Healing Abutment Distortion in Implant Prostheses: An In Vitro Study

被引:1
|
作者
Pandey, Chandrashekhar [1 ]
Bhattarai, Bishwa Prakash [1 ]
Riddhabhaya, Apiwat [2 ]
Wongsirichat, Natthamet [1 ]
Rokaya, Dinesh [1 ]
机构
[1] Walailak Univ, Int Coll Dent, Dept Clin Dent, Bangkok 10400, Thailand
[2] Walailak Univ, Int Coll Dent, Dept Oral Biosci, Bangkok 10400, Thailand
关键词
dental implants; implant abutment; healing abutment; screw thread; screw head; wear; distortion; scanning electron microscope; TORQUE; PRELOAD; REMOVAL; CYCLES; SCREWS; CONTAMINATION; MAINTENANCE; MISFIT; REUSE;
D O I
10.3390/jfb13030085
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Little has been studied regarding the repeated usage of healing abutments and their effects on the distortion of dental implant-healing abutment hex surfaces. Additionally, implant manufacturers do not provide specific guidelines on how many times a healing abutment can be used before discarding. Therefore, we evaluated the effect of repeated screwing-unscrewing of implant-healing abutment on the healing abutment hex surface and screw head. A total of 12 Biomate implants with 4 mm diameter and 13 mm length were inserted into a synthetic bone block. The standard healing abutments of 3 mm diameter and 4 mm length were screwed onto each implant using a torque ratchet at a final torque of 30 Ncm. Immediately, the abutments were unscrewed at 30 Ncm. Then, screwing-unscrewing was repeated for 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 times and the healing abutments were scanned under the scanning electron microscope for any distortion. Distortion was graded as 0, 1, 2, and 3. Data were analyzed using SPSS 24.0. Descriptive statistics were calculated. One-way ANOVA with post hoc using Tukey's HSD test was performed to analyze the difference in distortion at different screwing-unscrewing times. A significant level was selected at p-value = 0.05. It was found that distortion healing abutments screws were seen after 32 times screwing and unscrewing. There was a significant difference in the distortion (p-value < 0.05) after 24 times of repeated usage of healing abutment and at 160, 320, and 400 times. No surface distortions were observed at the healing abutment screw head at 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 cycles of screwing-unscrewing. It can be concluded that repeated screwing and unscrewing of the implant-healing abutments causes damage to the healing abutment hex surface. The distortion of healing abutments screws was seen after 32 times screwing and unscrewing. No surface distortions were observed on the healing abutment screw head until 400 times of screwing and unscrewing. Hence, the clinician should be cautious while using the healing abutments repeatedly.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Dental implant healing abutment decontamination: A systematic review of in vitro studies
    Kyaw, Thiha Tin
    Abdou, Ahmed
    Nakata, Hidemi
    Pimkhaokham, Atiphan
    INTERNATIONAL JOURNAL OF ORAL IMPLANTOLOGY, 2022, 15 (04) : 311 - 324
  • [2] Bioactive Healing Abutment as a Potential Tool for the Treatment of Peri-Implant Disease-In Vitro Study
    Iwanczyk, Bartlomiej
    Wychowanski, Piotr
    Minkiewicz-Zochniak, Anna
    Strom, Kamila
    Jarzynka, Sylwia
    Oledzka, Gabriela
    APPLIED SCIENCES-BASEL, 2020, 10 (15):
  • [3] Influence of abutment and connection type on implant-abutment connection rigidity: an in vitro study
    Mendes, Teresa Almeida
    Vilhena, Luis Miguel
    Portugal, Jaime
    Carames, Joao
    Ramalho, Amilcar Lopes
    Lopes, Luis Pires
    REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2022, 63 (03): : 134 - 140
  • [4] Gelatin-Curcumin Nanocomposites as a Coating for Implant Healing Abutment: In Vitro Stability Investigation
    Dizaj, Solmaz Maleki
    Torab, Ali
    Kouhkani, Shadi
    Sharifi, Simin
    Negahdari, Ramin
    Bohlouli, Sepideh
    Fattahi, Shirin
    Salatin, Sara
    CLINICS AND PRACTICE, 2023, 13 (01) : 88 - 101
  • [5] A TECHNIQUE TO MODIFY THE LENGTH OF AN IMPLANT HEALING ABUTMENT
    Yilmaz, Burak
    McGlumphy, Edwin
    Turkyilmaz, Ilser
    JOURNAL OF ORAL IMPLANTOLOGY, 2009, 35 (04) : 201 - 203
  • [6] Bacterial microleakage at the abutment-implant interface, in vitro study
    Larrucea, Carlos
    Conrado, Aparicio
    Olivares, Denise
    Padilla, Carlos
    Barrera, Andrea
    Lobos, Olga
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (03) : 360 - 367
  • [7] Bacterial Microleakage at the Implant-Abutment Interface: An In Vitro Study
    D'Ercole, Simonetta
    Dotta, Tatiane Cristina
    Farani, Marzieh Ramezani
    Etemadi, Niloofar
    Iezzi, Giovanna
    Comuzzi, Luca
    Piattelli, Adriano
    Petrini, Morena
    BIOENGINEERING-BASEL, 2022, 9 (07):
  • [8] Microleakage at the abutment implant interface: A comparative in vitro study.
    Abramovitch, I
    Gross, M
    Weiss, EI
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 3306 - 3306
  • [9] Probing and Scaling Instrumentation on Implant Abutment Surfaces: An In Vitro Study
    Fakhravar, Behnam
    Khocht, Ahmed
    Jefferies, Steven R.
    Suzuki, Jon B.
    IMPLANT DENTISTRY, 2012, 21 (04) : 311 - 316
  • [10] Effect of custom abutment data superimposition on the accuracy of implant abutment level scanning: An in vitro study
    Baek, Yeon-Wha
    Lim, Young-Jun
    Kim, Myung-Joo
    Kwon, Ho-Beom
    JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (06): : 1208 - 1217