Matrix effect; Isotope-dilution mass spectrometry; Deuterium-labeled internal standard; Pesticide; Matrix matching; Calibration curve; CERTIFIED REFERENCE MATERIAL; GEL-PERMEATION CHROMATOGRAPHY; ON-COLUMN INJECTION; RESIDUE ANALYSIS; PYRETHROID PESTICIDES; RESPONSE ENHANCEMENT; ORGANOPHOSPHORUS; SAMPLES; PERFORMANCE; SPLITLESS;
D O I
10.1016/j.chroma.2015.03.075
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The impact of the matrix effect in GC-MS quantification of pesticides in food using the corresponding isotope-labeled internal standards was evaluated. A spike-and-recovery study of nine target pesticides was first conducted using paste samples of corn, green soybean, carrot, and pumpkin. The observed analytical values using isotope-labeled internal standards were more accurate for most target pesticides than that obtained using the external calibration method, but were still biased from the spiked concentrations when a matrix-free calibration solution was used for calibration. The respective calibration curves for each target pesticide were also prepared using matrix-free calibration solutions and matrix-matched calibration solutions with blank soybean extract. The intensity ratio of the peaks of most target pesticides to that of the corresponding isotope-labeled internal standards was influenced by the presence of the matrix in the calibration solution; therefore, the observed slope varied. The ratio was also influenced by the type of injection method (splitless or on-column). These results indicated that matrix-matching of the calibration solution is required for very accurate quantification, even if isotope-labeled internal standards were used for calibration. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:109 / 116
页数:8
相关论文
共 32 条
[1]   Stability Monitoring of Pesticide Residues in a Chinese Cabbage Certified Reference Material [J].
Ahn, Seonghee ;
Kim, Byungjoo ;
Hwang, Eujin .
BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2011, 32 (04) :1365-1367
[2]  
[Anonymous], GUID VAL AN METH AGR
[3]  
[Anonymous], J FOOD HYG SOC JPN
[4]   13C labelled internal standards-A solution to minimize ion suppression effects in liquid chromatography-tandem mass spectrometry analyses of drugs in biological samples? [J].
Berg, Thomas ;
Strand, Dag Helge .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (52) :9366-9374
[5]   An efficient GC-IDMS method for determination of PBDEs and PBB in plastic materials [J].
Binici, Burcu ;
Bilsel, Mine ;
Karakas, Melis ;
Koyuncu, Ikbal ;
Goren, Ahmet C. .
TALANTA, 2013, 116 :417-426
[6]   Agriculture, pesticides, food security and food safety [J].
Carvalho, Fernando P. .
ENVIRONMENTAL SCIENCE & POLICY, 2006, 9 (7-8) :685-692
[7]   A multinational joint project on the evaluation of residual pesticide analysis in tea in the Asia Pacific region [J].
Dayarathna, Disni ;
Thirimanna, Champika ;
Mubarak, Azeez ;
Mackay, Lindsey ;
Rogerson, Jill ;
Tang, Hua ;
Feng, Jie ;
Sin, Della W. M. ;
Wong, Yiu-chung .
FOOD RESEARCH INTERNATIONAL, 2013, 53 (02) :931-937
[8]   EXPLANATION OF THE MATRIX-INDUCED CHROMATOGRAPHIC RESPONSE ENHANCEMENT OF ORGANOPHOSPHORUS PESTICIDES DURING OPEN-TUBULAR COLUMN GAS-CHROMATOGRAPHY WITH SPLITLESS OR HOT ON-COLUMN INJECTION AND FLAME PHOTOMETRIC DETECTION [J].
ERNEY, DR ;
GILLESPIE, AM ;
GILVYDIS, DM ;
POOLE, CF .
JOURNAL OF CHROMATOGRAPHY, 1993, 638 (01) :57-63
[9]   Matrix-induced effects:: a critical point in the gas chromatographic analysis of pesticide residues [J].
Hajslová, J ;
Holadová, K ;
Kocourek, V ;
Poustka, J ;
Godula, M ;
Cuhra, P ;
Kempny, M .
JOURNAL OF CHROMATOGRAPHY A, 1998, 800 (02) :283-295
[10]   Effect of residues remaining in the injection liner of a gas chromatograph on the quantification of polycyclic aromatic hydrocarbons by isotope dilution mass spectrometry using deuterium-labeled internal standards [J].
Itoh, Nobuyasu ;
Numata, Masahiko ;
Aoyagi, Yoshie ;
Yarita, Takashi .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1134 (1-2) :246-252