Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry

被引:8
作者
Baek, Sung Uk [1 ,2 ]
Ha, Ahnul [1 ,2 ]
Kim, Young Kook [1 ,2 ]
Jeoung, Jin Wook [1 ,2 ]
Park, Ki Ho [1 ,2 ]
机构
[1] Seoul Natl Univ, Dept Ophthalmol, Coll Med, Seoul, South Korea
[2] Seoul Natl Univ Hosp, Dept Ophthalmol, Seoul, South Korea
关键词
glaucoma; intraocular pressure; eye lids; GOLDMANN APPLANATION TONOMETER;
D O I
10.1136/bjophthalmol-2017-311587
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background/aims To investigate the effect of eyelid manipulation on the measurement of intraocular pressure (IOP) using two different tonometries (rebound tonometry (RT) vs Goldmann applanation tonometry (GAT)). Methods 103 patients with primary open-angle glaucoma were prospectively enrolled. For all of the patients, IOP measurements were performed in three different ways: (1) RT with lid manipulation (LM), (2) RT without LM and (3) GAT. The order of the three measurements was randomly selected. Additionally, the palpebral fissure height (PFH; elliptical space between upper and lower eyelids) was measured. Results The mean value of IOP measured by GAT was 13.972.80mm Hg, which was not significantly different from that by RT without LM (13.75 +/- 2.44mm Hg; P=0.096), but which was significantly lower than that by RT with LM (15.21 +/- 2.91mm Hg; P<0.001). On a Bland-Altman plot, RT with LM was overestimated relative to GAT (mean: -1.5) and RT without LM (mean: -1.2). Among the high IOPs (>20mm Hg), interestingly, those measured by RT without LM were significantly lower than those measured by GAT (P<0.001). In the subgroup analysis of PFH, the smaller the PFH, the more exaggerated the IOP difference between GAT (P=0.014) and RT with LM (P<0.001). Conclusion RT-measured IOP was significantly exaggerated when manipulation was applied to the eyelid. This overall trend was more pronounced when PFH was small. GAT-measured IOP, meanwhile, showed a good correlation with IOP measured using RT without LM.
引用
收藏
页码:1515 / 1519
页数:5
相关论文
共 50 条
  • [41] The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry
    Avitabile, Teresio
    Longo, Antonio
    Rocca, Daniela
    Amato, Roberta
    Gagliano, Caterina
    Castaing, Marine
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (04) : 585 - 591
  • [42] Applanation resonance tonometry for intraocular pressure in humans
    Hallberg, P
    Lindén, C
    Lindahl, OA
    Bäcklund, T
    Eklund, A
    PHYSIOLOGICAL MEASUREMENT, 2004, 25 (04) : 1053 - 1065
  • [43] Rebound tonometry intraocular pressure reference values for conscious adult Egyptian vultures (Neophron percnopterus)
    Prieto, Lucia Gomez
    Byron-Chance, Dominic
    Hollwarth, Ashton J.
    Dutton, Thomas A. G.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2025, 86 (04)
  • [44] Determination of Tear Production and Intraocular Pressure With Rebound Tonometry in Wild Humboldt Penguins (Spheniscus humboldti)
    Sheldon, Julie D.
    Adkesson, Michael J.
    Allender, Matthew C.
    Jankowski, Gwen
    Langan, Jennifer
    Cardena, Marco
    Cardenas-Alayza, Susana
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2017, 31 (01) : 16 - 23
  • [45] Intraocular Pressure in Captive Black-footed Penguins (Spheniscus demersus) Measured by Rebound Tonometry
    Mercado, Julio A.
    Wirtu, Gemechu
    Beaufrere, Hugues
    Lydick, Dianna
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2010, 24 (02) : 138 - 141
  • [46] Modeling and Simulation of Human Cornea for the Measurement of Intraocular Pressure (IOP) through Eyelid
    Mawale, Mahesh B.
    Kuthe, Abhaykumar
    Pawane, Padma G.
    Dahake, Sandeep W.
    CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES, 2018, 114 (02): : 173 - 189
  • [47] Rebound Tonometer: Ideal Tonometer for Measurement of Accurate Intraocular Pressure
    Suman, Suwarna
    Agrawal, Ajai
    Pal, Virendra K.
    Pratap, Vir B.
    JOURNAL OF GLAUCOMA, 2014, 23 (09) : 633 - 637
  • [48] Evaluation of the Effect of Body Position on Intraocular Pressure Measured with Rebound Tonometer
    Mayali, Huseyin
    Tekin, Beyza
    Kayikcioglu, Ozcan Rasim
    Kurt, Emin
    Ilker, Suleyman Sami
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2019, 49 (01): : 6 - 9
  • [49] Twenty-four hour intraocular pressure measurements and home tonometry
    Meier-Gibbons, Frances
    Berlin, Michael S.
    Toteberg-Harms, Marc
    CURRENT OPINION IN OPHTHALMOLOGY, 2018, 29 (02) : 111 - 115
  • [50] Effect of Repeated Applanation Tonometry on the Accuracy of Intraocular Pressure Measurements
    Ehrenberg, M.
    Gaton, D.
    Lusky, M.
    Dotan, G.
    Weinberger, D.
    Snir, M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)