Effect of the Anatomic Reservoir on Low-Flow Oxygen Delivery Via Nasal Cannula: Constant Flow Versus Pulse Flow With Portable Oxygen Concentrator

被引:19
|
作者
Zhou, Steven [1 ]
Chatburn, Robert L. [1 ,2 ]
机构
[1] Cleveland Clin, Resp Inst, Cleveland, OH 44195 USA
[2] Case Western Reserve Univ, Dept Med, Cleveland Clin, Lerner Coll Med, Cleveland, OH 44106 USA
关键词
lung simulator; nasal cannula; chronic obstructive pulmonary disease; oxygen therapy; OBSTRUCTIVE PULMONARY-DISEASE; LUNG-DISEASE; EXERCISE; SYSTEM; EFFICACY; REST; PERFORMANCE; THERAPY; COPD;
D O I
10.4187/respcare.02878
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The F-IO2 for a nasal cannula with constant flow (CF) depends on the anatomic reservoir (AR), which is affected by changes in frequency and end-expiratory flow. Conversely, pulse flow (PF) devices do not require the AR. The purpose of this study was to compare the F-IO2 delivered by a nasal cannula supplied by CF via oxygen tank with that delivered by PF delivered via portable oxygen concentrator. Hypotheses were (1) a lung model of COPD with non-zero end-expiratory flow decreases F-IO2 for CF more than for PF, and (2) CF and PF perform differently in terms of F-IO2 delivery, despite having equivalent settings. METHODS: Normal and COPD lung models were simulated (Ing Mar Medical ASL 5000) using published human data: normal: breathing frequency = 15 breaths/min, R-in = 4 cm H2O center dot s center dot L-1, R-out = 4 cm H2O center dot s center dot L-1, C = 60 mL center dot cm H2O-1, tidal volume (V-T) = 685 mL, P-max = 11.95 cm H2O, increase = 33%, and release = 28; COPD: breathing frequency = 20 breaths/min, R-in = 12 cm H2O center dot s center dot L-1, R-out = 25 cm H2O center dot s center dot L-1, C = 66 mL center dot cm H2O-1, V-T = 685 mL, P-max = 24.52 cm H2O, increase = 35%, and release = 23%. CF was 1-5 L/min. Portable oxygen concentrators used were Solo(2) (Invacare), XPO2 (Invacare), FreeStyle (AirSep), Focus (AirSep), One G3 (Inogen), and LifeChoice ActivOx (Inova Labs). RESULTS: CF produced significantly higher F-IO2 at all settings for normal lungs but lower for COPD lungs compared with Solo(2). COPD reduced the F-IO2 for CF but had a smaller variable effect for PF. Data show there is no equivalency between PF setting and CF rates for the portable oxygen concentrators tested. CONCLUSIONS: CF oxygen delivery via a nasal cannula is significantly reduced by elimination of the AR in a model of COPD, yielding clinically important decreases in F-IO2. PF (delivered with a portable oxygen concentrator) is relatively unaffected. This study supports the recommendation that clinicians and caretakers should titrate the PF setting to each patient's unique oxygen requirements.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 50 条
  • [1] Aerosol Delivery Through an Adult High-Flow Nasal Cannula Circuit Using Low-Flow Oxygen
    Madney, Yasmin M.
    Fathy, Maha
    Elberry, Ahmed A.
    Rabea, Hoda
    Abdelrahim, Mohamed E. A.
    RESPIRATORY CARE, 2019, 64 (04) : 453 - 461
  • [2] High-Flow and Low-Flow Oxygen Delivery by Nasal Cannula Evaluated in Infant and Adult Airway Replicas
    Sabz, Mozhgan
    Noga, Michelle L.
    Finlay, Warren H.
    Rouhani, Hossein
    Martin, Andrew R.
    RESPIRATORY CARE, 2024, 69 (04) : 438 - 448
  • [3] Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas
    Sabz, Mozhgan
    Tavernini, Scott
    Pillay, Kineshta
    Christianson, Cole
    Noga, Michelle
    Finlay, Warren H.
    Rouhani, Hossein
    Martin, Andrew R.
    RESPIRATORY RESEARCH, 2022, 23 (01)
  • [4] Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas
    Mozhgan Sabz
    Scott Tavernini
    Kineshta Pillay
    Cole Christianson
    Michelle Noga
    Warren H. Finlay
    Hossein Rouhani
    Andrew R. Martin
    Respiratory Research, 23
  • [5] In Vitro-In Silico Comparison of Pulsed Oxygen Delivery From Portable Oxygen Concentrators Versus Continuous Flow Oxygen Delivery
    Chen, John Z.
    Katz, Ira M.
    Pichelin, Marine
    Zhu, Kaixian
    Caillibotte, Georges
    Finlay, Warren H.
    Martin, Andrew R.
    RESPIRATORY CARE, 2019, 64 (02) : 117 - 129
  • [6] Using a high-flow nasal cannula provided superior results to low-flow oxygen delivery in moderate to severe bronchiolitis
    Milani, Gregorio P.
    Plebani, Anna M.
    Arturi, Elisa
    Brusa, Danila
    Esposito, Susanna
    Dell'Era, Laura
    Laicini, Emanuela A.
    Consonni, Dario
    Agostoni, Carlo
    Fossali, Emilio F.
    ACTA PAEDIATRICA, 2016, 105 (08) : E368 - E372
  • [7] Low-flow nasal cannula oxygen and potential nosocomial spread of COVID-19
    Goldhaber-Fiebert, Sara N.
    Greene, Jeremy A.
    Garibaldi, Brian T.
    BRITISH JOURNAL OF ANAESTHESIA, 2020, 125 (03) : E309 - E310
  • [8] Performance of Different Low-Flow Oxygen Delivery Systems
    Duprez, Frederic
    Dupriez, Florence
    De Greef, Julien
    Gabriel, Julie
    Bruyneel, Arnaud
    Reychler, Gregory
    De Terwangne, Christophe
    Poncin, William
    RESPIRATORY CARE, 2022, 67 (03) : 322 - 330
  • [9] Fraction of Inspired Oxygen With Low-Flow Versus High-Flow Devices: A Simulation Study
    Kojima, Yuki
    Sendo, Ryozo
    Okayama, Naoko
    Hamasaki, Junichiro
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (05)
  • [10] Performance of different low-flow oxygen delivery systems
    Poncin, William
    Reychler, Gregory
    De Terwangne, Christophe
    Marnette, Matthieu
    Duprez, Frederic
    EUROPEAN RESPIRATORY JOURNAL, 2021, 58