Noncontact Screening Methods for the Detection of Narrow Anterior Chamber Angles

被引:31
作者
Dabasia, Priya L. [1 ]
Edgar, David F. [1 ]
Murdoch, Ian E. [2 ]
Lawrenson, John G. [1 ]
机构
[1] City Univ London, London EC1V 0HB, England
[2] UCL Inst Ophthalmol, London, England
关键词
gonioscopy; van Herick test; advanced anterior segment imaging; diagnostic accuracy; angle closure glaucoma; OPTICAL COHERENCE TOMOGRAPHY; CLOSURE GLAUCOMA; OCCLUDABLE ANGLES; DEPTH MEASUREMENT; FLASHLIGHT TEST; CHINESE; GONIOSCOPY; PENTACAM; SPECIFICITY; SENSITIVITY;
D O I
10.1167/iovs.15-16727
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. Comparing diagnostic accuracy of biomicroscope techniques (van Herick and Smith's tests, evaluating limbal and central anterior chamber depth, respectively) and advanced imaging (Visante OCT and Pentacam) for detection of gonioscopically narrow anterior chamber angles (ACAs). METHODS. A total of 78 subjects with narrow or open ACAs underwent four index tests, performed on both eyes by examiners masked to other test results. Diagnostic performance was compared with gonioscopy, using International Society of Geographical and Epidemiological Ophthalmology (ISGEO) definition of primary angle closure and a classification based on clinical opinion of occludability. Data were analyzed using both the eye and the individual as unit of analysis. Sensitivity, specificity, and partial area under the receiver operating characteristic curve (AUROC) were generated. RESULTS. Using the eye as the unit of analysis, the van Herick grading cutoff of 25% or less and ISGEO gonioscopic classification achieved 80% (confidence interval [CI] 65 to 89) sensitivity and 92% specificity (CI 80 to 97) for narrow angle detection, with specificity reaching 97% (CI 87 to 100) for a cutoff of less than or equal to 15%. Notably, with a gonioscopic classification based on clinical opinion of occludability, van Herick (<= 25%) together with Smith's test (<= 2.50 mm) detected 100% of narrow angle subjects. Of the three Pentacam parameters, anterior chamber volume achieved highest test sensitivity of 85% (CI 70 to 94) using the ISGEO definition. Visante OCT ACA had greatest partial AUROC at 90% specificity, also yielding sensitivity and specificity greater than 85% using the Youden-derived cutoff of less than or equal to 20.7 degrees and ISGEO definition. CONCLUSIONS. Van Herick test and Visante OCT ACA exhibited best discrimination between narrow and open angles both alone, and in combination. Van Herick test affords advantages over Visante OCT, showing potential for identifying individuals who may benefit from further gonioscopic assessment in a case-finding or screening setting.
引用
收藏
页码:3929 / 3935
页数:7
相关论文
共 41 条
[1]  
Al-Mubrad Turki M., 2006, Indian Journal of Ophthalmology, V54, P165
[2]   Glaucoma anterior chamber morphometry based on optical Scheimpflug images [J].
Alonso, Ruiz Simonato ;
Ambrosio Junior, Renato ;
Paranhos Junior, Augusto ;
Sakaia, Lisandro Massanori ;
Ventura, Marcelo Palis .
ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2010, 73 (06) :497-500
[3]  
ALSBIRK PH, 1986, ACTA OPHTHALMOL, V64, P593
[4]   Comparing approaches to screening for angle closure in older Chinese adults [J].
Andrews, J. ;
Chang, D. S. ;
Jiang, Y. ;
He, M. ;
Foster, P. J. ;
Munoz, B. ;
Kashiwagi, K. ;
Friedman, D. S. .
EYE, 2012, 26 (01) :96-100
[5]   Anterior chamber depth measurement in clinical practice [J].
Barrett, BT ;
McGraw, PV ;
Murray, LA ;
Murgatroyd, P .
OPTOMETRY AND VISION SCIENCE, 1996, 73 (07) :482-486
[6]   Comparison of the scanning peripheral anterior chamber depth analyzer and the modified van Herick grading system in the assessment of angle closure [J].
Baskaran, Mani ;
Oen, Francis T. S. ;
Chan, Yiong-Huak ;
Hoh, Sek-Tien ;
Ho, Ching-Lin ;
Kashiwagi, Kenji ;
Foster, Pau J. ;
Aung, Tin .
OPHTHALMOLOGY, 2007, 114 (03) :501-506
[7]   Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD initiative [J].
Bossuyt, PM ;
Reitsma, JB ;
Bruns, DE ;
Gatsonis, CA ;
Glasziou, PP ;
Irwig, LM ;
Lijmer, JG ;
Moher, D ;
Rennie, D ;
de Vet, HCW .
ANNALS OF INTERNAL MEDICINE, 2003, 138 (01) :40-44
[8]   Single versus sequential testing with scanning peripheral anterior chamber depth analyser, IOLMaster and anterior segment optical coherence tomography for the detection of narrow angles [J].
Chang, Dolly S. ;
Sakata, Lisandro M. ;
Aung, Tin ;
He, Ming-Guang ;
Lavanya, Raghavan ;
Kashiwagi, Kenji ;
Friedman, David S. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2011, 95 (10) :1410-1414
[9]  
Congdon NG, 1996, ACTA OPHTHALMOL SCAN, V74, P113
[10]  
Devereux JG, 2000, ARCH OPHTHALMOL-CHIC, V118, P257