Knee Angle Measurement Device Using Fiber Bragg Grating Sensor

被引:33
作者
Pant, Shweta [1 ]
Umesh, Sharath [1 ]
Asokan, Sundarrajan [1 ]
机构
[1] Indian Inst Sci, Dept Instrumentat & Appl Phys, Bengaluru 560012, India
关键词
Fiber Bragg grating sensor; knee angle measurement device; gait analysis; POLYMER OPTICAL-FIBER; PHYSICAL-THERAPY; JOINT KINEMATICS; MOTION; GAIT; FABRICATION; RANGE; HIP;
D O I
10.1109/JSEN.2018.2875564
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Human joint kinematics has established its capacity as a vital tool for joint pathology and gait analysis. Range of motion (ROM) of a knee joint performed by flexion and extension is an essential tool to assess the efficacy of surgical interventions or therapeutic assessment. This paper proposes a novel wearable device for dynamic measurement of knee joint angle using fiber Bragg grating (FRG) sensor. The proposed knee angle measurement device (KAMD) can transduce the angular movement between the shin and the thigh into strain variations on a cantilever, which is acquired by the FRG sensor over it. The KAMD is also facilitated with tuneable sensitivity with three modes of high sensitivity mode, medium sensitivity mode, and low sensitivity mode, which can be opted based on application specific resolution and range of measurement. The ROM data obtained from the KAMD are validated against ROM obtained from commercially available Polhemus Sensors. Furthermore, the synchronization of the knee joint movement of both legs during walking along with stance and swing phase is illustrated, which shows the ability of KAMD to perform gait analysis. Employment of KAMDs on both legs eliminating the time synchronization complexity makes the proposed FBG-hased KAMD an efficient means to evaluate the knee angle from both legs simultaneously.
引用
收藏
页码:10034 / 10040
页数:7
相关论文
共 40 条
[21]   Intra-rater and inter-rater reliability of gait measurements with CODA mpx30 motion analysis system [J].
Maynard, V ;
Bakheit, AMO ;
Oldham, J ;
Freeman, J .
GAIT & POSTURE, 2003, 17 (01) :59-67
[22]   Accuracy of Femur Angles Estimated by IMUs During Clinical Procedures Used to Diagnose Femoroacetabular Impingement [J].
McGinnis, Ryan S. ;
Cain, Stephen M. ;
Tao, Sui ;
Whiteside, David ;
Goulet, Grant C. ;
Gardner, Elizabeth C. ;
Bedi, Asheesh ;
Perkins, Noel C. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (06) :1503-1513
[23]   A BRAGG GRATING-TUNED FIBER LASER STRAIN SENSOR SYSTEM [J].
MELLE, SM ;
ALAVIE, AT ;
KARR, S ;
COROY, T ;
LIU, KX ;
MEASURES, RM .
IEEE PHOTONICS TECHNOLOGY LETTERS, 1993, 5 (02) :263-266
[24]   Bragg grating inscription in various optical fibers with femtosecond infrared lasers and a phase mask [J].
Mihailov, Stephen J. ;
Grobnic, Dan ;
Smelser, Christopher W. ;
Lu, Ping ;
Walker, Robert B. ;
Ding, Huimin .
OPTICAL MATERIALS EXPRESS, 2011, 1 (04) :754-765
[25]  
MOREY WW, 1990, P SOC PHOTO-OPT INS, V1169, P98
[26]   WALKING PATTERNS IN HEALTHY OLD MEN [J].
MURRAY, MP ;
KORY, RC ;
CLARKSON, BH .
JOURNALS OF GERONTOLOGY, 1969, 24 (02) :169-&
[27]  
MURRAY MP, 1981, CLIN ORTHOP RELAT R, P119
[28]   Fiber Bragg gratings [J].
Othonos, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (12) :4309-4341
[29]  
Othonos A., 1999, ART H OPTO
[30]   NORMAL RANGE OF MOTION OF THE HIP, KNEE AND ANKLE JOINTS IN MALE-SUBJECTS, 30-40 YEARS OF AGE [J].
ROAAS, A ;
ANDERSSON, GBJ .
ACTA ORTHOPAEDICA SCANDINAVICA, 1982, 53 (02) :205-208