Chemiluminescence flow system for the determination of sulfite

被引:17
|
作者
Qin, W
Zhang, ZJ [1 ]
Zhang, CJ
机构
[1] Shaanxi Normal Univ, Dept Chem, Xian 710062, Peoples R China
[2] Nanjing Univ, Dept Chem, Nanjing 210008, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
D O I
10.1007/s002160051023
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel chemiluminescence(CL) flow system for sulfite is described based on electrostatically immobilized luminol on an anion exchange column. Sulfite is detected by the CL reaction with luminol bleeding from the column by hydrolysis. The calibration graph is linear in the range 3 x 10(-7) to 1 x 10(-5) mol/L, and the detection limit is 1 x 10(-7) mol/L. Interfering metal ions co-existing ill sample solutions could be effectively eliminated on-line by an upstream cation exchanger. A complete analysis could be performed in 1 min with a relative standard deviation of less than 5%. The system could be reused for over 50 h and has been applied successfully to the determination of sulfur dioxide in air.
引用
收藏
页码:824 / 826
页数:3
相关论文
共 50 条