Safety and utility of computer-aided shoulder arthroplasty

被引:52
作者
Edwards, T. Bradley [1 ]
Gartsman, Gary M. [1 ]
O'Connor, Daniel P. [1 ]
Sarin, Vineet K. [2 ]
机构
[1] Texas Orthoped Hosp, Fondren Orthoped Grp, Houston, TX 77030 USA
[2] Kinamed Navigat Syst LLC, Camarillo, CA USA
关键词
D O I
10.1016/j.jse.2007.10.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study evaluated the safety and utility of a novel, image-free, shoulder navigation system in a cadaver and in on initial cohort of shoulder arthroplasty patients. Shoulder arthroplasty was performed on a cadaver and 27 patients using an image-free navigation system (NaviPro (TM); Kinamed Navigation Systems LLC, Camarillo, CA). Optical trackers were attached to the proximal humerus and the coracoid process. Prior to and following humeral head resection, the anatomic neck axis (retroversion, inclination) and humeral head diameter were measured with the navigation system. Native glenoid surface orientation was registered, and a navigation tracker was attached to the glenoid reamer. The navigation system recorded change in inclination and version relative to the native glenoid during reaming. The cadaver results demonstrated that the trackers did not impede surgical performance and that system accuracy was 2.6 degrees +/- 2.5 degrees. In the clinical series, the navigation system reported the anatomic humeral neck measurements (retroversion 30.0 degrees +/- 16.0 degrees; inclination 137.0 degrees +/- 11.7 degrees), the humeral head diameters (major axis 46.2 mm +/- 4.8 mm; minor axis 43.2 mm +/- 3.8 mm), the humeral neck resection angles (retroversion 29.9 degrees +/- 15.1 degrees and inclination 135.6 degrees +/- 9.1 degrees), and glenoid reaming orientation relative to the native glenoid (+3.0 degrees +/- 6.3 degrees of version; -6.7 degrees +/- 4.4 degrees of inclination). This initial clinical experience with computer-aided shoulder navigation demonstrates that the procedure is safe and can provide valuable intraoperative measurements. With an anatomic humeral implant system, the navigation system provides real-time feedback on the humeral resection as it relates to anatomic neck geometry. The system also provides realtime angulation of the glenoid reamer relative to preoperative glenoid deformity.
引用
收藏
页码:503 / 508
页数:6
相关论文
共 14 条
  • [1] Computer aided surgery system for shoulder prosthesis placement
    Atmani, H.
    Merienne, F.
    Fofi, D.
    Trouilloud, P.
    [J]. COMPUTER AIDED SURGERY, 2007, 12 (01) : 60 - 70
  • [2] Early experience with computer-assisted shoulder hemiarthroplasty for fractures of the proximal humerus: Development of a novel technique and an in vitro comparison with traditional methods
    Bicknell, Ryan T.
    DeLude, Jen A.
    Kedgley, Angela E.
    Ferreira, Louis M.
    Dunning, Cynthia E.
    King, Graham J. W.
    Faber, Ken J.
    Johnson, James A.
    Drosdowech, Darren S.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (03) : 117S - 125S
  • [3] Arthroplasty of the shoulder
    Boileau, P.
    Sinnerton, R. J.
    Chuinard, C.
    Walch, G.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (05): : 562 - 575
  • [4] Comparison of a mechanical acetabular alignment guide with computer placement of the socket
    DiGioia, AM
    Jaramaz, B
    Plakseychuk, AY
    Moody, JE
    Nikou, C
    LaBarca, RS
    Levison, TJ
    Picard, F
    [J]. JOURNAL OF ARTHROPLASTY, 2002, 17 (03) : 359 - 364
  • [5] The reverse shoulder prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency - A minimum two-year follow-up study of sixty patients
    Frankle, M
    Siegal, S
    Pupello, D
    Saleem, A
    Mighell, M
    Vasey, M
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) : 1697 - 1705
  • [6] THE USE OF COMPUTERIZED-TOMOGRAPHY IN THE MEASUREMENT OF GLENOID VERSION
    FRIEDMAN, RJ
    HAWTHORNE, KB
    GENEZ, BM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) : 1032 - 1037
  • [7] Computer-assisted navigation increases precision of component placement in total knee arthroplasty
    Haaker, RG
    Stockheim, M
    Kamp, M
    Proff, G
    Breitenfelder, J
    Ottersbach, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (433) : 152 - 159
  • [8] The effects of glenoid component alignment variations on cement mantle stresses in total shoulder arthroplasty
    Hopkins, AR
    Hansen, UN
    Amis, AA
    Emery, R
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2004, 13 (06) : 668 - 675
  • [9] Neer C.S., 1990, Shoulder Reconstruction
  • [10] Learning the vagaries of computer-assisted total knee replacement
    Sikorski, JM
    Blythe, MC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (07): : 903 - 910