Integrated wearable and self-carrying active upper limb orthosis

被引:14
作者
Merchant, Roberto [1 ]
Cruz-Ortiz, David [1 ]
Ballesteros-Escamilla, Mariana [1 ]
Chairez, Isaac [1 ]
机构
[1] Inst Politecn Nacl UPIBI, Av Acueducto S-N, Ciudad De Mexico 07340, Mexico
关键词
Orthosis; upper limb pathologies; sliding-mode control; terminal sliding mode; wearable technology; SLIDING-MODE CONTROL; SYSTEMS; REHABILITATION;
D O I
10.1177/0954411917751001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this study was to develop a prototype of an orthotic system that can be used as a support tool in the rehabilitation of the upper limb. The construction of this device was motivated by the increasing number of subjects suffering from full or partial loss of the upper limb function as a consequence of spinal cord injuries, strokes, occupational syndromes and sports injuries. The majority of procedures used in upper limb rehabilitation consist of repetitive movements enforced by physiotherapists; a robotic device executing the same tasks seems to be a plausible solution if the orthosis can be programmed and controlled automatically. This study reports the mechanical design, electronic instrumentation and automatic control of an upper limb orthosis made of plastic polymer that makes the orthosis a wearable and self-carrying device. The orthosis consisted of a mechatronic device with five joints. The pieces made by a three-dimensional plastic printer were used to construct the device leading to a total weight of 2.6kg. The application of a robust automatic controller based on the sliding-mode theory forces the movement of the arm, while taking into account the constraints in each angular displacement of the orthosis. A set of reference trajectories designed to represent the usual movements of a healthy upper limb served for evaluating the controller execution. The orthosis was tested on 15 volunteers with a maximum experimental steady-state error of 2% in the angular deviation of all articulations with respect to their reference trajectories.
引用
收藏
页码:172 / 184
页数:13
相关论文
共 25 条
[1]   Review of arm motion analyses [J].
Anglin, C ;
Wyss, UP .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H5) :541-555
[2]  
[Anonymous], 2014, Attractive Ellipsoids in Robust Control, DOI DOI 10.1007/978-3-319-09210-2
[3]  
[Anonymous], 2006, ROBOT MODELING CONTR
[4]   Sterilization, toxicity, biocompatibility and clinical applications of polylactic acid polyglycolic acid copolymers [J].
Athanasiou, KA ;
Niederauer, GG ;
Agrawal, CM .
BIOMATERIALS, 1996, 17 (02) :93-102
[5]   3D printing: the dawn of a new era in manufacturing? [J].
Bogue, Robert .
ASSEMBLY AUTOMATION, 2013, 33 (04) :307-311
[6]  
DRILLIS R.J., 1966, BODY SEGMENT PARAMET
[7]   Non-singular terminal sliding mode control of rigid manipulators [J].
Feng, Y ;
Yu, XH ;
Man, ZH .
AUTOMATICA, 2002, 38 (12) :2159-2167
[8]  
Gopura R. A. R. C., 2009, 2009 IEEE International Conference on Rehabilitation Robotics: Reaching Users & the Community (ICORR), P178, DOI 10.1109/ICORR.2009.5209630
[9]  
Gordon Claire C., 1989, TECHNICAL REPORT
[10]   Current Hand Exoskeleton Technologies for Rehabilitation and Assistive Engineering [J].
Heo, Piwon ;
Gu, Gwang Min ;
Lee, Soo-Jin ;
Rhee, Kyehan ;
Kim, Jung .
INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2012, 13 (05) :807-824