrator. Hypotheses were (1) a lung model of COPD with non-zero end-expiratory flow decreases F-IO2 for CF more than for PF, and (2) CF and PF perform differently in terms of F-IO2 delivery, despite having equivalent settings. METHODS: Normal and COPD lung models were simulated (Ing Mar Medical ASL 5000) using published human data: normal: breathing frequency = 15 breaths/min, R-in = 4 cm H2O center dot s center dot L-1, R-out = 4 cm H2O center dot s center dot L-1, C = 60 mL center dot cm H2O-1, tidal volume (V-T) = 685 mL, P-max = 11.95 cm H2O, increase = 33%, and release = 28; COPD: breathing frequency = 20 breaths/min, R-in = 12 cm H2O center dot s center dot L-1, R-out = 25 cm H2O center dot s center dot L-1, C = 66 mL center dot cm H2O-1, V-T = 685 mL, P-max = 24.52 cm H2O, increase = 35%, and release = 23%. CF was 1-5 L/min. Portable oxygen concentrators used were Solo(2) (Invacare), XPO2 (Invacare), FreeStyle (AirSep), Focus (AirSep), One G3 (Inogen), and LifeChoice ActivOx (Inova Labs). RESULTS: CF produced significantly higher F-IO2 at all settings for normal lungs but lower for COPD lungs compared with Solo(2). COPD reduced the F-IO2 for CF but had a smaller variable effect for PF. Data show there is no equivalency between PF setting and CF rates for the portable oxygen concentrators tested. CONCLUSIONS: CF oxygen delivery via a nasal cannula is significantly reduced by elimination of the AR in a model of COPD, yielding clinically important decreases in F-IO2. PF (delivered with a portable oxygen concentrator) is relatively unaffected. This study supports the recommendation that clinicians and caretakers should titrate the PF setting to each patient's unique oxygen requirements.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 26 条
[1]   NEW OXYGEN CANNULA SYSTEM USING INTERMITTENT-DEMAND NASAL FLOW [J].
AUERBACH, D ;
FLICK, MR ;
BLOCK, AJ .
CHEST, 1978, 74 (01) :39-44
[2]  
Boatright J, 2012, RESP CARE PRINCIPLES, P282
[3]   PERFORMANCE OF A DEMAND OXYGEN SAVER SYSTEM DURING REST, EXERCISE, AND SLEEP IN HYPOXEMIC PATIENTS [J].
BOWER, JS ;
BROOK, CJ ;
ZIMMER, K ;
DAVIS, D .
CHEST, 1988, 94 (01) :77-80
[4]   COMPARISON OF 6 OXYGEN DELIVERY SYSTEMS FOR COPD PATIENTS AT REST AND DURING EXERCISE [J].
BRAUN, SR ;
SPRATT, G ;
SCOTT, GC ;
ELLERSIECK, M .
CHEST, 1992, 102 (03) :694-698
[5]   DEMAND OXYGEN DELIVERY FOR PATIENTS WITH RESTRICTIVE LUNG-DISEASE [J].
CARTER, R ;
TASHKIN, D ;
DJAHED, B ;
HATHAWAY, E ;
NICOTRA, MB ;
TIEP, BL .
CHEST, 1989, 96 (06) :1307-1311
[6]  
Chatburn RL, 2010, RESP CARE, V55, P433
[7]   Comparison of four demand oxygen delivery systems at rest and during exercise for chronic obstructive pulmonary disease [J].
Fuhrman, C ;
Chouaid, C ;
Herigault, R ;
Housset, B ;
Adnot, S .
RESPIRATORY MEDICINE, 2004, 98 (10) :938-944
[8]   RESPONSE OF NORMAL SUBJECTS TO INSPIRATORY RESISTIVE UNLOADING [J].
GALLAGHER, CG ;
SANII, R ;
YOUNES, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (03) :1113-1119
[9]   STEADY-STATE RESPONSE OF NORMAL SUBJECTS TO INSPIRATORY RESISTIVE LOAD [J].
HOF, VI ;
WEST, P ;
YOUNES, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 60 (05) :1471-1481
[10]   CLINICAL EFFICACY AND COST-BENEFIT OF PULSE FLOW OXYGEN IN HOSPITALIZED-PATIENTS [J].
KERBY, GR ;
ODONOHUE, WJ ;
ROMBERGER, DJ ;
HANSON, FN ;
KOENIG, GA .
CHEST, 1990, 97 (02) :369-372