Note: Curve fit models for atomic force microscopy cantilever calibration in water

被引:2
|
作者
Kennedy, Scott J. [1 ]
Cole, Daniel G. [2 ]
Clark, Robert L. [3 ]
机构
[1] Duke Univ, Dept Mech Engn & Mat Sci, Ctr Biol Inspired Mat & Mat Syst, Durham, NC 27708 USA
[2] Univ Pittsburgh, Dept Mech Engn & Mat Sci, Swanson Sch Engn, Pittsburgh, PA 15260 USA
[3] Univ Rochester, Hajim Sch Engn & Appl Sci, Rochester, NY 14627 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2011年 / 82卷 / 11期
基金
美国国家科学基金会;
关键词
D O I
10.1063/1.3661130
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Atomic force microscopy stiffness calibrations performed on commercial instruments using the thermal noise method on the same cantilever in both air and water can vary by as much as 20% when a simple harmonic oscillator model and white noise are used in curve fitting. In this note, several fitting strategies are described that reduce this difference to about 11%. (C) 2011 American Institute of Physics. [doi: 10.1063/1.3661130]
引用
收藏
页数:3
相关论文
共 50 条
  • [21] Calibration-Independent Atomic Force Microscopy
    Corredera, Carmen Suay
    Pimenta-Lopes, Carolina
    Velazquez-Carreras, Diana
    Sanchez-Ortiz, David
    Alegre-Cebollada, Jorge
    BIOPHYSICAL JOURNAL, 2019, 116 (03) : 428A - 428A
  • [22] Calibration of frictional forces in atomic force microscopy
    Ogletree, DF
    Carpick, RW
    Salmeron, M
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (09): : 3298 - 3306
  • [23] Lateral force microscope calibration using a modified atomic force microscope cantilever
    Reitsma, M. G.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (10):
  • [24] An improved calibration method for friction force in atomic force microscopy
    Tribology Research Institute, National Traction Power Laboratory, Southwest Jiaotong University, Chengdu 610031, China
    Mocaxue Xuebao/Tribology, 2007, 27 (05): : 472 - 476
  • [25] Traceable Lateral Force Calibration (TLFC) for Atomic Force Microscopy
    Bhattacharjee, Arnab
    Garabedian, Nikolay T.
    Evans, Christopher L.
    Burris, David L.
    TRIBOLOGY LETTERS, 2020, 68 (04)
  • [26] A theoretical model for the cantilever motion in contact-resonance atomic force microscopy and its application to phase calibration in piezoresponse force and electrochemical strain microscopy
    Bradler, Stephan
    Kachel, Stefan Renato
    Schirmeisen, Andre
    Roling, Bernhard
    JOURNAL OF APPLIED PHYSICS, 2016, 120 (16)
  • [27] Traceable Lateral Force Calibration (TLFC) for Atomic Force Microscopy
    Arnab Bhattacharjee
    Nikolay T. Garabedian
    Christopher L. Evans
    David L. Burris
    Tribology Letters, 2020, 68
  • [28] Micro-fabricated piezoelectric cantilever for atomic force microscopy
    Watanabe, S
    Fujii, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (11): : 3898 - 3903
  • [29] WHOLE ELECTRONIC CANTILEVER CONTROL FOR ATOMIC-FORCE MICROSCOPY
    TAKATA, K
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (09): : 2598 - 2600
  • [30] Massively parallel cantilever-free atomic force microscopy
    Wenhan Cao
    Nourin Alsharif
    Zhongjie Huang
    Alice E. White
    YuHuang Wang
    Keith A. Brown
    Nature Communications, 12