A Simulation Study to Estimate Optimum LOR Angular Acceptance for the Image Reconstruction with the Total-Body J-PET

被引:6
|
作者
Dadgar, Meysam [1 ,2 ]
Parzych, Szymon [1 ,2 ]
Ardebili, Faranak Tayefi [1 ,2 ]
机构
[1] Jagiellonian Univ, Marian Smoluchowski Inst Phys, Krakow, Poland
[2] Jagiellonian Univ, Total Body Jagiellonian PET Lab, Krakow, Poland
来源
MEDICAL IMAGE UNDERSTANDING AND ANALYSIS (MIUA 2021) | 2021年 / 12722卷
关键词
Acceptance angle; Total Body J-PET; Sensitivity; Spatial resolution; PLASTIC SCINTILLATOR; GATE;
D O I
10.1007/978-3-030-80432-9_15
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
One of the directions in today's development of PET scanners is to increase their axial field of view (AFOV). Currently limited to several centimeters, AFOV of the clinically available PET tomographs results in a very low sensitivity (similar to 1%) and requires an extended time for a scan of a whole human body. While these drawbacks are addressed in the so-called, Total Body PET concept (scanner with a significantly elongated field of view), it creates new challenges not only in the mechanical construction but also in the image reconstruction and event selection. The possibility of taking into account of large angle variety of lines of responses (LORs) contributes positively to the sensitivity of the tomograph. However, at the same time, the most oblique LORs have an unfavorable influence on the spatial resolution due to the parallax error and large contribution to the scatter fraction. This forces to determine a new factor - acceptance angle - which is a maximum azimuthal angle for which the LORs are still taken into image reconstruction. Correct determination of such factor is imperative to maximize the performance of a Total Body PET system since it introduces a trade-off between the two main characteristics of scanners: sensitivity and spatial resolution. This work has been dedicated to the estimation of the optimal acceptance angle for the proposed by the Jagiellonian PET (J-PET) Collaboration Total Body tomograph. J-PET Collaboration introduces a novel, cost-effective approach to PET systems development with the use of organic scintillators. This simulation study provides evidence that the 45 degrees acceptance angle cut can be an appropriate choice for the investigated scanner.
引用
收藏
页码:189 / 200
页数:12
相关论文
共 24 条
  • [1] RECONSTRUCTION OF THE NEMA IEC BODY PHANTOM FROM J-PET TOTAL-BODY SCANNER SIMULATION USING STIR
    Kopka, P.
    Klimaszewski, K.
    ACTA PHYSICA POLONICA B, 2020, 51 (01): : 357 - 360
  • [2] Realistic total-body J-PET geometry optimization: Monte Carlo study
    Baran, Jakub
    Krzemien, Wojciech
    Parzych, Szymon
    Raczynski, Lech
    Bala, Mateusz
    Coussat, Aurelien
    Chug, Neha
    Czerwinski, Eryk
    Curceanu, Catalina Oana
    Dadgar, Meysam
    Dulski, Kamil
    Eliyan, Kavya
    Gajewski, Jan
    Gajos, Aleksander
    Hiesmayr, Beatrix C.
    Kacprzak, Krzysztof
    Kaplon, Lukasz
    Klimaszewski, Konrad
    Korcyl, Grzegorz
    Kozik, Tomasz
    Kumar, Deepak
    Niedzwiecki, Szymon
    Panek, Dominik
    del Rio, Elena Perez
    Rucinski, Antoni
    Sharma, Sushil
    Shivani, Roman Y.
    Shopa, Roman Y.
    Skurzok, Magdalena
    Stepien, Ewa
    Tayefiardebili, Faranak
    Tayefiardebili, Keyvan
    Wislicki, Wojciech
    Moskal, Pawel
    MEDICAL PHYSICS, 2025,
  • [3] SYNTHESIS AND CHARACTERIZATION OF PLASTIC SCINTILLATORS FOR THE TOTAL-BODY J-PET SCANNER
    Kaplon, L.
