摘要
Background. Procalcitonin (PCT) increases in septic patients, and is not transformed into calcitonin (CT). We found in septic patients, a significant increase of CT, as determined by an immunoassay using polyclonal antibodies. We compare determination using polyclonal and monoclonal AB. Methods. We included 34 patients: 17 with clinical signs of sepsis, a positive haemoculture (PCT > 0.5 mu g/L) and 17 without them (PCT < 0.1 mu g/L). CT was determined by two above-mentioned methods. The influence on CT levels was observed after using the high-concentration PCT calibrator addition to a mixed serum sample with a low concentration of CT. The dilution test of the high-concentration calibrator PCT was performed by an IBL calibrator, with a zero calcitonin concentration. Results. In the septic patients we found an interference in calcitonin determination using the polyclonal AB (IRMA); 24.1-718 mu g/L, proportional to the PCT levels (r = 0.814, p < 0.0001). When using the monoclonal AB (ELISA), the calcitonin levels < 6.5-46.3 ng/L, and no interference of PCT was observed. In the non-septic group, we did not record any PCT interference using either the polyclonal or the monoclonal AB, and the CT levels were within the reference ranges using the two methods (r = 0.997, p < 0.0001). The recovery and dilution tests confirmed interference by PCT on the calcitonin determination with the polyclonal antibody. Conclusions. Results show that in septic patients there is visible interference of PCT in the calcitonin determination, principally in the IRMA method (polyclonal AB); while no such relationship was observed in the ELISA method (monoclonal AB).
引用
收藏
页码:157 / 162
页数:6
相关论文
共 50 条
  • [31] DETERMINATION OF ALPHA-FETOPROTEIN IN SERUM WITH AN AUTOMATED IMMUNOMETRIC ASSAY
    VILJA, P
    TANHUANPAA, P
    TUOHIMAA, P
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1988, 26 (05): : 336 - 336
  • [32] EVALUATION OF A NEW DIGOXIN IMMUNOMETRIC ASSAY IN INFANTS NOT RECEIVING DIGOXIN
    LEE, KR
    LINK, AB
    PHELPS, SJ
    CLINICAL RESEARCH, 1991, 39 (04): : A832 - A832
  • [33] CLINICAL-APPLICATION OF A SENSITIVE NONISOTOPIC IMMUNOMETRIC ASSAY OF THYROTROPIN
    SHEPPARD, MC
    BLACK, EG
    CLINICAL CHEMISTRY, 1987, 33 (01) : 179 - 181
  • [34] Use of free radical chemistry in an immunometric assay for 17β-estradiol
    Buscarlet, L
    Volland, H
    Dupret-Carruel, J
    Jolivet, M
    Grassi, J
    Créminon, C
    Taran, P
    Pradelles, P
    CLINICAL CHEMISTRY, 2001, 47 (01) : 102 - 109
  • [35] Can procalcitonin replace calcitonin as a neuroendocrine tumor marker?
    Preissner, C. M.
    Finseth, M. S.
    Grebe, S. K.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A117 - A118
  • [36] Development and Application of a Novel Sensitive Immunometric Assay for Calcitonin in a Large Cohort of Patients with Medullary and Differentiated Thyroid Cancer, Thyroid Nodules, and Autoimmune Thyroid Diseases
    Camacho, Cleber P.
    Lindsey, Susan C.
    Kasamatsu, Teresa S.
    Machado, Alberto L.
    Martins, Joao Roberto M.
    Biscolla, Rosa Paula M.
    Dias da Silva, Magnus R.
    Vieira, Jose Gilberto H.
    Maciel, Rui M. B.
    EUROPEAN THYROID JOURNAL, 2014, 3 (02) : 117 - 124
  • [37] Performance Evaluation of Affinity Column Mediated Immunometric Assay for Tacrolimus
    Chung, Jae-Woo
    An, Dongheui
    Song, Junghan
    Chung, Hee-Jung
    Park, Hae-Il
    Lee, Woochang
    Chun, Sail
    Min, Won-Ki
    KOREAN JOURNAL OF LABORATORY MEDICINE, 2009, 29 (05): : 415 - 422
  • [38] DIRECT IMMUNOMETRIC ASSAY OF ACTIVE RENIN IN HUMAN-PLASMA
    MENARD, J
    GUYENNE, TT
    CORVOL, P
    PAU, B
    SIMON, D
    RONCUCCI, R
    JOURNAL OF HYPERTENSION, 1985, 3 : S275 - S278
  • [39] Usefulness of a rapid immunometric assay for intraoperative parathyroid hormone measurements
    Ohe, MN
    Santos, RO
    Kunii, IS
    Carvalho, AB
    Abrahao, M
    Cervantes, O
    Lazaretti-Castro, M
    Viera, JGH
    BRAZILIAN JOURNAL OF MEDICAL AND BIOLOGICAL RESEARCH, 2003, 36 (06) : 715 - 721
  • [40] Performance Evaluation of an Affinity Column Mediated Immunometric Assay for Tacrolimus
    Chung, Hee-Jung
    Chung, Jae-Woo
    An, Dongheui
    Song, Junghan
    Park, Hae-Il
    Lee, Woochang
    Chun, Sail
    Min, Won-Ki
    THERAPEUTIC DRUG MONITORING, 2009, 31 (05) : 639 - 639