IMAGE REGISTRATION BASED 3D TEE-EM CALIBRATION

被引:6
|
作者
Huang, Xishi [1 ,2 ]
Gutierrez, Luis F. [1 ]
Stanton, Douglas [1 ]
Kim, Peter C. W. [2 ]
Jain, Ameet [1 ]
机构
[1] Philips Res N Amer, Briarcliff Manor, NY USA
[2] Hosp Sick Children, Toronto, ON M5G 1X8, Canada
来源
2010 7TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO | 2010年
关键词
US calibration; 3D TEE; electromagnetic tracking; motion-based calibration; image registration based calibration; HAND-EYE CALIBRATION; ULTRASOUND;
D O I
10.1109/ISBI.2010.5490212
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Real-time 3D trans-esophageal echo (TEE) is being integrated into routine clinical practice for minimally invasive cardiac therapy. Electromagnetically (EM) tracking of the TEE probe inside the body may facilitate intra-procedural heart surgery and therapy on a beating heart. This requires an accurate calibration between the 3D TEE and the EM tracking system. The objective of calibration is to find the transformation that converts the coordinates of voxels in ultrasound volumes into the coordinate system of a position sensor attached to the TEE probe. An image registration based ultrasound calibration method is presented and validated using phantoms. The experiments indicate a post-calibration image reconstruction precision of 1.56 mm and accuracy of 0.85 mm (relative accuracy of 1.86%), and TRE of 2.37 mm. The potential impact of this work is that it can provide an easy to use calibration for clinical use.
引用
收藏
页码:1209 / 1212
页数:4
相关论文
共 50 条
  • [1] Calibration method by image registration with synthetic image of 3D model
    Tamaki, T
    Yamamoto, M
    IEICE TRANSACTIONS ON INFORMATION AND SYSTEMS, 2003, E86D (05): : 981 - 985
  • [2] Registration of Image and 3D LIDAR Data from Extrinsic Calibration
    Hu, Zhaozheng
    Li, Yicheng
    Hu, Yuezhi
    Huang, Gang
    3RD INTERNATIONAL CONFERENCE ON TRANSPORTATION INFORMATION AND SAFETY (ICTIS 2015), 2015, : 102 - 106
  • [3] An image registration approach to automated calibration for freehand 3D ultrasound
    Blackall, JM
    Rueckert, D
    Maurer, CR
    Penney, GP
    Hill, DLG
    Hawkes, DJ
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2000, 2000, 1935 : 462 - 471
  • [4] Progressive 3D biomedical image registration network based on deep self-calibration
    Sun, Rui
    Wu, Jun
    Miao, Yongchun
    Ouyang, Lei
    Qu, Lei
    FRONTIERS IN NEUROINFORMATICS, 2022, 16
  • [5] Mutifractals based multimodal 3D image registration
    Palanivel, Dhevendra Alagan
    Natarajan, Sivakumaran
    Gopalakrishnan, Sainarayanan
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2019, 47 : 126 - 136
  • [6] 3D TEE Registration with MR for Cardiac Interventional Applications
    Woo, Jonghye
    Parthasarathy, Vijay
    Sandeep, Dalal
    Jain, Ameet
    MEDICAL IMAGING 2010: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, 2010, 7625
  • [7] Automatic 3D ultrasound calibration for image guided therapy using intramodality image registration
    Schlosser, Jeffrey
    Kirmizibayrak, Can
    Shamdasani, Vijay
    Metz, Steve
    Hristov, Dimitre
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (21): : 7481 - 7496
  • [8] Accelerated 3D image registration
    Vester-Christensen, Martin
    Erbou, Soren G.
    Darkner, Sune
    Larsen, Rasmus
    MEDICAL IMAGING 2007: IMAGE PROCESSING, PTS 1-3, 2007, 6512
  • [9] Super resolution for fundoscopy based on 3D image registration
    Hernandez-Matas, Carlos
    Zabulis, Xenophon
    2014 36TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2014, : 6332 - 6338
  • [10] Intensity-based 3D local image registration
    Song, Huajun
    Qiu, Peihua
    PATTERN RECOGNITION LETTERS, 2017, 94 : 15 - 21