    ACTA PHYSICA POLONICA B, 2020, 51 (01): : 225 - 230
  • [4] ESTIMATION OF 511 keV GAMMA SCATTER FRACTION IN WLS LAYER IN TOTAL-BODY J-PET; A SIMULATION STUDY
    Ardebili, K. Tayefi
    Niedzwiecki, S.
    Moskal, P.
    ACTA PHYSICA POLONICA B PROCEEDINGS SUPPLEMENT, 2022, 15 (04)
  • [5] Simulating NEMA characteristics of the modular total-body J-PET scanner-an economic total-body PET from plastic scintillators
    Moskal, P.
    Kowalski, P.
    Shopa, R. Y.
    Raczynski, L.
    Baran, J.
    Chug, N.
    Curceanu, C.
    Czerwinski, E.
    Dadgar, M.
    Dulski, K.
    Gajos, A.
    Hiesmayr, B. C.
    Kacprzak, K.
    Kaplon, L.
    Kisielewska, D.
    Klimaszewski, K.
    Kopka, P.
    Korcyl, G.
    Krawczyk, N.
    Krzemien, W.
    Kubicz, E.
    Niedzwiecki, Sz
    Parzych, Sz
    Raj, J.
    Sharma, S.
    Shivani, S.
    Stepien, E.
    Tayefi, F.
    Wislicki, W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (17):
  • [6] Investigation of Novel Preclinical Total Body PET Designed With J-PET Technology: A Simulation Study
    Dadgar, M.
    Parzych, S.
    Tayefi Ardebili, F.
    Baran, J.
    Chug, N.
    Curceanu, C.
    Czerwinski, E.
    Dulski, K.
    Eliyan, K.
    Gajos, A.
    Hiesmayr, B. C.
    Kacprzak, K.
    Kaplon, L.
    Klimaszewski, K.
    Konieczka, P.
    Korcyl, G.
    Kozik, T.
    Krzemien, W.
    Kumar, D.
    Niedzwiecki, S.
    Panek, D.
    Perez del Rio, E.
    Raczynski, L.
    Sharma, S.
    Shopa, R. Y.
    Skurzok, M.
    Tayefi Ardebili, K.
    Vandenberghe, S.
    Wislicki, W.
    Stepien, E. L.
    Moskal, P.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (02) : 124 - 131
  • [7] Optimization of the WLS design for positron emission mammography and Total-Body J-PET systems
    Georgadze, Anzori
    Shivani
    Ardebili, Keyvan Tayefi
    Moskal, Pawel
    BIO-ALGORITHMS AND MED-SYSTEMS, 2023, 19 (01) : 114 - 123
  • [8] New Challenges for PET Image Reconstruction for Total-Body Imaging
    Efthimiou, Nikos
    PET CLINICS, 2020, 15 (04) : 453 - 461
  • [9] Technical Attenuation Length Measurement of Plastic Scintillator Strips for the Total-Body J-PET Scanner
    Kaplon, Ukasz
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2020, 67 (10) : 2286 - 2289
  • [10] TOF MLEM Adaptation for the Total-Body J-PET With a Realistic Analytical System Response Matrix
    Shopa, R. Y.
    Baran, J.
    Klimaszewski, K.
    Krzemien, W.
    Raczynski, L.
    Wislicki, W.
    Brzezinski, K.
    Chug, N.
    Coussat, A.
    Curceanu, C.
    Czerwinski, E.
    Dadgar, M.
    Dulski, K.
    Gajewski, J.
    Gajos, A.
    Hiesmayr, B. C.
    Valsan, E. Kavya
    Korcyl, G.
    Kozik, T.
    Kumar, D.
    Kaplon, L.
    Moskal, G.
    Niedzwiecki, S.
    Panek, D.
    Parzych, S.
    Rio, E. Perez del
    Rucinski, A.
    Sharma, S.
    Shivani, S.
    Silarski, M.
    Skurzok, M.
    Stepien, E.
    Ardebili, F. Tayefi
    Ardebili, K. Tayefi
    Moskal, P.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (05) : 509 - 520