Sensing stability and dynamic response of the F-Scan in-shoe sensing system: A technical note

被引:0
作者
Sumiya, T
Suzuki, Y
Kasahara, T
Ogata, H
机构
[1] Hamamatsu Rosai Hosp, Dept Rehabil, Hamamatsu, Shizuoka, Japan
[2] Univ Occupat & Environm Hlth, Dept Rehabil Med, Kitakyushu, Fukuoka 807, Japan
[3] Rosai Rehabil Engn Ctr, Nagoya, Aichi 455, Japan
来源
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT | 1998年 / 35卷 / 02期
关键词
foot; gait analysis; insole sensors; orthotic devices; plantar pressure;
D O I
暂无
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
The objective of this study was to gain a better understanding of the F-Scan to determine its appropriate clinical application. Vertical pressure was applied to a sensor foil over the range of 10-80 kPa with or without the intervention of 0.2-0,8 mm thick felt. Sensor sensitivity reached a maximum without. the felt, and decreased with increasing felt thickness, stabilizing at 48-74% of the maximum level when felt thickness exceeded 0.4 mm. This sensitivity change was caused by the slight difference in thickness of sensing areas from that of non-sensing areas. Dynamic response time was delayed by a mean of 0.32 s. Although the cause of this dynamic response delay remains unclear, it was considered to be inappropriate for accurate dynamic measurements. Therefore, rather than using F-Scan measurement to accurately obtain actual values, it should be used for relative comparisons of the plantar pressure distributions under constant conditions.
引用
收藏
页码:192 / 200
页数:9
相关论文
共 11 条
  • [1] BAUMANN W, 1992, VDI BERICHT, V940, P413
  • [2] Baumann W., 1994, MED ORTH TECH, V114, P6
  • [3] CAVANAGH P.R., 1992, INT J CLIN FOOT SCI, V2, P185, DOI [10.1016/0958-2592(92)90047-S, DOI 10.1016/0958-2592(92)90047-S, 10.1016/0958-2592, DOI 10.1016/0958-2592]
  • [4] PLANTAR PRESSURE MEASUREMENTS AND THE PREVENTION OF ULCERATION IN THE DIABETIC FOOT
    DUCKWORTH, T
    BOULTON, AJM
    BETTS, RP
    FRANKS, CI
    WARD, JD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (01): : 79 - 85
  • [5] FREDERICK EC, 1993, SENSORS JUN, P30
  • [6] RELIABILITY OF PRESSURE MEASUREMENTS - THE EMED F SYSTEM
    HUGHES, J
    PRATT, L
    LINGE, K
    CLARK, P
    KLENERMAN, L
    [J]. CLINICAL BIOMECHANICS, 1991, 6 (01) : 14 - 18
  • [7] LORD M, 1994, J REHABIL RES DEV, V31, P214
  • [8] McPoil T. G., 1995, LOWER EXTREMITY, V2, P95
  • [9] A PORTABLE INSOLE PLANTAR PRESSURE MEASUREMENT SYSTEM
    WERTSCH, JJ
    WEBSTER, JG
    TOMPKINS, WJ
    [J]. JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 1992, 29 (01): : 13 - 18
  • [10] Young C R, 1993, Clin Podiatr Med Surg, V10, P455