Implementation of a High-Rate USB Data Acquisition System for PET and SPECT Imaging

被引:16
作者
Proffitt, J. [1 ]
Hammond, W. [1 ]
Majewski, S. [1 ]
Popov, V. [1 ]
Raylman, R. R. [2 ]
Weisenberger, A. G. [1 ]
机构
[1] Thomas Jefferson Natl Accelerator Facil, Detector & Imaging Grp, Newport News, VA 23606 USA
[2] West Virginia Univ, Ctr Adv Imaging, Dept Radiol, Morgantown, WV USA
来源
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6 | 2006年
关键词
D O I
10.1109/NSSMIC.2006.356521
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We made substantial progress with a flexible high-rate USB data acquisition system developed for gamma-ray imaging detectors. Hardware consists of 16-channel data acquisition modules installed on USB carrier boards. One, two, and four-module units were developed. USB data rate was increased to over 30 MB/s and a 16-channel configuration achieved a trigger rate of over 700 kHz. Several high-resolution single-gamma detectors and two high-rate PET detectors were instrumented. The detectors use various configurations of Hamamatsu H8500, H9500, and Burle 85002-800 PSPMTs. System channels were expanded by synchronizing additional acquisition units. A Java client-server system was developed to link acquisition computers over Gigabit Ethernet. A Kmax tool was developed to process and display images during acquisition. C and Java utilities were developed to assist development and diagnostics.