摘要
Objective: To evaluate the safety and feasibility of transcatheter closure of atrial septal defect (ASD) without balloon sizing. Methods: A total of 243 patients (group I), aged 2.1-76 years (median 22 years), underwent transcatheter closure of ASD without balloon sizing. The maximal diameter of the defect was measured on transesophageal echocardiographic (TEE) images. The size of device selected was generally 4-6 mm and 5-8 mm larger than the maximal diameter, if the defect was <14 mm and >= 14 mm, respectively. The results of ASD closure in group I were compared with those of 271 patients (group II, median age 11 years) who underwent ASD closure with balloon sizing prior to the study period. Results: Of the 243 patients in group I, the maximal defect diameter ranged from 5.2 to 37 mm (mean 17.5 +/- 6.6 mm, median 17 mm). A total of 247 Amplatzer septal occluders were deployed in 240 patients. Two patients were found to develop distal embolization of a device the next day. Therefore, failure occurred in five patients. Comparing the results between group I and group II, there was no significant difference in success rate (238/243 vs. 263/271), incidence of embolization (2/243 vs. 2/271) and complete closure rate at 3-month follow-up (94.1% vs. 95.8%). There is significant difference in mean age (26.6 +/- 20.2 vs. 19.1 +/- 17.6), maximal defect diameter (17.5 +/- 6.6 vs. 14.1 +/- 5.9 mm) and Q(p)/Q(s) ratio (2.77 +/- 1.11 vs. 2.48 +/- 0.97) between group I and II. The mean diameter of device used was significantly larger in group I than in group II (23.1 +/- 8.1 vs. 19.6 +/- 7 mm, P < 0.001). Conclusions: Balloon sizing may not be necessary in transcatheter closure of ASD. (C) 2008 Wiley-Liss, Inc.
引用
收藏
页码:214 / 221
页数:8
相关论文
共 25 条
  • [1] Assessment of atrial septal defect size with 3D-transesophageal echocardiography:: Comparison with balloon method
    Abdel-Massih, T
    Dulac, Y
    Taktak, A
    Aggoun, Y
    Massabuau, P
    Elbaz, M
    Carrié, D
    Acar, P
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2005, 22 (02): : 121 - 127
  • [2] Amin Z, 2005, J AM COLL CARDIOL, V45, p317A
  • [3] Erosion of amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk
    Amin, Z
    Hijazi, ZM
    Bass, JL
    Cheatham, JP
    Hellenbrand, WE
    Kleinman, CS
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) : 496 - 502
  • [4] Transcatheter closure as standard treatment for most interatrial defects:: experience in 200 patients treated with the Amplatzer™ Septal Occluder
    Berger, F
    Ewert, P
    Björnstad, PG
    Dähnert, I
    Krings, G
    Brilla-Austgenat, I
    Vogel, M
    Lange, PE
    [J]. CARDIOLOGY IN THE YOUNG, 1999, 9 (05) : 468 - 473
  • [5] New echocardiographic diameter for amplatzer sizing in adult patients with secundum atrial septal defect: Preliminary results
    Carcagni, A
    Presbitero, P
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (03) : 409 - 414
  • [6] Transcatheter atrial septal defect closure: Modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer septal occluder
    Carlson, KM
    Justino, H
    O'Brien, RE
    Dimas, VV
    Leonard, GT
    Pignatelli, RH
    Mullins, CE
    Smith, EO
    Grifka, RG
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 66 (03) : 390 - 396
  • [7] Early and late complications associated with transcatheter occlusion of secundum atrial septal defect
    Chessa, M
    Carminati, M
    Butera, G
    Bini, RM
    Drago, M
    Rosti, L
    Giamberti, A
    Pomè, G
    Bossone, E
    Frigiola, A
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (06) : 1061 - 1065
  • [8] Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder
    Divekar, A
    Gaamangwe, T
    Shaikh, N
    Raabe, M
    Ducas, J
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (08) : 1213 - 1218
  • [9] DU ZD, 1944, J AM COLL CARDIOL, V39, P1936
  • [10] Du Zhong-Dong, 2002, J Interv Cardiol, V15, P287, DOI 10.1111/j.1540-8183.2002.tb01105.x