Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups

被引:90
作者
Currier, Barbara H. [1 ]
Currier, John H. [1 ]
Mayor, Michael B. [1 ]
Lyford, Kimberly A. [1 ]
Collier, John P. [1 ]
Van Citters, Douglas W. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
D O I
10.2106/JBJS.F.00336
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Crossfire cross-linked polyethylene is produced differently from other cross-linked polyethylene materials; a below-melt-temperature annealing process is used with the goal of avoiding compromised mechanical properties. The present study was performed to evaluate retrieved Crossfire acetabular cups to determine whether they had oxidized and to what extent oxidation might have influenced their clinical performance. Methods: Eleven acetabular cups were received at retrieval and a twelfth acetabular cup was received two years post-retrieval over a period of four years. None were retrieved because of polyethylene wear or fatigue. The cups had been in vivo from 0.1 to 5.3 years. Each was examined visually, clinical fatigue damage was rated, and oxidation was measured with use of Fourier transform infrared spectroscopy. Results: The cups exhibited oxidation that varied with its location on the cup: the oxidation value was generally low on the articular surface but more than an order of magnitude higher value on the rim. Maximum rim oxidation correlated significantly with the time in vivo (Spearman rho = 0.734, p = 0.010). Oxidation was identified visually by a white band in thin sections on the rim of seven of the cups and on the articular surface of one of these seven cups. Six of the seven cups also exhibited clinical fatigue damage. Eight of the twelve cups exhibited evidence of impingement or dislocation. Conclusions: Acetabular cups made of Crossfire polyethylene oxidized to a measurable degree. The oxidation-related reduction of polyethylene mechanical properties was sufficient to allow the fatigue damage seen in these retrieved cups. Clinical Relevance: The oxidation measured in acetabular cups made of Crossfire polyethylene makes them susceptible to fatigue after as little as three years in service.
引用
收藏
页码:2023 / 2029
页数:7
相关论文
共 18 条
  • [1] *ASTM, 2000, F64800E1 ASTM
  • [2] *ASTM, 2003, 210201 ASTM F, P1680
  • [3] Comparison of cross-linked polyethylene materials for orthopaedic applications
    Collier, JP
    Currier, BH
    Kennedy, FE
    Currier, JH
    Timmins, GS
    Jackson, SK
    Brewer, RL
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (414) : 289 - 304
  • [4] Currier BH, 2000, J BIOMED MATER RES, V53, P143, DOI 10.1002/(SICI)1097-4636(2000)53:2<143::AID-JBM3>3.0.CO
  • [5] 2-5
  • [6] CURRIER BH, 2007, IN PRESS J ARTHROPLA
  • [7] Five-year experience with Crossfire® highly cross-linked polyethylene
    D'Antonio, JA
    Manley, MT
    Capello, WN
    Bierbaum, BE
    Ramakrishnan, R
    Naughton, M
    Sutton, K
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) : 143 - 150
  • [8] A review of current cross-linked polyethylenes used in total joint arthroplasty
    Harris, WH
    Muratoglu, OK
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (430) : 46 - 52
  • [9] RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES
    HOOD, RW
    WRIGHT, TM
    BURSTEIN, AH
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05): : 829 - 842
  • [10] In vivo degradation of polyethylene liners after gamma sterilization in air
    Kurtz, SM
    Rimnac, CM
    Hozack, WJ
    Turner, J
    Marcolongo, M
    Goldberg, VM
    Kraay, MJ
    Edidin, AA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (04) : 815 - 823