We tested an X-ray interferometer fabricated to improve the spatial resolution of phase-contrast X-ray computed tomography (CT). The main factor limiting the spatial resolution is the diffraction phenomenon in the crystal wafer of the interferometer. Improvement of the spatial resolution was attempted by thinning the wafer. A preliminary test was performed by checking interference patterns. Thinning the wafer from 1 to 0.24 mm showed clear improvement in the image quality.
引用
收藏
页码:98 / 103
页数:6
相关论文
共 12 条
[1]  
Ando M, 1972, P 6 INT C XRAY OPTIC, P63
[2]   DYNAMICAL DIFFRACTION OF X RAYS BY PERFECT CRYSTALS [J].
BATTERMAN, BW ;
COLE, H .
REVIEWS OF MODERN PHYSICS, 1964, 36 (03) :681-&
[3]   X-ray microtomography (mu CT) using phase contrast for the investigation of organic matter [J].
Beckmann, F ;
Bonse, U ;
Busch, F ;
Gunnewig, O .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (04) :539-553
[4]   AN X-RAY INTERFEROMETER [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 6 (08) :155-&
[5]  
HIRANO K, UNPUB JPN J APPL PHY
[6]  
Kinoshita K., 1992, Proceedings of the SPIE - The International Society for Optical Engineering, V1741, P287
[7]   The possibility of phase-contrast X-ray microtomography [J].
Momose, A ;
Hirano, K .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 1999, 38 :625-629
[8]   PHASE-CONTRAST RADIOGRAPHS OF NONSTAINED RAT CEREBELLAR SPECIMEN [J].
MOMOSE, A ;
FUKUDA, J .
MEDICAL PHYSICS, 1995, 22 (04) :375-379
[9]   Phase-contrast X-ray computed tomography for observing biological soft tissues [J].
Momose, A ;
Takeda, T ;
Itai, Y ;
Hirano, K .
NATURE MEDICINE, 1996, 2 (04) :473-475