Metal-on-metal bearings total hip arthroplasty: The cobalt and chromium ions release concern

被引:137
作者
Delaunay, C. [1 ]
Petit, I. [2 ]
Learmonth, I. D. [3 ]
Oger, P. [4 ]
Vendittoli, P. A. [5 ]
机构
[1] De Yvette Private Hosp, F-91160 Longjumeau, France
[2] Biomnis Biol Lab, F-94200 Ivry, France
[3] Southmead Hosp, Avon Orthopaed Ctr, Bristol Implant Res Ctr, Bristol BS10 5NB, Avon, England
[4] Versailles Hosp, Dept Orthopaed Surg, F-78150 Le Chesnay, France
[5] Univ Montreal, Maisonneuve Rosemont Hosp, Orthopaed Div, Montreal, PQ H1T 2M4, Canada
关键词
Total hip; arthroplasty; Metal-on-metal bearings; Hypersensitivity; Pseudotumors; Resurfacing; TOTAL KNEE ARTHROPLASTY; MINIMUM FOLLOW-UP; RESURFACING ARTHROPLASTY; WEAR DEBRIS; INFLAMMATORY PSEUDOTUMOR; SURFACE REPLACEMENT; PERIPHERAL-BLOOD; SERUM COBALT; IN-VITRO; OSTEOLYSIS;
D O I
10.1016/j.otsr.2010.05.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
With certain concerns recently reported on metal-on-metal bearing couples in total hip arthroplasty, this study's objective is to review the current knowledge concerning release of metal ions and its potential consequences. Each metal-on-metal implant presents different tribological properties. The analytical techniques for metals are accurate and the Co ion rates seem acceptable up to 2 mu g/L. A delayed type IV hypersensitivity reaction (atypical lymphocytic vasculitis-associated lesion [ALVAL]) may be the source of arthroplasty failure. Idiosyncratic, it remains unpredictable even using cutaneous tests and apparently is rare (0.3%). Today, there are no scientific or epidemiologic data supporting a risk of carcinogenesis or teratogenesis related to the use of a metal-on-metal bearings couple. Solid pseudotumors nearly exclusively are observed with resurfacing procedures, carrying a high annual revision rate in women under 40 years of age, occurring particularly in cases of acetabular malposition and with use of cast molded Cr-Co alloys. Osteolysis manifests through complete and progressive radiolucent lines or through cavitary lesions stemming from ALVAL-type alterations or impingement problems or implant incompatibility. The formation of wear debris exceeding the biological tolerance is possible with implant malposition, subluxation, and jamming of the femoral head in cases of cup deformity. Moreover, each implant presents different metal ion production; assessment of their performance and safety is required before their clinical use. With the knowledge available today, metal-on-metal bearing couples are contraindicated in cases of metal allergies or end stage renal dysfunction and small size resurfacing should cautiously be used. (C) 2010 Published by Elsevier Masson SAS.
引用
收藏
页码:894 / 904
页数:11
相关论文
共 92 条
[1]   Eleven years of experience with metal-on-metal hybrid hip resurfacing - A review of 1000 Conserve Plus [J].
Amstutz, Harlan C. ;
Le Duff, Michel J. .
JOURNAL OF ARTHROPLASTY, 2008, 23 (06) :36-43
[2]  
[Anonymous], 1979, Clin Orthop Relat Res
[3]   How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? [J].
Back, DL ;
Young, AA ;
Shimmin, AJ .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (438) :177-181
[4]   Cementless cups do not increase osteolysis risk in metal-on-metal total hip arthroplasty [J].
Beldame, J. ;
Carreras, F. ;
Oger, P. ;
Beaufils, P. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2009, 95 (07) :478-490
[5]   A benign psoas mass following metal-on-metal resurfacing of the hip [J].
Boardman, DR ;
Middleton, FR ;
Kavanagh, TG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (03) :402-404
[6]   Does the placenta inhibit the passage of chromium and cobalt after metal-on-metal total hip arthroplasty? [J].
Brodner, W ;
Grohs, JG ;
Bancher-Todesca, D ;
Dorotka, R ;
Meisinger, V ;
Gottsauner-Wolf, F ;
Kotz, R .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :102-106
[7]   Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[8]   Biological effects of clinically relevant wear particles from metal-on-metal hip prostheses [J].
Brown, C ;
Fisher, J ;
Ingham, E .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :355-369
[9]   Total hip wear debris presenting as lower extemity swelling [J].
Butler, RA ;
Barrack, RL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :142-145
[10]   Osteolysis in patients with a metal-on-metal hip arthroplasty [J].
Carr, Ashley M. ;
Desteiger, Richard .
ANZ JOURNAL OF SURGERY, 2008, 78 (03) :144-147