Position detection accuracy of a novel linac-mounted intrafractional x-ray imaging system

被引:18
作者
Fast, Martin F. [1 ]
Krauss, Andreas [1 ]
Oelfke, Uwe [1 ]
Nill, Simeon [1 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, D-69120 Heidelberg, Germany
关键词
intrafractional imaging; in-line geometry; adaptive radiotherapy; tumor tracking; REAL-TIME TUMOR; MULTILEAF COLLIMATOR TRACKING; CONE-BEAM CT; DYNAMIC MLC TRACKING; SIEMENS; 160; MLC; GUIDED RADIOTHERAPY; MOTION COMPENSATION; RESPIRATORY MOTION; KV; LOCALIZATION;
D O I
10.1118/1.3665712
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors have developed a system that monitors intrafractional target motion perpendicular to the treatment beam with the aid of radioopaque markers by means of separating kV image and megavoltage (MV) treatment field on a single flat-panel detector. Methods: They equipped a research Siemens Artiste linear accelerator (linac) with a 41 x 41 cm(2) a-Si flat-panel detector underneath the treatment head. The in-line geometry allows kV (imaging) and MV (treatment) beams to share closely aligned beam axes. The kV source, usually mounted directly across from the flat-panel imager, was retracted toward the gantry by 13 cm to intentionally misalign kV and MV beams, resulting in a geometric separation of MV treatment field and kV image on the detector. Two consecutive images acquired within 140 ms (the first with MV-only and the second with kV and MV signal) were subtracted to generate a kV-only image. The images were then analyzed "online" with an automated threshold-based marker detection algorithm. They employed a 3D and a 4D phantom equipped with either a single radioopaque marker or three Calypso beacons to mimic respiratory motion. Measured room positions were either cross-referenced with a phantom voltage signal (single marker) or the Calypso system. The accuracy of the back-projection (from detected marker positions into room coordinates) was verified by a simulation study. Results: A phantom study has demonstrated that the imaging framework is capable of automatically detecting marker positions and sending this information to the tracking tool at an update rate of 7.14 Hz. The system latency is 86.9 +/- 1.0 ms for single marker detection in the absence of MV radiation. In the presence of a circular MV field of 5 cm diameter, the latency is 87.1 +/- 0.9 ms. The total RMS position detection accuracy is 0.20 mm (without MV radiation) and 0.23 mm (with MV). Conclusions: Based on the evaluated motion patterns and MV field size, the positional accuracy and system latency indicate that this system is suitable for real-time adaptive applications. (C) 2012 American Association of Physicists in Medicine. [DOI: 10.1118/1.3665712]
引用
收藏
页码:109 / 118
页数:10
相关论文
共 43 条
[1]   Accuracy of a wireless localization system for radiotherapy [J].
Balter, JM ;
Wright, JN ;
Newell, LJ ;
Friemel, B ;
Dimmer, S ;
Cheng, Y ;
Wong, J ;
Vertatschitsch, E ;
Mate, TP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (03) :933-937
[2]   Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors [J].
Berbeco, RI ;
Jiang, SB ;
Sharp, GC ;
Chen, GTY ;
Mostafavi, H ;
Shirato, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (02) :243-255
[3]   A robotic approach to 4D real-time tumor tracking for radiotherapy [J].
Buzurovic, I. ;
Huang, K. ;
Yu, Y. ;
Podder, T. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (05) :1299-1318
[4]   REAL-TIME TARGET POSITION ESTIMATION USING STEREOSCOPIC KILOVOLTAGE/MEGAVOLTAGE IMAGING AND EXTERNAL RESPIRATORY MONITORING FOR DYNAMIC MULTILEAF COLLIMATOR TRACKING [J].
Cho, Byungchul ;
Poulsen, Per Rugaard ;
Sawant, Amit ;
Ruan, Dan ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (01) :269-278
[5]   Real-time tumor tracking using sequential kV imaging combined with respiratory monitoring: a general framework applicable to commonly used IGRT systems [J].
Cho, Byungchul ;
Poulsen, Per Rugaard ;
Keall, Paul J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12) :3299-3316
[6]   FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING [J].
Cho, Byungchul ;
Poulsen, Per R. ;
Sloutsky, Alex ;
Sawant, Amit ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03) :859-867
[7]   An analysis of the treatment couch and control system dynamics for respiration-induced motion compensation [J].
D'Souza, Warren D. ;
McAvoy, Thomas J. .
MEDICAL PHYSICS, 2006, 33 (12) :4701-4709
[8]   Estimation of CT cone-beam geometry using a novel method insensitive to phantom fabrication inaccuracy: Implications for isocenter localization accuracy [J].
Ford, J. Chetley ;
Zheng, Dandan ;
Williamson, Jeffrey F. .
MEDICAL PHYSICS, 2011, 38 (06) :2829-2840
[9]   A fast, accurate, and automatic 2D-3D image registration for image-guided cranial radiosurgery [J].
Fu, Dongshan ;
Kuduvalli, Gopinath .
MEDICAL PHYSICS, 2008, 35 (05) :2180-2194
[10]   Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization [J].
Gayou, Olivier ;
Miften, Moyed .
MEDICAL PHYSICS, 2007, 34 (08) :3183-3192