Electrode fabrication and interface optimization for imaging of evoked peripheral nervous system activity with electrical impedance tomography (EIT)

被引:25
作者
Chapman, Christopher A. R. [1 ]
Aristovich, Kirill [1 ]
Donega, Matteo [2 ]
Fjordbakk, Cathrine T. [3 ]
Stathopoulou, Thaleia-Rengina [3 ]
Viscasillas, Jaime [3 ]
Avery, James [1 ]
Perkins, Justin D. [3 ]
Holder, David [1 ]
机构
[1] UCL, Dept Med Phys & Biomed Engn, Gower St, London WC1E 6BT, England
[2] Galvani Bioelect, Neuromodulat Devices Team, Stevenage SG1 2NY, Herts, England
[3] Royal Vet Coll, Dept Clin Sci & Serv, Hatfield AL9 7TA, Herts, England
基金
英国工程与自然科学研究理事会;
关键词
electrical impedance tomography; neural recording; peripheral nerve cuff; electrode characterization; laser-based fabrication; NEURAL STIMULATION; FIBERS; BRAIN;
D O I
10.1088/1741-2552/aae868
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Non-invasive imaging techniques are undoubtedly the ideal methods for continuous monitoring of neural activity. One such method, fast neural electrical impedance tomography (EIT) has been developed over the past decade in order to image neural action potentials with non-penetrating electrode arrays. Approach. The goal of this study is two-fold. First, we present a detailed fabrication method for silicone-based multiple electrode arrays which can be used for epicortical or neural cuff applications. Secondly, we optimize electrode material coatings in order to achieve the best accuracy in EIT reconstructions. Main results. The testing of nanostructured electrode interface materials consisting of platintun, iridium oxide, and PEDOT:pTS in saline tank experiments demonstrated that the PEDOT:pTS coating used in this study leads to more accurate reconstruction dimensions along with reduced phase separation between recording channels. The PEDOT:pTS electrodes were then used in vivo to successfully image and localize the evoked activity of the recurrent laryngeal fascicle from within the cervical vagus nerve. Significance. These results alongside the simple fabrication method presented here position EIT as an effective method to image neural activity.
引用
收藏
页数:11
相关论文
共 44 条
[1]   GREIT: a unified approach to 2D linear EIT reconstruction of lung images [J].
Adler, Andy ;
Arnold, John H. ;
Bayford, Richard ;
Borsic, Andrea ;
Brown, Brian ;
Dixon, Paul ;
Faes, Theo J. C. ;
Frerichs, Inez ;
Gagnon, Herve ;
Gaerber, Yvo ;
Grychtol, Bartlomiej ;
Hahn, Guenter ;
Lionheart, William R. B. ;
Malik, Anjum ;
Patterson, Robert P. ;
Stocks, Janet ;
Tizzard, Andrew ;
Weiler, Norbert ;
Wolf, Gerhard K. .
PHYSIOLOGICAL MEASUREMENT, 2009, 30 (06) :S35-S55
[2]   Imaging fast neural traffic at fascicular eve with electrical impedance tomography: proof of principle in rat sciatic nerve [J].
Aristovich, Kirill ;
Donega, Matteo ;
Blochet, Camille ;
Avery, James ;
Hannan, Sana ;
Chew, Daniel J. ;
Holder, David .
JOURNAL OF NEURAL ENGINEERING, 2018, 15 (05)
[3]   Imaging fast electrical activity in the brain with electrical impedance tomography [J].
Aristovich, Kirill Y. ;
Packham, Brett C. ;
Koo, Hwan ;
dos Santos, Gustavo Sato ;
McEvoy, Andy ;
Holder, David S. .
NEUROIMAGE, 2016, 124 :204-213
[4]   Investigation of potential artefactual changes in measurements of impedance changes during evoked activity: implications to electrical impedance tomography of brain function [J].
Aristovich, Kirill Y. ;
Dos Santos, Gustavo S. ;
Holder, David S. .
PHYSIOLOGICAL MEASUREMENT, 2015, 36 (06) :1245-1259
[5]   A method for reconstructing tomographic images of evoked neural activity with electrical impedance tomography using intracranial planar arrays [J].
Aristovich, Kirill Y. ;
dos Santos, Gustavo Sato ;
Packham, Brett C. ;
Holder, David S. .
PHYSIOLOGICAL MEASUREMENT, 2014, 35 (06) :1095-1109
[6]   A Versatile and Reproducible Multi-Frequency Electrical Impedance Tomography System [J].
Avery, James ;
Dowrick, Thomas ;
Faulkner, Mayo ;
Goren, Nir ;
Holder, David .
SENSORS, 2017, 17 (02)
[7]   Flexible, all-polymer microelectrode arrays for the capture of cardiac and neuronal signals [J].
Blau, Axel ;
Murr, Angelika ;
Wolff, Sandra ;
Sernagor, Evelyne ;
Medini, Paolo ;
Iurilli, Giuliano ;
Ziegler, Christiane ;
Benfenati, Fabio .
BIOMATERIALS, 2011, 32 (07) :1778-1786
[8]  
Chapman C A R, 2017, ADV FUNCT MATER, V4
[9]  
Cogan SF, 2004, P ANN INT IEEE EMBS, V26, P4153
[10]   Neural stimulation and recording electrodes [J].
Cogan, Stuart F. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2008, 10 :275-309