Healing Abutment Distortion in Implant Prostheses: An In Vitro Study

被引:1
|
作者
Pandey, Chandrashekhar [1 ]
Bhattarai, Bishwa Prakash [1 ]
Riddhabhaya, Apiwat [2 ]
Wongsirichat, Natthamet [1 ]
Rokaya, Dinesh [1 ]
机构
[1] Walailak Univ, Int Coll Dent, Dept Clin Dent, Bangkok 10400, Thailand
[2] Walailak Univ, Int Coll Dent, Dept Oral Biosci, Bangkok 10400, Thailand
关键词
dental implants; implant abutment; healing abutment; screw thread; screw head; wear; distortion; scanning electron microscope; TORQUE; PRELOAD; REMOVAL; CYCLES; SCREWS; CONTAMINATION; MAINTENANCE; MISFIT; REUSE;
D O I
10.3390/jfb13030085
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Little has been studied regarding the repeated usage of healing abutments and their effects on the distortion of dental implant-healing abutment hex surfaces. Additionally, implant manufacturers do not provide specific guidelines on how many times a healing abutment can be used before discarding. Therefore, we evaluated the effect of repeated screwing-unscrewing of implant-healing abutment on the healing abutment hex surface and screw head. A total of 12 Biomate implants with 4 mm diameter and 13 mm length were inserted into a synthetic bone block. The standard healing abutments of 3 mm diameter and 4 mm length were screwed onto each implant using a torque ratchet at a final torque of 30 Ncm. Immediately, the abutments were unscrewed at 30 Ncm. Then, screwing-unscrewing was repeated for 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 times and the healing abutments were scanned under the scanning electron microscope for any distortion. Distortion was graded as 0, 1, 2, and 3. Data were analyzed using SPSS 24.0. Descriptive statistics were calculated. One-way ANOVA with post hoc using Tukey's HSD test was performed to analyze the difference in distortion at different screwing-unscrewing times. A significant level was selected at p-value = 0.05. It was found that distortion healing abutments screws were seen after 32 times screwing and unscrewing. There was a significant difference in the distortion (p-value < 0.05) after 24 times of repeated usage of healing abutment and at 160, 320, and 400 times. No surface distortions were observed at the healing abutment screw head at 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 cycles of screwing-unscrewing. It can be concluded that repeated screwing and unscrewing of the implant-healing abutments causes damage to the healing abutment hex surface. The distortion of healing abutments screws was seen after 32 times screwing and unscrewing. No surface distortions were observed on the healing abutment screw head until 400 times of screwing and unscrewing. Hence, the clinician should be cautious while using the healing abutments repeatedly.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Bacterial Leakage in Implants With Different Implant-Abutment Connections: An In Vitro Study
    Assenza, Bartolomeo
    Tripodi, Domenico
    Scarano, Antonio
    Perrotti, Vittoria
    Piattelli, Adriano
    Iezzi, Giovanna
    D'Ercole, Simonetta
    JOURNAL OF PERIODONTOLOGY, 2012, 83 (04) : 491 - 497
  • [22] Microleakage of the prosthetic abutment/implant interface with internal and external connection: In vitro study
    Larrucea Verdugo, Carlos
    Jaramillo Nunez, Guido
    Acevedo Avila, Ariel
    Larrucea San Martin, Carlo
    CLINICAL ORAL IMPLANTS RESEARCH, 2014, 25 (09) : 1078 - 1083
  • [23] Implant Abutment Deformation During Prosthetic Cylinder Screw Tightening: An In Vitro Study
    Moretti Neto, Rafael Tobias
    Moura, Marcio Silva
    Capello Souza, Edson Antonio
    Rubo, Jose Henrique
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2009, 22 (04) : 391 - 395
  • [24] Bacterial Colonization of the Dental Implant Fixture-Abutment Interface: An In Vitro Study
    Tesmer, Michael
    Wallet, Shannon
    Koutouzis, Theofilos
    Lundgren, Tord
    JOURNAL OF PERIODONTOLOGY, 2009, 80 (12) : 1991 - 1997
  • [25] Implant-Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study
    de Vasconcelos, John Eversong Lucena
    De Matos, Jefferson David Melo
    Queiroz, Daher Antonio
    Lopes, Guilherme da Rocha Scalzer
    De Lacerda, Bruna Caroline Goncalves Vasconcelos
    Bottino, Marco Antonio
    Turssi, Cecilia Pedroso
    Basting, Roberta Tarkany
    do Amaral, Flavia Lucisano Botelho
    Franca, Fabiana Mantovani Gomes
    MATERIALS, 2022, 15 (15)
  • [26] Implant-Abutment Connection Deformation After Prosthetic Procedures: An In Vitro Study
    Micarelli, Costanza
    Canullo, Luigi
    Iannello, Giuliano
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2015, 28 (03) : 282 - 286
  • [27] Manual closing torque in five implant abutment systems: An in vitro comparative study
    Gross, M
    Kozak, D
    Laufer, BZ
    Weiss, EI
    JOURNAL OF PROSTHETIC DENTISTRY, 1999, 81 (05): : 574 - 578
  • [28] The Evaluation of Unscrewing Torque Values of Implant-Abutment Connections: An In Vitro Study
    Bruna, Ezio
    Fabianelli, Andrea
    Mastriforti, Giacomo
    Papacchini, Federica
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2017, 30 (01) : 30 - 32
  • [29] Implant deformation and implant-abutment fracture resistance after standardized artificial aging: An in vitro study
    Zhai, Zhihao
    Tamaki, Nakano
    Chen, Yuming
    Shota, Watanabe
    Takashi, Matsuoka
    Shoichi, Ishigaki
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2023, 25 (01) : 107 - 117
  • [30] Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study
    Bianchini, Marco Aurelio
    De Bortoli Junior, Nilton De Bortoli
    Dedavid, Berenice Anina
    De Aza, Piedad N.
    Gehrke, Sergio Alexandre
    PLOS ONE, 2023, 18 (01):