Pediatric intraocular lens power calculations

被引:16
|
作者
O'Hara, Mary A. [1 ,2 ]
机构
[1] Univ Calif Davis, Dept Ophthalmol, Sacramento, CA 95817 USA
[2] Univ Calif Davis, Dept Pediat, Sacramento, CA 95817 USA
关键词
intraocular lens power calculation; myopic shift; pediatric cataract surgery; pediatric intraocular lens implantation; AXIAL LENGTH; IMPLANTATION; ACCURACY; ERROR; KERATOMETRY; FORMULAS; OUTCOMES; EYES;
D O I
10.1097/ICU.0b013e32835622f8
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose of review To implant an appropriate intraocular lens (IOL) in a child, we must measure the eye well, calculate the IOL power accurately and predict the refractive change of the pseudophakic eye to maturity. The present review will concentrate on recent studies dealing with these issues. Recent findings Immersion A-scan biometry is superior in measuring the axial length of children. Current IOL power calculation formulas are very accurate in adults, but significantly less accurate in children. Several studies point to the high prediction errors encountered particularly in shorter eyes with all available IOL formulas. Postoperative refraction target remains controversial, but low degrees of overcorrection (i.e. hyperopia) may not adversely affect eventual best-corrected visual acuity. Summary Although pediatric IOL power calculations suffer from significant prediction error, these errors can be decreased by careful preoperative measurements. IOL power calculation formulas are most accurate in the older, more 'adult'-sized eye. The smallest eyes have the most prediction error with all available formulas. Individual circumstances and parental concerns must be factored into the choice of a postoperative refractive target.
引用
收藏
页码:388 / 393
页数:6
相关论文
共 50 条
  • [41] Intraocular lens calculations in patients with keratoectatic disorders
    Smith, Ryan G.
    Knezevic, Alexander
    Garg, Sumit
    CURRENT OPINION IN OPHTHALMOLOGY, 2020, 31 (04) : 284 - 287
  • [42] Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm
    Wang, Li
    Shirayama, Mariko
    Ma, Xingxuan Jack
    Kohnen, Thomas
    Koch, Douglas D.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (11): : 2018 - 2027
  • [43] Intraocular Lens Power Selection after Radial Keratotomy
    Geggel, Harry S.
    OPHTHALMOLOGY, 2015, 122 (05) : 897 - 902
  • [44] A comparative analysis of 12 intraocular lens power formulas
    Pereira, Austin
    Popovic, Marko M.
    Ahmed, Yusuf
    Lloyd, John C.
    El-Defrawy, Sherif
    Gorfinkel, John
    Schlenker, Matthew B.
    INTERNATIONAL OPHTHALMOLOGY, 2021, 41 (12) : 4137 - 4150
  • [45] Intraocular Lens Power Measured by Partial Coherence Interferometry
    Hsieh, Yi-Ting
    Wang, I-Jong
    OPTOMETRY AND VISION SCIENCE, 2012, 89 (12) : 1697 - 1701
  • [46] Intraocular lens power calculation in keratoconus; A review of literature
    Ghiasian, Leila
    Abolfathzadeh, Navid
    Manafi, Navid
    Hadavandkhani, Ali
    JOURNAL OF CURRENT OPHTHALMOLOGY, 2019, 31 (02): : 127 - 134
  • [47] Effects of Contact Lens Wear on Biometry Measurements for Intraocular Lens Calculations
    Meyer, Jay J.
    Kim, Michelle J.
    Kim, Terry
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2018, 44 : S255 - S258
  • [48] Fluctuations in corneal curvature limit predictability of intraocular lens power calculations
    Norrby, Sverker
    Hirnschall, Nino
    Nishi, Yutaro
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (02): : 174 - 179
  • [49] Intraocular lens power calculations in short eyes using 7 formulas
    Gokce, Sabite E.
    Zeiter, John H.
    Weikert, Mitchell P.
    Koch, Douglas D.
    Hill, Warren
    Wang, Li
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (07): : 892 - 897
  • [50] Accuracy of Intraocular Lens Power Calculation Formulas in Pediatric Cataract Patients: A Systematic Review and Meta-Analysis
    Zhong, Yueyang
    Yu, Yibo
    Li, Jinyu
    Lu, Bing
    Li, Su
    Zhu, Yanan
    FRONTIERS IN MEDICINE, 2021, 8