Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry

被引:8
|
作者
Baek, Sung Uk [1 ,2 ]
Ha, Ahnul [1 ,2 ]
Kim, Young Kook [1 ,2 ]
Jeoung, Jin Wook [1 ,2 ]
Park, Ki Ho [1 ,2 ]
机构
[1] Seoul Natl Univ, Dept Ophthalmol, Coll Med, Seoul, South Korea
[2] Seoul Natl Univ Hosp, Dept Ophthalmol, Seoul, South Korea
关键词
glaucoma; intraocular pressure; eye lids; GOLDMANN APPLANATION TONOMETER;
D O I
10.1136/bjophthalmol-2017-311587
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background/aims To investigate the effect of eyelid manipulation on the measurement of intraocular pressure (IOP) using two different tonometries (rebound tonometry (RT) vs Goldmann applanation tonometry (GAT)). Methods 103 patients with primary open-angle glaucoma were prospectively enrolled. For all of the patients, IOP measurements were performed in three different ways: (1) RT with lid manipulation (LM), (2) RT without LM and (3) GAT. The order of the three measurements was randomly selected. Additionally, the palpebral fissure height (PFH; elliptical space between upper and lower eyelids) was measured. Results The mean value of IOP measured by GAT was 13.972.80mm Hg, which was not significantly different from that by RT without LM (13.75 +/- 2.44mm Hg; P=0.096), but which was significantly lower than that by RT with LM (15.21 +/- 2.91mm Hg; P<0.001). On a Bland-Altman plot, RT with LM was overestimated relative to GAT (mean: -1.5) and RT without LM (mean: -1.2). Among the high IOPs (>20mm Hg), interestingly, those measured by RT without LM were significantly lower than those measured by GAT (P<0.001). In the subgroup analysis of PFH, the smaller the PFH, the more exaggerated the IOP difference between GAT (P=0.014) and RT with LM (P<0.001). Conclusion RT-measured IOP was significantly exaggerated when manipulation was applied to the eyelid. This overall trend was more pronounced when PFH was small. GAT-measured IOP, meanwhile, showed a good correlation with IOP measured using RT without LM.
引用
收藏
页码:1515 / 1519
页数:5
相关论文
共 50 条
  • [31] INTRAOCULAR PRESSURE IN CAPTIVE AMERICAN FLAMINGOS (PHOENICOPTERUS RUBER) AS MEASURED BY REBOUND TONOMETRY
    Molter, Christine M.
    Hollingsworth, Steven R.
    Kass, Philip H.
    Chinnadurai, Sathya K.
    Wack, Raymund F.
    JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2014, 45 (03) : 664 - 667
  • [32] Intraocular pressure measurement - Comparison of dynamic contour tonometry and Goldmann applanation tonometry
    Schneider, E
    Grehn, F
    JOURNAL OF GLAUCOMA, 2006, 15 (01) : 2 - 6
  • [33] Comparison of intraocular pressure measurement by Dynamic Contour Tonometry and Goldmann applanation tonometry
    DeBry, PW
    Krishna, R
    Willoughby, TL
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U372 - U372
  • [34] Evaluation of intraocular pressure in conscious ferrets (Mustela putorius furo) by means of rebound tonometry and comparison with applanation tonometry
    Di Girolamo, N.
    Andreani, V.
    Guandalini, A.
    Selleri, P.
    VETERINARY RECORD, 2013, 172 (15) : 396 - +
  • [35] Measuring Intraocular Pressure After Intrastromal Corneal Ring Segment Implantation With Rebound Tonometry and Goldmann Applanation Tonometry
    Arribas-Pardo, Paula
    Mendez-Hernandez, Carmen
    Cuina-Sardina, Ricardo
    Fernandez-Perez, Cristina
    Diaz-Valle, David
    Garcia-Feijoo, Julian
    CORNEA, 2015, 34 (05) : 516 - 520
  • [36] Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients
    Vincent, Stephen J.
    Vincent, Roslyn A.
    Shields, David
    Lee, Graham A.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (04): : E163 - E170
  • [37] Comparison of intraocular pressure measurements in healthy subjects taken by rebound tonometry, dynamic contour tonometry and applanation tonometry at a hospital in Mexico City
    Perez Gudino, Itzel Montserrat
    Escalante-Castanon, Mariana
    Hernandez-Oteyza, Alejandra
    Jimenez-Arroyo, Jesus
    Garcia-Huerta, Magdalena
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [38] Does Rebound Tonometry Probe Misalignment Modify Intraocular Pressure Measurements in Human Eyes?
    Beasley, Ian G.
    Laughton, Deborah S.
    Coldrick, Benjamin J.
    Drew, Thomas E.
    Sallah, Marium
    Davies, Leon N.
    JOURNAL OF OPHTHALMOLOGY, 2013, 2013
  • [39] INTRAOCULAR PRESSURE IN CLINICALLY NORMAL BROOK TROUT (SALVELINUS FONTINALIS) BY MEANS OF REBOUND TONOMETRY
    Keeney, Caitlin Hepps
    Vorbach, Bryan
    Clayton, Leigh
    Seeley, Kathryn
    JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2019, 50 (01) : 107 - 110
  • [40] Comparison of Intraocular Pressure Measurements Obtained by Rebound, Noncontact, and Goldmann Applanation Tonometry in Children
    Vodencarevic, Amra Nadarevic
    Jusufovic, Vahid
    Terzic, Svjetlana
    Halilbasic, Meliha
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 163 : 192 - 192