Portable High-Frequency Electrodynamic Force Standards

被引:0
作者
Chijioke, Akobuije [1 ]
Vlajic, Nicholas [1 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
来源
2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018) | 2018年
关键词
dynamic force; electrodynamic; force calibration; force metrology; force standard; kibble balance; watt balance;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We describe high-frequency electrodynamic force standards. Such standards can be calibrated voice coils, or high-frequency Kibble (watt) balances. We discuss the design of such devices and present example operating parameters for a 10 N Kibble balance force standard designed for operation to frequencies up to 20 kHz.
引用
收藏
页数:2
相关论文
共 4 条
[1]  
Chijioke A., HIGH BANDWIDTH UNPUB
[2]   A watt balance based on a simultaneous measurement scheme [J].
Fang, H. ;
Kiss, A. ;
Picard, A. ;
Stock, M. .
METROLOGIA, 2014, 51 (02) :S80-S87
[3]  
Kibble B.P., 1976, Atomic Masses and Fundamental Constants 5, P545, DOI [10.1007/978-1-4684-2682-380, DOI 10.1007/978-1-4684-2682-380]
[4]   A simultaneous moving and weighing technique for a watt balance at room temperature [J].
Robinson, I. A. .
METROLOGIA, 2012, 49 (01) :108-112