Implementation and evaluation of a protocol management system for automated review of CT protocols

被引:11
作者
Grimes, Joshua [1 ]
Leng, Shuai [1 ]
Zhang, Yi [1 ]
Vrieze, Thomas [1 ]
McCollough, Cynthia [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
关键词
computed tomography; protocol monitoring; protocol management; radiation safety; OPTIMIZATION;
D O I
10.1120/jacmp.v17i5.6164
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Protocol review is important to decrease the risk of patient injury and increase the consistency of CT image quality. A large volume of CT protocols makes manual review labor-intensive, error-prone, and costly. To address these challenges, we have developed a software system for automatically managing and monitoring CT protocols on a frequent basis. This article describes our experiences in the implementation and evaluation of this protocol monitoring system. In particular, we discuss various strategies for addressing each of the steps in our protocol-monitoring workflow, which are: maintaining an accurate set of master protocols, retrieving protocols from the scanners, comparing scanner protocols to master protocols, reviewing flagged differences between the scanner and master protocols, and updating the scanner and/or master protocols. In our initial evaluation focusing only on abdomen and pelvis protocols, we detected 309 modified protocols in a 24-week trial period. About one-quarter of these modified protocols were determined to contain inappropriate (i.e., erroneous) protocol parameter modifications that needed to be corrected on the scanner. The most frequently affected parameter was the series description, which was inappropriately modified 47 times. Two inappropriate modifications were made to the tube current, which is particularly important to flag as this parameter impacts both radiation dose and image quality. The CT protocol changes detected in this work provide strong motivation for the use of an automated CT protocol quality control system to ensure protocol accuracy and consistency.
引用
收藏
页码:523 / 533
页数:11
相关论文
共 9 条
[1]  
Cody DD, 2012, 2012 COMPUTED TOMOGR
[2]  
Cody DD, 2013, J APPL CLIN MED PHYS, V14, P3, DOI 10.1120/jacmp.v14i5.4462
[3]   CT Radiation Dose Optimization and Tracking Program at a Large Quaternary-Care Health Care System [J].
Goenka, Ajit H. ;
Dong, Frank ;
Wildman, Bonnie ;
Hulme, Katie ;
Johnson, Paul ;
Herts, Brian R. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2015, 12 (07) :703-710
[4]   CT Protocol Review and Optimization [J].
Kofler, James M. ;
Cody, Dianna D. ;
Morin, Richard L. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2014, 11 (03) :267-270
[5]   Methods for CT Automatic Exposure Control Protocol Translation Between Scanner Platforms [J].
McKenney, Sarah E. ;
Seibert, J. Anthony ;
Lamba, Ramit ;
Boone, John M. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2014, 11 (03) :285-291
[6]  
Szczykutowicz TP, 2015, J APPL CLIN MED PHYS, V16, P228, DOI [10.1120/jacmp.v16i4.5412, 10.1120/jacmp.v16i2.5023]
[7]   On the Same Page-Physicist and Radiologist Perspectives on Protocol Management and Review [J].
Szczykutowicz, Timothy P. ;
Siegelman, Jenifer .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2015, 12 (08) :808-814
[8]   FDA Investigates the Safety of Brain Perfusion CT [J].
Wintermark, M. ;
Lev, M. H. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (01) :2-3
[9]   Data-Driven CT Protocol Review and Management-Experience From a Large Academic Hospital [J].
Zhang, Da ;
Savage, Cristy A. ;
Li, Xinhua ;
Liu, Bob .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2015, 12 (03) :267-272