Icare-Pro Rebound Tonometer Versus Hand-held Applanation Tonometer for Pediatric Screening

被引:12
作者
Arribas-Pardo, Paula [1 ]
Mendez-Hernandez, Carmen [2 ]
Valls-Ferran, Isabel [1 ]
Puertas-Bordallo, Diego [1 ]
机构
[1] Univ Autonoma Madrid, Hosp Infantil Univ Nino Jesus, Dept Ophthalmol, Madrid, Spain
[2] Univ Complutense Madrid, Dept Ophthalmol, Hosp Clin San Carlos, Inst Hlth Care Res, Madrid, Spain
关键词
NORMAL INTRAOCULAR-PRESSURE; GOLDMANN APPLANATION; NONCONTACT TONOMETER; CHILDREN;
D O I
10.3928/01913913-20180621-01
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare intraocular pressure (IOP) measurements obtained using the new rebound tonometer Icare-Pro (Icare,Tiolat Oy, Helsinki, Finland) and the hand-held version of the Goldmann applanation tonometer (Perkins; Clement Clarke, Haag-Streit, Harlow, United Kingdom) in healthy children during clinical practice. Methods: In this prospective study, three IOP measurements were made using each tonometer in a single session, starting with the 'care-Pro. Participants were 173 non-anesthetized patients aged 1 to 16 years. Measurements were made in both eyes but only data for the right eye were entered in the analysis. Central corneal thickness, anterior chamber depth, and axial length were also measured in each patient. Data were compared by determining interclass correlation coefficients (ICCs) for each tonometer and representing the differences detected as Bland-Altman plots. Results: Good linear correlation was observed between IOP readings obtained using the Perkins and (care-Pro tonometers (r = 0.87, P < .001), although the 'care-Pro readings were slightly higher (mean IOP difference: 0.26 +/- 1.58 mm Hg, P = .037). The 95% limits of agreement between the two methods were 2.8 to -3.4 mm Hg. The ICC was 0.857 (95% confidence interval: 0.810 to 0.893), indicating good agreement. For both tonometers, a low but significant correlation was detected between IOP and central corneal thickness or age. However, no correlation of IOP was found with axial length or anterior chamber depth. Conclusions: Pediatric IOP measurements determined using the new Icare-Pro rebound tonometer showed good correlation with those obtained using the handheld Perkins applanation tonometer in a routine clinical examination with no need for general anesthesia.
引用
收藏
页码:382 / 386
页数:5
相关论文
共 19 条
[1]   A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma [J].
AlHarkan, Dora H. ;
Al-Shamlan, Fatemah T. ;
Edward, Deepak P. ;
Khan, Arif O. .
MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2016, 23 (02) :183-186
[2]   Measuring Intraocular Pressure After Intrastromal Corneal Ring Segment Implantation With Rebound Tonometry and Goldmann Applanation Tonometry [J].
Arribas-Pardo, Paula ;
Mendez-Hernandez, Carmen ;
Cuina-Sardina, Ricardo ;
Fernandez-Perez, Cristina ;
Diaz-Valle, David ;
Garcia-Feijoo, Julian .
CORNEA, 2015, 34 (05) :516-520
[3]   Rebound tonometry versus Goldmann tonometry in school children: feasibility and agreement of intraocular pressure measurements [J].
Barranco Esporcatte, Bruno Leonardo ;
Santos Lopes, Flÿvio Siqueira ;
Netto, Camila Fonseca ;
Reboucas-Santos, Vespasiano ;
Dias, Diego Torres ;
Marujo, Fabio Iglesias ;
Rolim-de-Moura, Christiane .
ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2015, 78 (06) :359-362
[4]   The Icare-Pro Rebound Tonometer Versus the Hand-held Applanation Tonometer in Congenital Glaucoma [J].
Borrego Sanz, Lara ;
Morales-Fernandez, Laura ;
Martinez de-la-Casa, Jose M. ;
Saenz-Frances, Federico ;
Fuentes, Manuel ;
Garcia-Feijoo, Julian .
JOURNAL OF GLAUCOMA, 2016, 25 (02) :149-154
[5]   Rebound tonometry over an air-filled anterior chamber in the supine child after intraocular surgery [J].
Davis, Ryan ;
Jiramongkolchai, Kim ;
Silverstein, Evan ;
Freedman, Sharon F. .
JOURNAL OF AAPOS, 2016, 20 (02) :159-164
[6]   Intraocular Pressure in Children: The Effect of Body Position as Assessed by Icare and Tono-Pen Tonometers [J].
Dosunmu, Eniolami O. ;
Marcus, Inna ;
Tung, Irene ;
Thiamthat, Warakorn ;
Freedman, Sharon F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 158 (06) :1348-1352
[7]   Icare ONE Rebound Versus Goldmann Applanation Tonometry in Children With Known or Suspected Glaucoma [J].
Gandhi, Nandini G. ;
Prakalapakorn, Sasapin G. ;
El-Dairi, Mays A. ;
Jones, Sarah K. ;
Freedman, Sharon F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 154 (05) :843-849
[8]   Comparison of the Icare rebound tonometry with the Goldmann applanation tonometry in a pediatric population [J].
Grigorian, Florin ;
Grigorian, A. Paula ;
Li, Ang ;
Sattar, Abdus ;
Krishna, Rohit ;
Olitsky, Scott E. .
JOURNAL OF AAPOS, 2015, 19 (06) :572-574
[9]   Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects [J].
Guler, Mete ;
Bilak, Semsettin ;
Bilgin, Burak ;
Simsek, Ali ;
Capkin, Musa ;
Reyhan, Ali Hakim .
JOURNAL OF GLAUCOMA, 2015, 24 (08) :613-618
[10]  
Hladikova E, 2014, Cesk Slov Oftalmol, V70, P90