Comparison of Second-Generation Stents for Application in the Superficial Femoral Artery: An In Vitro Evaluation Focusing on Stent Design

被引:35
|
作者
Mueller-Huelsbeck, Stefan [1 ]
Schaefer, Philipp J. [2 ]
Charalambous, Nikolas [2 ]
Yagi, Hiroshi [3 ]
Heller, Martin [2 ]
Jahnke, Thomas [2 ]
机构
[1] Ev Luth Diakonissenanstalt Flensburg, Dept Diagnost & Intervent Radiol Neuroradiol, Acad Hosp Flensburg, D-24939 Flensburg, Germany
[2] Univ Hosp Schleswig Holstein, Dept Radiol, Kiel, Germany
[3] Terumo, Tokyo, Japan
关键词
superficial femoral artery; stent; nitinol; experimental study; finite element analysis; stent design; fatigue testing; stent fracture; compression; bending; torsion; NITINOL STENT; IMPLANTATION; FRACTURES; LESIONS; QUANTIFICATION; ANGIOPLASTY; DISEASE; TRIAL;
D O I
10.1583/10-3069.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To examine and compare in an ex vivo study different nitinol stent designs intended for the superficial femoral artery (SFA) with regard to the appearance of fracture. Methods: Seven different 8-x40-mm nitinol stents were evaluated (Misago, Absolute, Smart, Luminexx, Sentinol, Lifestent NT, and Sinus-Superflex). Finite element analysis (FEA) was used for digitalized stent design comparison; the strain during stent movement was calculated for bending, compression, and torsion. Additional mechanical fatigue tests for bending (70 degrees), compression (40%), and torsion (twisted counterclockwise by 180 degrees) were performed up to 650,000 cycles or until a fracture was observed. Results:The FEA bending test showed that only the Misago, LifeStent, and Absolute stents presented no zones of high strain; in the torsion test, the Smart stent also had no zones of high strain. Macroscopic evaluation after mechanical bending indicated that the LifeStent performed the best (no stent fracture after 650,000 cycles). Misago and Absolute stents showed fractures at 536,000 cycles and 456,667 cycles, respectively (range 320,000-650,000 cycles). After compression and torsion testing, Misago showed no stent fracture after 650,000 cycles. The worst performing stent was Luminexx during all test cycles. Conclusion: The 7 SFA stents showed differences in the incidence of high strain zones, which indicates a potential for stent fracture, as demonstrated by the mechanical fatigue tests. Differences in stent design might play a major role in the appearance of stent strut fracture related to restenosis and reocclusion. J Endovasc Ther. 2010;17:767-776
引用
收藏
页码:767 / 776
页数:10
相关论文
共 50 条
  • [21] Randomized comparison of ePTFE/nitinol self-expanding stent graft vs prosthetic femoral-popliteal bypass in the treatment of superficial femoral artery occlusive disease
    McQuade, Karen
    Gable, Dennis
    Holman, Stephen
    Pearl, Greg
    Theune, Brian
    JOURNAL OF VASCULAR SURGERY, 2009, 49 (01) : 109 - 116
  • [22] Comparison of First and Second-generation Drug-eluting Stents in Patients Presenting With Anemia
    Wanha, Wojciech
    Pluta, Aleksandra
    Marcinkiewicz, Kamil
    Kawecki, Damian
    Ochala, Andrzej
    Tendera, Michal
    Wojakowski, Wojciech
    CIRCULATION, 2013, 128 (22)
  • [23] Comparison of Optical Coherence Tomographic Assessment between First- and Second-Generation Drug-Eluting Stents
    Kim, Byeong-Keuk
    Kim, Jung-Sun
    Park, Junbeom
    Ko, Young-Guk
    Choi, Donghoon
    Jang, Yangsoo
    Hong, Myeong-Ki
    YONSEI MEDICAL JOURNAL, 2012, 53 (03) : 524 - 529
  • [24] Impact of Paclitaxel-Eluting Balloons Compared to Second-Generation Drug-Eluting Stents for of In-Stent Restenosis in a Primarily Acute Coronary Syndrome Population
    Marquis-Gravel, Guillaume
    Matteau, Alexis
    Potter, Brian J.
    Gobeil, Francois
    Noiseux, Nicolas
    Stevens, Louis-Mathieu
    Mansour, Samer
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2017, 109 (04) : 277 - 283
  • [25] Comparison of clinical outcomes between first-generation and second-generation drug-eluting stents in type 2 diabetic patients
    Jeong, Han Saem
    Cho, Jae Young
    Kim, Eun Ji
    Yu, Cheol Woong
    Ahn, Chul-Min
    Park, Jae Hyoung
    Hong, Soon Jun
    Lim, Do-Sun
    CORONARY ARTERY DISEASE, 2013, 24 (08) : 676 - 683
  • [26] Characteristics of early versus late in-stent restenosis in second-generation drug-eluting stents: an optical coherence tomography study
    Song, Lei
    Mintz, Gary S.
    Yin, Dong
    Yamamoto, Myong Hwa
    Chin, Chee Yang
    Matsumura, Mitsuaki
    Kirtane, Ajay J.
    Parikh, Manish A.
    Moses, Jeffrey W.
    Ali, Ziad A.
    Shlofmitz, Richard A.
    Maehara, Akiko
    EUROINTERVENTION, 2017, 13 (03) : 294 - 302
  • [27] Comparison of in-stent neoatherosclerosis and tissue characteristics between early and late in-stent restenosis in second-generation drug-eluting stents: an optical coherence tomography study
    Sabbah, Mahmoud
    Kadota, Kazushige
    El-Eraky, Azza
    Kamal, Hanan M.
    Abdellah, Ahmed-Tageldien
    El Hawary, Ahmed
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2017, 33 (10) : 1463 - 1472
  • [28] Treatment of drug-eluting stent restenosis: Comparison between drug-eluting balloon versus second-generation drug-eluting stents from a retrospective observational study
    Basavarajaiah, Sandeep
    Naganuma, Toru
    Latib, Azeem
    Sticchi, Alessandro
    Ciconte, Giuseppe
    Panoulas, Vasilis
    Chieffo, Alaide
    Montorfano, Matteo
    Carlino, Mauro
    Colombo, Antonio
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2016, 88 (04) : 522 - 528
  • [29] Safety and Effectiveness of Second-Generation Drug-Eluting Stents in Patients With Left Main Coronary Artery Disease
    Lee, Pil Hyung
    Kwon, Osung
    Ahn, Jung-Min
    Lee, Cheol Hyun
    Kang, Do-Yoon
    Lee, Jung-Bok
    Kang, Soo-Jin
    Lee, Seung-Whan
    Kim, Young-Hak
    Lee, Cheol Whan
    Park, Seong-Wook
    Park, Duk-Woo
    Park, Seung-Jung
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (08) : 832 - 841
  • [30] Mechanisms and Patterns of Intravascular Ultrasound In-Stent Restenosis Among Bare Metal Stents and First- and Second-Generation Drug-Eluting Stents
    Goto, Kosaku
    Zhao, Zhijing
    Matsumura, Mitsuaki
    Dohi, Tomotaka
    Kobayashi, Nobuaki
    Kirtane, Ajay J.
    Rabbani, LeRoy E.
    Collins, Michael B.
    Parikh, Manish A.
    Kodali, Susheel K.
    Leon, Martin B.
    Moses, Jeffrey W.
    Mintz, Gary S.
    Maehara, Akiko
    AMERICAN JOURNAL OF CARDIOLOGY, 2015, 116 (09) : 1351 - 1357